Page 21

 
© Adam Equipment Company 2010                        19 
 
  Next  the  beeper  setting  is  displayed.  Press  [Up]  or [Down]   to  scroll 
through the options –    
“BUZZER OFF”   (Beeper set to off at all times) 
“BUZZER In”   (Beeper  will  sound  when  the  weight  on  the  pan  is 
                              
stable and within the limit)     
“BUZZER Out”    (Beeper  will  sound  when  the  weight  o n  the  pan  is 
stable and outside the set limits) 
 
   Confirm the beeper setting by...

Page 22

 
© Adam Equipment Company 2010                        20 
MANUAL MODE 
  
When the balance is in the MANUAL mode –
  
 
If  [Enter]  is pressed when “rUN” is selected, balance will di splay “START ?” 
 
Place the item on the pan and press [Enter] again 
 
After  the  configured  Delay  and  test  time  are  comput ed  by  the  balance,  the 
result  will  be  displayed  by  showing  “FINISHED  xx.xx  x”  after  a  pre-set  time 
Delay and Sampling period (see the section 9.4.4.1 Animal Weighing Setup...

Page 23

 
© Adam Equipment Company 2010                        21 
 
  Press  [Enter]  to  select  “SEtUP”  to  set  up  the  balance  for  animal  
weighing 
 
  Use  the  [Up]  or [Down]  key  to  scroll  through  the  options  for  setting  up 
the mode. The display will show “MODE AUTO” or “MOD E MANUAL”. 
 
  Select “MANUAL” or “AUTO” 
   If “AUTO” is selected- 
  Threshold
 (For AUTO mode only) 
 
  The  display  will  next  show  “THRESHOLD  XX”  where  XX   is  the 
minimum weight of the item...

Page 24

 
© Adam Equipment Company 2010                        22 
9.4.5  Net/Total     
This  function  allows  the  user  to  see  the  total  of  a ll the  net  weights,  the  value 
of  which  is  displayed  on  smaller  digits  above  the  m ain  display.  The  current 
net  weight  is  shown  on  the  main  display.  The  [Tare]  key  functions  the  same 
way as in the normal weighing.    
Steps:   
  Press  [Mode]  and  then  [Up]  or [Down]   to  select  Net/Total.    When 
“NET/TOT” is displayed press...

Page 25

 
© Adam Equipment Company 2010                        23 
immerse  the  sample  in  the  chosen  liquid.  The  densit y  of  the  liquid  must  be 
known or determined from a look-up table. 
 
For  water  and  ethanol  the  density  will  be  calculate d  based  on  the 
temperatures entered using the numeric entry method  (see section 7.1) 
 
Steps:  
  Press  [Mode] and then  [Up] and  [Down]  keys to select density 
 
  Press  [Enter] to enter Density mode. 
 
  Press  the  [Up]  and  [Down]...

Page 26

 
© Adam Equipment Company 2010                        24 
DENSITY OF LIQUID     
  
When  finding  the  density  of  a  liquid,  it  is  necessa ry  to  weigh  a  sample  of 
known volume in air and then in the liquid. The vol ume of the sample must be 
entered by the user. The last known volume is store d for use at any time. 
  
If  using  the  density  determination  kit,  the  volume  of  the  plumb  is  marked  on 
its support, i.e. 10.123    
Steps:  
  Press  [Mode] and then  [Up] and  [Down]...

Page 27

 
© Adam Equipment Company 2010                        25 
10.0  CALIBRATION 
 
The  PGW  XXXe  series  can  only  be  calibrated  with  an  external  mass.    The 
PGW  XXXi  series  can  be  calibrated  with  either  an  in ternal  mass  (default 
method)  or  an external mass  (if  setup by  the  user  t o do this).     See  the  User 
Parameters section. 
 
10.1  MANUAL CALIBRATION 
 
Pressing the  [Cal] key will start calibration. Calibration can also be  initiated by 
a  change  in  internal...

Page 28

 
© Adam Equipment Company 2010                        26 
 
10.2  AUTOMATIC CALIBRATION 
 
The  balance  will  have  the  ability  to  calibrate  (or  ask  for  calibration)  when  the 
balance  has  automatic  calibration  enabled  and  the  c onditions  of  the 
automatic calibration have been meet.  
 
Conditions that will cause an automatic calibration  are:  
1.  Internal temperature change greater than a prese t amount, typically 2ºC 
for Precision balances.  
2.  Time since last calibration...

Page 29

 
© Adam Equipment Company 2010                        27 
   
11.0  RS-232 INTERFACE 
 
The balances have the ability to send or receive da ta over the serial interface.  
The weighing data can be sent over the interface ei ther automatically or when 
the user presses the  [Print] key. 
 
The user has control over what data is to be printe d.  
  
The following gives a description of the RS-232 int erface. 
 
HARDWARE  
The RS-232 interface is a simple 3 wire connection.  The input and output...

Page 30

 
© Adam Equipment Company 2010                        28 
Unit weight  Unit wt: xxx.xxx g 
Count  Count: xxxx pcs 
Reference weight  Ref. wt: xxx.xxx g 
Percent  Percent: xx.xxx % 
Checkweigh lower limit  Low: xxx.xxx g 
Checkweigh upper limit  High: xxx.xxx g 
A blank line printed   only.  
Any  of  these  can  be  printed  on  any  of  the  15  lines  available.    Not  all  items 
need to be used and any one can be used more than o nce (see section 13.3). 
 
The  data  for  each  form  will  be...