Home > MSI > Scale > MSI 3460 Challenger 3 Manual

MSI 3460 Challenger 3 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual MSI 3460 Challenger 3 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 62 MSI manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							USER
    F8) To enter a calibrationweight other than
    capacity, press  USER. T h e
    display far left digit will
    flash zero indicating that
    a number should be
    entered.
    blinking
    Error Correction: If you input a wrong
    value, press  ZERO to step back one
    digit and reenter.
    USER
    FUSER
    F
    TARE
    
    USER
    F
    TARE
    
    
    TAREUSER
    F
    TAREUSER
    F
    5 times
    USER
    F
    TARE
    fixed blinking blinking
    blinking
    blinking blinkingblinking
    fixed
    fixed
    fixed3DVVTARE
    9) Press the  USER key to
    scroll the number and
    the TARE  key to enter
    each digit of the
    calibration weight.  
    In this example, we’ll enter 2500
    kg on a 5000 kg capacity scale.
    Do not push the  TARE key two
    times in a row until the entire
    number is showing.
    To add a decimal point, push the
    POWER  key while the number is
    blinking.
    10) When the entire value of
    the test weight is displayed
    and the weight and scale
    are stable, press  TARE to
    finish off the weight entry.
    If the resultant cal value is
    within limits, the display will
    read PASS briefly.
    11) The display now reads Load2. The 3460
    allows multipoint
    calibration. If more cal
    points are desired (up to
    3 additional) press
    TARE. If a single point
    cal is all that’s needed,
    pressZERO  and jump to
    step 15.
    12) Load the scale with the next test weight. The display
    offers the reading that will
    occur without adjusting the
    value. If the weight value
    shown is acceptable push
    TARE and skip to step 14.
    If the calibrated value needs
    adjusting, go to step 13.
    Test weights can be any in-capacity
    weight, the order is not important.
    The 3460 supports 5-Point Cal, Zero
    and up to 4 span points.
    13) Press the  USER key to
    scroll the number and
    the TARE  key to enter
    each digit of the
    calibration weight value.
    When the entire calibration
    weight is displayed, push
    TARE  a second time to
    finalize the calibration span
    point. If the resultant cal
    value is within limits, the
    display will read “PASS”
    briefly.
    
    Example Value only.
    /RDGTARE
    To set Next
    Span Point
    To skip additional
    Span Points
    USER
    F
    ZERO
    To scroll the digit
    TARE
    To Enter the digit
    into the cal weight
    value
    TARE
    To complete the cal
    weigh value entry3DVV
    You can cancel calibration by
    pressing Power and the scale
    will reset to the previous
    calibration constants.
    Calibration      27
    Figure 5-2. Standard Calibration Procedure, Continued. 
    						
    							14) The display now readsLoAd3 or  LoAd4.
    Repeat steps 12 and 13
    for additional span points
    press TARE or press
    ZERO  if you are finished.
    After LoAD4 has passed, the 3460
    will automatically advance to the next
    step.
    /RDG
    /RDG
    or
    TARE
    To enter another
    span point
    15) The display shows  CAL’d
    to indicate that the
    calibration was successful.
    16) Press TARE. The display
    flashes C-CAL followed
    by the C-CAL number.
    Make a note of this number.
    If the X1000 annunciator is lit,
    also make a note of that.
    This number is used to calibrate when
    test weights are unavailable. C-CAL
    does not replicate multipoint
    calibrations.
    17) Press  ZEROto store the
    calibration constants, or
    after eight seconds, the
    calibration constants will
    automatically stored. After
    the store message is
    displayed, the scale goes
    to the next item on the cal
    menu, SEtuP.
    18) Press ZERO to exit the Cal
    menus and start up the
    standard weight display.
    The 3460 is ready for use.
    19) After ensuring the calibration is acceptable,
    replace the hex seal screw.
    If the scale is being used as a Legal-
    for-Trade device, place a lead-wire
    seal through the cal screw over to
    the adjacent screw.
      &&D/
    Example Value only.
     VWRUH
    ZERO
    ZERO
    &$/G
    6HWYS
    TAREFlashing
    ZERO
    LO BATT
    CAL
    X1000
    MSI-3460Model No.
    To skip the C-CAL step
    28     MSI-3460 Operator’s Manual
    Figure 5-3. Standard Calibration Procedure, Continued. 
    						
