Home > Adam Equipment > Scale > Adam Equipment CBD Bench Counting Scales CBDREVBAE2 User Manual

Adam Equipment CBD Bench Counting Scales CBDREVBAE2 User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Adam Equipment CBD Bench Counting Scales CBDREVBAE2 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 35 Adam Equipment manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							 
    © Adam Equipment Company 2011  Page  
    29 
     
    Sample of output:   
    Using M+ function Using PRINT function 
    DATE   01/27/2011  
    TIME   12:38:49 
    LOCAL SCALE 
    SID: ABS123  
    UID: 123ABC  
    NAME:Text  
      4    S. No.  
     1.234 kg TARE 
    12.456 kg NET 
    1.1234  g U.W.  
        11 PCS  
    TOTAL 
    - - - - - - - - - - - -  
     49.824 kg TW 
         44 TPC 
          4 No.    DATE   01/27/2011  
    TIME   12:39:58 
    LOCAL SCALE 
    SID: ABS123  
    UID: 123ABC   
    NAME:Text  
     1.234 kg TARE 
    12.456 kg NET 
    1.1234  g U.W.  
        11 PCS    
     
    9.1  INPUT COMMANDS FORMAT 
     
    The scale can be controlled with the following comm
    ands.  
     
    Input Commands:  
    ·  The  scale  has  a  number  of  commands  to  either  cause   an  action  or  to 
    enter  data  into  memory.      The  commands  are  all  uppe r  case  and  are 
    summarised below. 
     
    ·   All  commands  are  terminated  by  a  carriage  return  ( Enter  button  on  PC 
    keyboard) with the line feed optional.   
     
    ·   If  an  illegal  command  is  received  or  a  command  can not  be  carried  out 
    then  send  the  command  back  with  the  addition  of  ER  in front of it.  For 
    example if the command is NN send back ER N N.  
    						
    							 
    © Adam Equipment Company 2011  Page  
    30 
     
    Basic Commands:  
    PLUxx  Select PLU from scale memory to be used 
    T  Tare current weight value 
    T123.456 preset tare value is 123.456 
    Z Zero the display 
    P Print using selected format 
    M+ Store current results into memory and print 
    MR Recall memory values to scale display 
    MC  Clear memory 
    U123.456 Store unit weight of 123.456  
    (grams if in kilograms or pounds if currently in po unds)
     
    S123  Enter sample size of 123 parts.   
    Same as pressing [Smpl] key.
     
    SL Select local scale to be used 
    SR Select remote scale to be used 
     
    Immediate Printing Commands:     
    Command  Output from scale 
    \L  Scale: Local or Remote  
    \I  ID number same as PUID below 
    \S  Scale number same as PSID below 
    \N  Net weight 
    \G  Gross Weight 
    \T  Tare weight 
    \U  Unit weight 
    \P  Count 
    \C  Total Count 
    \W  Total Weight 
    \M  Number of items stored in memory  
    \B  A blank line printed 
    9.2  STORING DATA VIA RS232    
    To store data the commands are:  
    SUIDxxxxxx   Store user ID data 
    SSIDxxxxxx   Store scale ID data 
    SPLUxx,xxxxxxxxxxxx     Store text data for PLUxx  
     
        
    When  PLU  text  data  is  stored  the  Scale  stores  curre nt  unit  weight  and 
    current tare value along with the PLU number.  
       
    For  the  SPLU  command  the  data  is:  PLU  number  (2  cha racters),  (Comma) 
    description (max 12 characters).  
    If the fields are less than the maximum all charact ers need not be used.   
    						
    							 
    © Adam Equipment Company 2011  Page  
    31 
     
      
    9.3  PLU ENTRY USING RS-232 INTERFACE    
    This method enables the user to send the data from  a PC program as well as 
    from  the  keypad.  The  most  common  PLUs  can  be  stored   and  recalled  from 
    the  scale  memory.  Other  PLU  data  can  be  stored  on  a   PC,  then  the 
    Description  (text  data),  Unit  Weight  and  Tare  Weigh t  data  can  be  sent  from 
    the  PC  for  the  PLUxx  (where  xx  can  be  any  desired  n umber  up  to  99).  This 
    can then be used or over-written during operation.   
    OPERATION:   
    ·  Send Tare Weight data to set any tare value to be  stored with PLU.   
    i.e. “ T 1.45 ” . If no tare is needed then you may send “ T 0” to delete 
    any present tare data. 
     
    ·  Send the Unit Weight to be stored with PLU. i.e. “ U0.015”  
     
    ·  Send PLU text data to be stored with current Tare  and U/W values. 
    i.e. “ SPLU27, M4 Nut Brass ”  
     
    Any  information  stored  for  the  Tare  Weight,  Descrip tion  and  the  Unit  weight 
    will be stored in the scale and can be recalled for  use in the future. 
      
      
    						
    							 
    © Adam Equipment Company 2011  Page  
    32 
     
       
    10.0 CALIBRATION 
     
    ACTION  DISPLAYS 
    Press the  [Tare/Zero]  key during the self-
    test at power on. The scale will ask to  
    enter the password. 
     
