P a g e 29 © Adam Equipment Company 2010 12.1.1 F1 -CALIBRATION To enter the calibration parameter, press the [Tare] key when “F1” “CAL” is displayed. The scale should be calibrated using a mass of appr oximately 2/3Maximum to Maximum. If the scale has been calibrated previous ly the value will be stored. Details of calibration are given is section 8. 12.1.2 F2 –PIN To enter this parameter, press the [Tare] key when “F2” “Pin” is shown. The “Weight” display will show “Pin 1” Enter the new password. The “Unit Weight” display will show dashes. Press [Tare] . The “Weight” display will change to “Pin 2 “, Enter the password again and press [Tare]. The display will show “donE” indicating the new password has been accepted and will return to the menu. “F3” “Unit” is shown. Record the new password in a secured place. Press [Zero] to return to weighing. 12.1.3 F3 –Unit To enter this parameter, press the [Tare] key when “F3” “unit” is shown. This parameter allows the user to enable or disable the weighing units available in his scales. The display will show the first available weighing unit and if it is On or Off. Press the [U.Wt./Units] key to change the value. Press [Tare] to accept the displayed setting and go to the next uni t. Press the [Zero] key to return to the menus. Press the [Zero] key again to return to normal weighing.
P a g e 30 © Adam Equipment Company 2010 12.2 USING “2006” TO ENTER THE SERVICE PARAMETERS Before setting the parameters for units, decimal po int, capacity and increment the user should make sure they understand how the scale should be configured. The scale is limited to 30,000 scale divisions. To dete rmine the number of scale divisions divide the capacity by the desired increment. For example a 500kg scale reading to 0.01kg would have 50,000 divisions, a value not all owed. However a 500kg scale with interval of 0.02kg is 25,000 divisions, an acc eptable number. The order of the settings will request the defaul t weighing unit first. This is to assign a weighing unit to the first calibration, re gardless of what he user may select later. If the indicator has been configured for use in the USA the choice can be grams, kilograms or pounds. For the rest of the wor ld it is kilograms or grams only. This setting should match the type of masses you ha ve for calibration, imperial or metric. The next 2 parameters will set the decimal point po sition, capacity and increment. It will be necessary for the settings to be within all owable limits for a setting to be accepted. For example, it is only allowed that th e grams can be set to 45,000 or less. If you try to set the capacity to 50,000 gra ms it is not allowed, use kilograms for the default unit and set the capacity to 50kg. When setting the interval the options shown will only be those that allow the num ber of divisions to be 30,000 or less. For example setting capacity to 5000g and t he decimal point to 0.0, the increment of 1 (=0.1g) is not allowed. Only the in crement of 2, 5, 10 or 20 is allowed. PROCEDURE: Press the [Tare] key during the display counting when turned on. Th e display will ask for a code number, “Pin “ on the Weight Display. Enter the number 2006 when “Pin “ is displayed and then press [Tare]. The displays will show the first parameters, called “F1” “CAL”. To select another parameter press the [U.Wt./Units] key to advance through the parameters. Press the [Tare] key to enter a parameter. To exit a parameter, pre ss the [Zero] key. The Weight window will show the parameter number and the Unit Price and Count windows will show the word describing the function. When a parameter is entered by pressing the [Tare] key, the displays will guide you through the parameter selected and t he options available.
P a g e 31 © Adam Equipment Company 2010 The parameters available are: “F1” “Unit” Default Weighing Unit “F2” “dPP” Decimal Point Position “F3” “CAP” Select capacity and increment “F4” “CAL” To enter the Calibration “F5” “init” “ZEro” Initial Zero Range “F6” “rE” “ Zero” Re-Zero range “F7” “SCSivE” “ tArE” Successive Tare Enable “F8” “Ad” “CoUnt” Display the A/D counts “F9” “Pin” To set the password “F10” “Lvd” “mode” Low voltage detection “F11” “trACk” “ZEro” Zero track setting 12.2.1 F1 –DEFAULT WEIGHING UNIT To set the weighing unit that will be used initiall y for calibration and setting the calibration. This weighing unit will always be avai lable to the user. To enter this parameter, press the [Tare] key when “F1” is shown. The display will show the current setting. Press the [U.Wt./Units] key to change the value. Select from the units displayed, typically kilograms (kg), grams (g) o r pounds (lb) for USA sales and kilogram or grams for the rest of the wor ld. Press [Tare] to accept the displayed value. 12.2.2 F2 –DECIMAL POINT POSITION To set the value for the decimal point location. T he options are 0, 0.0, 0.00, 0.000, 0.0000 To enter this parameter, press the [Tare] key when “F2” is shown. The display will show the current setting.
P a g e 32 © Adam Equipment Company 2010 Press the [U.Wt./Units] key to change the value. Select from 0, 0.0, 0.00, 0.000, 0.0000 Press [Tare] to accept the displayed value. 12.2.3 F3 – CAPACITY To enter this parameter, press the [Tare] key when “F3” is shown. The display will show the current capacity. Enter the numeric values using the keypad. The s cale will check the number of divisions n = maximum/increment is less than 30,000 divisions. Press [Tare] to accept the displayed value. NOTE: The scale use load cells suitable for the capacit y selected. The display then will go to the increment setting. To set the value for the increment of the scale. F or example 100kg x 0.01kg the increment is 10grams. but the last digi t increments by 1. The display will show the current increment value. Press the [U.Wt./Units] key to change the value. The indicator will determine what the minimum and maximum increment is possible for the capacity and decimal point you have set. Press [Tare] to accept the displayed value. 12.2.4 F4 -CALIBRATION To enter the calibration parameter, press the [Tare] key when “F4” is displayed. The scale will be calibrated using 2 ma sses of approximately 1/3Maximum and Maximum. If the scale has been cali brated once the values will be stored. If this is the first time t he scale is calibrated the user must enter the values for the calibration mass es. The display will instruct you to remove any weight from the scale, “UnloAd” . Press [Tare]. The display will tell you to add the first weight to the scale: “LoAd 1” “ 40” “kiloS” Add the weight shown, wait for stability then press the [Tare] key.