    							Calibration     29
    5.1 Initial Calibration
    Use this procedure only if the capacity and count by (d) needs to be modified. The 
    initial steps of the initial calibration will  totally erase user setups as well as any 
    previous calibration. 
    1)   Turn the 3460 off.
    2) Remove the seal screw. With  a small screwdriver press and 
    hold the  CAL switch behind 
    the seal screw.
    3) Still holding down the  CAL
    switch, press the  POWER
    switch.
    4) The display blinks  rESEt. 
    5) To reset all calibration  constants and setup 
    parameters, press  TARE.
    6) The 3460 requests a  confirmation by displaying 
    Sure? .
    To cancel the Reset press 
    the  POWER  key.
    7) To complete the reset,  press 
    TARE . The Calibrate menu 
    appears. You must now 
    recalibrate the system. 
    8) Press TARE to start the initial 
    calibration  procedure. The 
    display shows Unit. You 
    select the units you wish to 
    calibrate in.
    9) Press TARE to select the  calibration unit.
    10) Select the  USER
     key to choose between lb and kg. When the 
    desired unit is shown, press 
    TARE.
    11)  Next, set the capacity in the  selected units. Capacity must 
    be set no higher than the load 
    cell rated capacity.
    12) Press TARE to enter the 
    capacity setting screen. A 
    capacity of 10000 is the initial 
    value.
    If 10000 units is the desired 
    capacity press TARE and skip to 
    step 16. 
    13) To change the capacity, press  USER .
    Initial CalibrationResetting Capacity and Countby (d)
    Cal Switch
    LO BA
    CAL
    Remove
    Hex Seal
    ScrewPress Switch
    behind the hole
    with a small
    ScrewdriverCAL
    UHVHWblinking
    6YUH
    &$/
    8QLW
    8QLWblinking
    kg    lbUSER
    F
    &$3
    TARE
    TARE
    TARE
    TARE
    TARE
    TARE
    blinking
    USER
    Fblinking
    Error Correction: If you input a 
    wrong value, press  ZERO to step 
    back one digit and reenter. 
    To enter a decimal point, push 
    POWER  while the digit is 
    blinking.
    Figure 5-4. Initial Calibration 
    						
    							30    MSI-3460 Operator’s Manual
    Figure 5-5. 
    14) The first digit blinks. Use the 
    USER  key to scroll through 
    the numbers. When the 
    desired number is shown, 
    push TARE.
    In this example, we’ll enter 
    2500 as a capacity.
    USER
    FUSER
    F
    TAREUSER
    Ffixed blinking blinking
    blinking
    
    TARE
    
    
    TAREUSER
    F
    TAREUSER
    F
    5 times
    USER
    F
    TAREblinking
    blinking
    blinkingfixed
    fixed
    fixed
    Continue inputing the desired 
    capacity using the  USER key 
    for scrolling the number and 
    the TARE  key to store the 
    number.
    15) Finalize the capacity value  by pressing the  TA R E
     key on an unblinking display. In our 
    example, once the number 
    2500 is fixed on the display, 
    pressTARE to store the 
    capacity value.
    16) Next the scale division size  ‘d’ is set. Press  TARE to 
    begin.
     In this example we’ll set 
    the ‘d’ to 2 units.
    17) Select the  USER key to 
    scroll through the 
    recommended scale 
    divisions.
    The first ‘d’ offered is 
    the standard division for the given 
    capacity. Setting a ‘d’ size that 
    results in total resolution higher than 
    1:5000 is not recommended for 
    stability reasons. 
    18) When the desired scale ‘d’  is displayed, press  TA R E.The UnLd display 
    appears and the scale is 
    ready for calibration. 
    Follow the standard 
    calibration procedure 
    starting at step 3.
    TAREG
    TARE
    blinkingUSER
    FblinkingUSER
    F
    XQ/G
    TARE
    Capacity is set.
    ‘d’ is set. Proceed to Standard 
    Calibration starting at Step 3.
    Initial Calibration, Continued 
    5.2 Guidelines for Capacity and Resolution
    Crane scales are subject to forces that regular floor scales do not see. Many bridge 
    cranes, hoist cranes, and mobile cranes la
    ck rigidity and tend to bounce or swing 
    when loads are lifted.  For this reason, MSI recommends that resolution is kept in the 
    1:2000 to 1:3000 range. Some improvement in stability can be achieved by increasing 
    the filtering. However, you should never program resolution that is far greater than 
    you need. If the  MSI-3460 display is never stable, it is recommended that the 
    resolution is reduced and/ or filtering increased.  
    						