      
    “ Pi n” 
    Default Password is 0000.  
    Enter “0” four times. 
    Password can be changed in technical 
    parameters.  
    Press  [Print] . 
        
    “ Pi n”     “ - - - -” 
    Select the scale to be set up by using the  
    [Local/Remote]
     key. 
    Press the  [Print] key to enter the 
    technical section.  
     
    “tECH”  “   LocAL”   “      “    
    “tECH”  “rEmo tE”   “      “   
     
    Use the  [U.Wt.] to select the weighing unit 
    to be used for setting  the calibration for  
    the scale. The arrow in the “ Weight” 
    window will indicate the unit selected. 
    Press the  [Print] key to continue. 
     
      “tECH”  “   Uni t”   “      “   
       
     
    You will enter the first parameter - 
    Calibration. Display will ask to unload any 
    weight on the platform.  
    Press the  [Print] key to continue.  “nO L o Ad”     
       
    Load the calibration weight onto the scale  
      
    If the mass is different than the value  
    shown press the 
    [CE] key then enter the 
    calibration weight to be used .    
    Then press  [Print]. 
     
    “LoAd   ”    “ 0010”    “ KiLoS”            
    						
    							 
    © Adam Equipment Company 2011  Page  
    33 
     
     
    The Calibration is complete, the scale  
    will run the self-test during which the 
    weight should be removed.   
    If calibration is not possible an error 
    message will be shown  
    “- - - -”    “  - - - -”    “ - - - - ”  
    “SPAn”    “ PASS”    “        ”    
       
    “SPAn”    “ FAiL”    “ LoW   ”  
    or  
    “HiGH 
      
    If The Calibration is being done while in the TECH  parameters (section 12)  the linearity is 
    set at the same time as calibration by performing t he calibration in 2 steps,   
     
    Display will ask to unload any weight on  
    the platform.  
    Press the  [Print] key to continue.  “nO L o Ad”     
       
    Load the first calibration weight onto the  
    scale   
    If the mass is different than the value  
    shown press the 
    [CE] key then enter 
    the calibration weight to be used .    
    Then press  [Print] to continue or press 
    [Zero] to complete with only one mass  
    being used. 
     
    “LoAd  1 ”    “ 0010”    “ KiLoS”           
    Load the second  calibration weight  
    onto the scale   
    If the mass is different than the value 
    shown press the [CE] key then enter 
    the calibration weight to be used .  
    Then press [Print].   “LoAd  2”    “ 0030”    “ KiLoS”           
    The Calibration is complete, the scale  
    will run the self-test during which the 
    weight should be removed.   
    “- - - -”    “  - - - -”    “ - - - - ”  
    “SPAn”    “ PASS”    “        ”    
    If calibration is not possible an 
    error message will be shown  
    “SPAn”    “ FAiL”    “ LoW   ”  
    or  
    “HiGH  
    						
    							 
    © Adam Equipment Company 2011  Page  
    34 
     
     
    11.0  ERROR CODES 
     
    During  the  initial  power-on  testing  or  during  the  o peration,  the  scale  may 
    show  an  error  message.    The  meaning  of  the  error  me ssages  are  described 
    below. 
     
    If  an  error  message  is  shown,  repeat  the  procedure  that  caused  the 
    message,  such  as  turning  the  balance  on,  calibratio n  or  any  other  functions.  
    If  the  error  message  is  still  shown  then  contact  yo ur  dealer  for  further 
    support.   
      
    ERROR CODE  DESCRIPTION  POSSIBLE CAUSES 
    Err  4 Initial Zero is greater than the  
    permissible value (typically 4% 
    of maximum capacity) when 
    power is turned on or when the  
    [Zero]  key is pressed,  Weight on the pan when turning  
    the scale on.  
    Excessive weight on the pan 
    when zeroing the scale.  
    Improper calibration of the scale.  
    Damaged load cell.  
    Damaged Electronics.  
    Err 5 Keyboard error.    Improper operation of the scale. 
     
    Err 6 A/D count is not correct when  
    turning the scale on. Platform is not installed.  
    Load cell may be damaged.  
    Electronics may be damaged.  
    FAIL HiGH or  
    FAIL LoW Calibration error  Improper calibration.    
    If  the  problem  persists  contact 
    your  dealer  or  Adam  Equipment 
    for assistance.  
     
      
    						
    							 
    © Adam Equipment Company 2011  Page  
    35 
     
     
    12.0 TECHNICAL PARAMETERS 
     
    The  technical  parameters  accessed  via  the  “ tECH”  prompt  at  the  end  of  user 
    parameters  and  are  password  controlled  to  prevent  u nauthorised  access.  
    These  parameters  set  the  metrology  for  the  scales.      Each  scale  is  set 
    independently.  The  parameters  will  set  capacity,  di vision,  decimal  point 
    position,  initial  zero  range,  auto  and  manual  zero  range  as  well  as  factory 
    calibration.   
     
    ACTION  DISPLAYS 
    When “ tEch” is displayed, press the 
    [Print]  key. The scale will ask you 
    to enter the password. 
     
