Page 1

 
© Adam Equipment Company 2010                    
  
Adam Equipment 
   
 
 
 
 
 
 
                        PGW SERIES 
                                                       (P.N. 3016610939, Revision A, July 2010) 
  
 
 
 
 
 
 
                                                 So ftware rev.: 4.22 & above 
       

Page 2

 
© Adam Equipment Company 2010                    
    

Page 3

 
© Adam Equipment Company 2010                        1              
TABLE OF CONTENTS 
  
1.0  KNOW YOUR BALANCE ............................. ................................................................ 3 
2.0  SPECIFICATIONS ................................ ....................................................................... 4 
3.0  UNPACKING THE BALANCE ......................... ............................................................. 6 
4.0  LOCATING THE BALANCE...

Page 4

 
© Adam Equipment Company 2010                        2               

Page 5

 
© Adam Equipment Company 2010                        3              
1.0  KNOW YOUR BALANCE 
 
Thank you for selecting the PGW Balance.    
This  Instruction  Manual  will  familiarise  you  with  t he  installation,  accessories, 
trouble-shooting,  after  sales  service  information,  general  maintenance  of  the 
balance, etc. and will guide you through the variou s applications. 
 
Please read this Manual thoroughly before starting  the operation. If you need 
any clarifications, feel free to...

Page 6

 © Adam Equipment Company 2010                                                                             
4 
   2.0  SPECIFICATIONS   
Model*  PGW 
153e PGW 
253e  PGW 
453e  PGW 
603e  PGW 
753e  PGW 
1502e  PGW 
2502e  PGW 
3502e  PGW 
4502e  PGW 
6002e Maximum  
capacity  150g  250g  450g  600g  750g  1500g  2500g  3500g  4500g  60
00g Readability 
0.001g  0.01g Tare range  Full Repeatability  
(S.D.) 0.001g 
0.01g Linearity (±)  0.002g  0.02g Units of  
measure grams, milligrams (for 0.001g units...

Page 7

 © Adam Equipment Company 2010                                                                             
5 
  
Housing 
Die cast aluminium housing (glass weighing chamber  for models with 1mg readability) Pan size 
125×145mm 140 x 
140mm  192×192mm Overall  
dimensions   
(w x d x h)   
251 × 358 × 104 mm Net weight  4 kg Applications  Weighing, Check weighing, Percentage w eighing, Parts counting, 
Animal / dynamic weighing, Net / total and Density  determination 
 
                *
Models are...

Page 8

 
© Adam Equipment Company 2010                        6 
                       
     
3.0  UNPACKING THE BALANCE 
 
Remove the balance from the packing by carefully li
fting it out of the box.   
Inside the box you will find everything needed to s tart using the balance- 
 
 
 
  AC adapter 
   Stainless Steel Top Pan 
   Plastic sub-pan 
   Draught shield (for mg models only) 
  This User Manual   
4.0  LOCATING THE BALANCE 
  
       
 
   
 
 
 
 
 
   The  balance  should  not  be  placed  in...

Page 9

 
© Adam Equipment Company 2010                        7 
   
5.0  SETTING UP THE BALANCE 
5.1  ASSEMBLING THE BALANCE 
 
  Locate the balance on a solid surface, free from v ibration  
  For PGWi models, tilt the balance 
on it’s side and remove the two  
shipping screws and nuts before  
use. If you need to ship the 
balance, replace the screws but do 
not over-tighten which may damage 
the internal mechanisms 
 
 
  For  the  low  capacity  scales  (0.001g  except  the  PGW   753)  remove  the...

Page 10

 
© Adam Equipment Company 2010                        8 
5.2  LEVELLING THE BALANCE 
   
 After  placing  the  balance  in  a  suitable  place,  
level it by using the spirit level on the rear of t
he 
balance.  To  level  the  balance  turn  the  two 
adjustable  feet  at  the  rear  of  the  balance  until 
the bubble in the spirit level is centred.  
5.3  WARM-UP TIME 
 
 
 
   
 
 
Attach  the  power  supply  cable  to  the  connector  
on  the  rear  of  the  balance.  Plug  the  power 
supply...