    							Calibration     31
    Due to LFT requirements and general scale de sign criteria, the weight must be stable 
    for certain features to work: ZERO – weight  must be stable to be zeroed. TARE – 
    weight must be stable to be tared. TOTAL – weight must be stable to be added to the 
    total registers. One way to improve the stability is to increase the filtering, at the risk 
    of increasing settling time. The other is to increase the ‘d’ (reduce resolution). The 
    third way is to increase the 
    motion window. The  MSI-3460  defaults to ±1d as a 
    motion window. It can be changed at MSI to  a higher value if desired. Often ±3d is 
    chosen for bridge cranes as these tend to ha ve a lot of bounce to them. This of course 
    carries an accuracy penalty adding ±3 d to the total accuracy of the scale if the zero or 
    tare operation happens to capture  the weight in a valley or peak.
    Setting capacity is dictated primarily by th e 
     capability of the load cell. MSI supplies 
    the  MSI-3460  in many capacities. 
    Never set the capacity of the scale higher than 
     the rating of the load cell.
    Due to excellent linearity of the MSI S-beam  load 
     cell, it is acceptable to set lower 
    capacities to better match the crane the  MSI-3460 is used on. For example, if the hoist 
    is rated for 9000 lb. you can use an MSI 10000 lb. MSI 3460 and reset the capacity to 
    9000 lb. so that the scale will indicate overload at 9000 lb. instead of 10000 lb. 
    Derating as much as 50% of the capacity is usually acceptable, but the scale may be 
    less stable if the ‘d’ is decreased.
    The capacity of all  MSI-3460 systems 
     is rated approximately 20% higher than the 
    rated capacity in pounds. This is to allow the kg capacity to be exactly 1/2 the number 
    of the pounds capacity. 
    5.3 C-Cal Calibration
    When adequate test weight s are not available, the MSI-3460 can be calibrated using a 
    programmed constant calibration which is referr ed to as C-Cal. To use C-Cal, a C-Cal 
    number must be known from a previous  calibration. MSI supplies replacement load 
    cells for the  MSI-3460 with the C-Cal value stamped on the serial number label. 
    When a calibration is performed with test weights, a new C-Cal is generated. C-Cal 
    can be used when the electronics are repl aced to get an approximate calibration that 
    may be suitable for non Legal-for-Trade applications.
    ImportantThe C-Cal number must be known prior to starting this 
    procedure. For a MSI-3460 with its original load cell, MSI prints 
    this number on the calibration r ecord, the serial number tag, and 
    on the calibration log found inside the battery compartment. 
    C-Calibration reduces slightly the absolute accuracy of th e system if the electronics 
    are replaced or a new load cell is installed, and is intended for non-critical use only. 
    Legal-for-Trade installations require that the  MSI-3460 is calibrated using test 
    weights. If a system was originally multi-point calibrated, the C-CAL calibration will 
    erase the additional span points, as C-Cal is only a two point calibration (Zero and 
    Span at 10% of capacity). 
    						
    							TARE
     XQ/G
    
    &&$/
    TARE
    USER
    F
    1)Remove the seal screw. With
    a small screwdriver press the
    cal switch behind the seal
    screw.
    The CAL setup menu
    appears.
    2)Press the  USER key to
    scroll to the C-Cal menu
    selection.
    3)Press TARE to start the C-
    Cal procedure.
    If this calibration is using a cleared
    3460, the INITIAL CALIBRATION
    procedure will start here. You must
    set the cal unit, capacity, and count-
    before proceding to step 4,
    4)The display reads  UnLd
    (unload) indicating you should
    remove all weight from the
    hook.
    5)Press TARE. The 3460 sets
    the zero calibration point.
    6)If the zero is in range, the scale
    will display PASS.
    7)Next the display reads
    CCAL? to ask that you look
    up the C-Cal value to be used.
    8)Press TARE. The 3460 is
    ready for numeric entry of the
    C-CAL value.
    9)To enter an C-Cal value
    press USER. The display
    far left digit will flash zero
    indicating that a number
    should be entered.
    &$/
    