      
    “ Pi n” 
    Default Password is 0000.  
    Enter “0” four times. 
    2006 will override any other user  
    password. Press  [Print]. 
        
    “ Pi n”     “ ----” 
    Select the scale to be set up by  
    using the 
    [Local/Rem]  key. 
     
    Press the  [Print] to enter the 
    technical section for that scale.  
     
    “t ECH”  “   Lo c A L”   “      “    
    “t ECH”  “r E mo t E”   “      “   
     
    Use the  [U.Wt./Units]  to select the 
    weighing unit to be used for setting  
    up the parameters for the scale.  
    The arrow in the “ Weight” window 
    will indicate the unit selected. 
    Press the  [Print] key to continue. 
     
      “t ECH”  “   Uni t”   “      “   
       
     
    Press the  [U.Wt./Units]  key to 
    scroll through the menus for the  
    chosen scale. User   [Print] to 
    select and set parameters and the  
    [Tare/Zero ] to escape as without 
    changing data as with the user  
    parameter section.    “Cn t”  
        
    Both  scales  have  the  same  menus 
    except that the capacity of the local 
    scale  is  fixed  and  cannot  be 
    entered. 
      
    						
    							 
    © Adam Equipment Company 2011  Page  
    36 
     
     
    Press the 
    [U.Wt/Unit]  to go to next 
    menu. Press  [Print] to enter. 
    Use  [U.Wt/Unit]  to select the 
    decimal point position and press  
    [Print].  
    To change the remote scale cap 
    press CE to clear and then enter  
    new value followed by  [Print]. 
          “
    CAP ”.  Scale capacity setting.  
     
    “dESC” “  0.00”      
    “LOAd” “0060”    
    Press  [U.Wt/Unit]  to go to next 
    menu. Press  [Print] to enter. 
    Use  [U.Wt/Unit]  to select the 
    division which you would like the  
    display to increment in, followed by  
    [Print].  
      “inC    5” 
    Press the 
    [U.Wt/Unit]  to go to next 
    menu, use  [Print] key to enter. 
    Use  [U.Wt/Unit]  to increment the 
    values then  [Print].  “AZt”
         Autozero tracking range.  
        
    Select from  0.5d
    0.5d 0.5d
    0.5d, 1d, 2d, 4,d  
     
    Press the  [U.Wt/Unit]  to go to next 
    menu, use  [Print] key to enter. 
    Use  [U.Wt/Unit]  to increment the 
    values then  [Print].  “0 Auto”
            Zero auto range at 
    power on.    
    Select from  0%, to  10%  
      
     
     
    Press the  [U.Wt/Unit]  to go to next 
    menu, use  [Print] key to enter. 
    Use  [U.Wt/Unit]  to increment the 
    values then  [Print].  “0 
    manl ”  Zero manual range. 
      
    Select from  0%, to  10%  
      
     
     
    Press the  [U.Wt/Unit]  to go to next 
    menu, use  [Print] key to enter. 
    Enter the new PIN number and  
    press  [Print]. 
    Confirm the new PIN number and 
    press  [Print].  “
    Pin ” Password number for “ tEch”. 
      
    “Pin1”   
    “Pin2” 
    “donE”   
    Press  [U.Wt./Units]  to continue 
    use  [Print]  key to enter. 
    Escape the parameters with the  
    [Zero/Tare]   “
    CAL ”  Enter calibration section 
    where the local or remote scale is  
    selected.  See section 10. 
                
    						
    							 
    © Adam Equipment Company 2011  Page  
    37 
      
    						
    							©
    
     Adam Equipment Company 2011  Page  
    38 
    13.0 REPLACEMENT PARTS AND ACCESSORIES 
    If you need to order any spare parts and accessories, contact your supplier or Adam 
    Equipment. A partial list of such items is as follows- 
    ·Power Supply Module
    · Main Power cord
    · Replacement Battery
    · Stainless Steel Pan ·
    In use cover
    · Printer, etc.
    14.0 SERVICE INFORMATION 
    This  manual  covers  the  details  of  operation.  If  you  have  a  problem  with  the 
    scale that is not directly addressed by this manual then contact your supplier 
    for  assistance.  In  order  to  provide  further  assistance,  the  supplier  will  need 
    the following information which should be kept ready: 
    A. Details of your company 
     -Name of your company: 
     -Contact person’s name: 
     -Contact telephone, e-mail,   fax or any other methods: 
    B.  Details of the unit purchased
     
       (This  part  of  information  should  always  be  available  for  any  future 
    correspondence.  We  suggest  you  to  fill  in  this  form  as  soon  as  the  unit  is 
    received and keep a print-out in your record for ready reference.) 
    Model name of the scale: CBD____ 
    Serial number of the unit: 
    Software revision number  
    (Displayed when power is first turned on): 
    Date of Purchase: 
    Name of the supplier and place: 
    C
    
    .   Brief description of the problem
     
     Include any recent history of the unit. For example:    -Has it been working since it’s delivered 
      -Has it been in contact with water  -Damaged from a fire 
    -Electrical Storms in the area 
    -Dropped on the floor, etc.   
    						
    All Adam Equipment manuals Comments (0)