P a g e 33 © Adam Equipment Company 2010 The display will tell you to add the second weigh t to the scale: “LoAd 2 “ 100” “kiloS” Remove the weight. The display will show “SPAn” “PASS” if the calibration is OK. Or it will show “SPAn” “FAIL” if calibration could not be completed correctly. A fter calibration the scale will return to normal operati on. If you wish to set any of the following parameters enter the Service m enu again and skip the first settings by pressing the [U.Wt./Unit] key. 12.2.5 F5 –INITIAL ZERO RANGE To enter this parameter, press the [Tare] key when “F5” is shown. The display will show the current initial zero rang e. Press the [U.Wt./Units] key to change the value and press [Tare] to accept the value. Press [Zero] to return to weighing. 12.2.6 F6 -RE-ZERO RANGE To enter this parameter, press the [Tare] key when “F6” is shown. The display will show the current re-zero range. Press the [U.Wt./Units] key to change the value. Press [Tare] to accept the value. Press [Zero] to return to weighing. 12.2.7 F7 -SUCCESSIVE TARE To enter this parameter, press the [Tare] key when “F7” is shown. The display will show if the successive tare is on or off. Press the [U.Wt./Units] key to change the value. Press [Tare] to accept the displayed value. Press [Zero] to return to weighing.
P a g e 34 © Adam Equipment Company 2010 12.2.8 F8 –ADC COUNTS To enter this parameter, press the [Tare] key when “F8” is shown. This parameter allows you to view the A/D counts fr om the internal A/D converter. This can be an aid to service. Press the [Tare] key to return to the PARAMETER menu. Press the [Zero] key to return to weighing. Typical value at zero is 30,000-90,000 (approx.) Ty pical value at full capacity is 500,000 (approx.) 12.2.9 F9 –PIN This parameter allows setting of password to enter the user parameters, see section 12.1. The default value is “0000”. To enter this parameter, press the [Tare] key when “F9” “Pin” i s shown. The “Weight” display will show “Pin 1” Enter the new password. The “Unit Weight” display will show dashes. Press [Tare] . The “Weight” display will change to “Pin 2 “, Enter the password again and press [Tare]. The display will show “donE” indicating the new password has been accepted and will return to the menu. “F9” “Pin” is shown. Record the new password in a secured place. Press [Zero] to return to weighing. 12.2.10 F10 –LOW VOLTAGE DETECTION This parameter allows detection of low voltage when the battery wears down. To enter this parameter, press the [Tare] key when “F10” is shown. The display will show if the LVD Mode is set to “on” or “oFF” . Press the [U.Wt./Units] key to change the value. Press [Tare] to accept the displayed value. Press [Zero] to return to weighing.
P a g e 35 © Adam Equipment Company 2010 12.2.11 F11 –ZERO TRACK SETTING This parameter allows the amount of zero track corr ection. To enter this parameter, press the [Tare] key when “F11” is shown. The display will show the current setting, i.e. “ 0 .5d “ resenting one-half a division. Other possible settings are 1d, 2d, 4 d and Off. Press the [U.Wt./Units] key to change the value. Press [Tare] to accept the displayed value. Press [Zero] to return to weighing.
P a g e 36 © Adam Equipment Company 2010 13.0 REPLACEMENT PARTS AND ACCESSORIES If you need to order any spare parts and accessories, contact your supplier or Adam Equipment. A partial list of such items is as follo ws- Main Power cord Power supply (GC-a) Replacement Battery In use cover Printer
P a g e 38 © Adam Equipment Company 2010 14.0 SERVICE INFORMA\bION This ma\f\bal covers the details of operatio\f. If yo\b have a problem with the scale that is \fot directly addressed by this ma\f\bal the\f co\ftact yo\br s\bpplier for assista\fce. I\f order to provide f\brther assista\fce, the s\bpplier will \feed the followi\fg i\fformatio\f which sho\bld be kept ready: A. Details of your company -Name of yo\br compa\fy: -Co\ftact perso\f’s \fame: -Co\ftact telepho\fe, e-mail, fax or a\fy other methods: B. Details of the unit purchased (This part of i\fformatio\f sho\bld always be available for a\fy f\bt\bre correspo\fde\fce. We s\bggest yo\b to fill i\f this form as soo\f as the \b\fit is received a\fd keep a pri\ft-o\bt i\f yo\br record for ready refere\fce.) Model \fame of the scale: GC _____ Serial \f\bmber of the \b\fit: Software revisio\f \f\bmber (Displayed whe\f power is first t\br\fed o\f): Date of P\brchase: Name of the s\bpplier a\fd place: C. Brief description of the problem I\fcl\bde a\fy rece\ft history of the \b\fit. For example: -Has it bee\f worki\fg si\fce it’s delivered -Has it bee\f i\f co\ftact with water -Damaged from a fire -Electrical Storms i\f the area -Dropped o\f the floor, etc.