    C-Cal Calibration Procedure
    blinking
    3$66
    No weight
    on hook
    Flashes
    Zero
    Input C-Cal value
    blinking
    Cal Switch
    LO BA
    CAL
    Remove
    Hex Seal
    ScrewPress Switch
    behind the hole
    with a small
    Screwdriver
    &&D/
    USER
    F
    TAREblinking
    CAL
    32     MSI-3460 Operator’s Manual
    Figure 5-6. C-Cal Calibration Procedure 
    						
    							TARE
    TARE
    TARE
    USER
    F
    10) Press the USER key to
    scroll  the number and the
    TARE  key to enter each
    digit of the C-Cal value.
    In this example, we’ll enter the C-
    Cal value of 2503. However, it
    should be noted that most C-CAL values are 5 digits.
    Do not push the  TARE key two
    times in a row until the entire
    number is entered. C-Cal’s lessthan 4 digits will not be accepted.
    If the number is too small the
    display will read “LittL” and goback to the “C-CAL” message.
    Error Correction during number
    entry: If you input a wrong value,
    press  ZERO to step back one
    digit and reenter.
    For the rare C-CAL number that
    includes a decimal point  and the
    X1000 annunciator, press the POWER  key to enter the decimal
    point followed by an additional press of the POWER key to inputthe X1000 factor.
    11) Once the entire number
    is entered (not blinking)
    press the  TARE key to
    complete the entry of the
    C-Cal value.
    12) If the Calibration was successful, the display
    will read  PASS followed
    by  CAL’D.
    13) C-Cal is complete. Press ZERO . The next cal
    setup menu item
    appears.
    14) Press ZERO to store the
    calibration and return to
    scale operation.
    USER
    FUSER
    F
    TARE
    
    USER
    F
    
    USER
    F
    5 times
    USER
    F
    TARE
    fixed blinking blinking
    blinking
    blinking blinkingblinking
    USER
    F
    USER
    Fblinking
    3$66TARE
    ZEROStore and return toweight display
    6WRUH
    fixed
    
    blinking
    blinking
    fixed
    fixed
    fixed
    USER
    Fblinking
    6HWYS
    &$/·
    ZEROnext Menu item
    Calibration      33
    Figure 5-7. C-Cal Calibration Procedure, Continued. 
    						
    							34    MSI-3460 Operator’s Manual
    5.4 Calibration Setup Menu
    The calibration setup menu contains two items: standard and auto zero maintenance 
    “Auto0.” In addition, more menus will appear  that are transferred from the main setup 
    menu when Legal for Trade settings are used.
    &$/
    &&D/
    6HWYS
    6WDQG
    $YWR
    LQGYV +E U
    XQLW
    2Q 2IICal Standard
    Auto Zero
    Maintenance
    Calibrate with Weights
    Calibrate with Constant Industrial Setting 
    (default)
    Handbook 44, NTEP
    R-76, OIML
    Industrial, restricted to
    1 Weight Unit
    AZM On
    AZM Off
    USERF
    Cal Switch
    LO BA
    CAL
    Remove
    Hex Seal
    ScrewPress Switch
    behind the hole
    with a small
    ScrewdriverCAL
    Figure 5-8. Calibration Setup 
    5.5 Standard
    There are four selections  in the standard menu.
    1. Industrial (Indu5) - This is the most common setting for the  MS
     I-3460. With the 
    industrial standard, you have full range zero, access to units switching, filters, 
    and peak hold.
    2. Handbook 44 (Hb-44) - Sets the scale to  enabl
     e only approved features per the 
    NTEP HB-44 rules and regulations. Access is denied to peak hold, and the zero 
    range may be limited. The filter menu is moved to the cal setup menu, so filters 
    are only accessible through the cal seal. 
    3. R-76 (r-76) - Sets the scale to enable  o
     nly approved features per OIML R-76. 
    Only kg weight units are available. The zero range is limited to 5% (-2+3% 
    relative to calibrate zero).  Net/Gross function is temporary. Once net weight is 
    established, pushing an F key set for Net/Gross will cause a maximum five 
    second display of the gross weight. You  must clear the tare to display gross 
    weight constantly. Other metrologica l aspects are changed to meet R-76 
    requirements.
    4. One Unit (1unit)-The one unit standard is exactly the same as industrial, except  uni
     ts switching is inhibited. This is useful for metric only countries. Another use 
    of the one unit standard is to allow the scale to be calibrated in units other than 
    lb or kg, since convers ions are eliminated.
    Contact MSI for more information on the standards settings.  
    						
    							TARE
    If you are already in the Cal menu 
    skip to step 2
    1) Remove the seal screw. With  a small screwdriver press the 
    cal switch behind the seal 
    screw. The CAL setup menu 
    appears.
    2) Press the  USER key to 
    scroll to  Setup.
    3) Press TARE to enter the cal 
    setup menu.
    4) Press TARE to enter the 
    standard menu. The current 
    standard setting is displayed.
    5) Scroll to the desired standard  pressing the  USER key. 
    6) Press TARE to set the 
    standard. The display goes to 
    the next item in the CAL setup 
    menu.
    7) Either press  ZERO twice to 
    exit setup and store all 
    changes, or continue to 
    another setup menu item 
    using the  USER key.
    &$/
    Weight Standard Setup
    Cal Switch
    LO BA
    CAL
    Remove
    Hex Seal
    ScrewPress Switch
    behind the hole
    with a small
    Screwdriver
    VHWYS
    6WDQG
    LQGXVTARE
    XQLW
    $YWRTARENext Cal Setup Menu Item
    ZERO
    Store and return to 
    weight display
    &D/
    ZERO
    6WRUH
    Calibrate Menu
    CAL
    USER
    F
    USER
    F
    Calibration      35
    Figure 5-9. Weight Standard Setup 
    5.6 Auto Zero Maintenance
    The MSI-3460  employs an autozeroing maintenanc e mechanism to adjust the zero 
    reading to the center-of-zero (COZ). COZ is defined as the weight reading is within 1/
    4 ‘d’ of zero. AZM continuously adjusts zer o to maintain COZ. It is recommended 
    that AZM is on to maintain the highest acc uracy. However, there are circumstances 
    when it should be turned off. This can happen when minor variations of weight occur 
    while picking up scale attach ments and the variations fall within the AZM capture 
    window. The AZM capture window (usually 1 ‘d’) and capture time (usually 8 
    seconds) can be adjusted by MSI to meet  custom requirements. The settings of AZM 
    are dictated in Legal for Trade  standards and cannot be adjusted.  
    						
    							TARE
    If you are already in the Cal menu 
    skip to step 2
    1) Remove the seal screw. With  a small screwdriver press the 
    cal switch behind the seal 
    screw. The CAL setup menu 
    appears.
    2) Press the  USER key to 
    scroll to  Setup.
    3) Press TARE to enter the cal 
    setup menu.
    4) Press USER to scroll to the 
    Auto0 menu.
    5) Press TARE to enter the auto-
    zero menu. The current setting 
    is displayed (blinking)
    6) Change the setting between  On or OFF pressing the 
    USER key.
    7) Press TARE to set the 
    autozero. The display goes to 
    the next item in the CAL setup 
    menu.
    8) Either press  ZERO twice to 
    exit setup and store all 
    changes, or continue to 
    another setup menu item 
    using the  USER Key.
    &$/
    Auto Zero Maintenance Setup
    Cal Switch
    LO BA
    CAL
    Remove
    Hex Seal
    ScrewPress Switch
    behind the hole
    with a small
    Screwdriver
    VHWYS
    6WDQG
    $YWR
    2II
    6WDQG
    TARENext Cal Setup Menu Item
    ZERO
    Store and return to 
    weight display
    &D/
    ZERO
    6WRUH
    Calibrate Menu
    USER
    F
    USER
    F
    TAREBlinking
    2QBlinking
    CAL
    USER
    F
    36     MSI-3460 Operator’s Manual
    Figure 5-10. Auto Zero Maintenance  
    						
    All MSI manuals Comments (0)