Home > Adam Equipment > Scale > Adam Equipment SHS Crane Scales SHSSHSaREVB User Manual

Adam Equipment SHS Crane Scales SHSSHSaREVB User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Adam Equipment SHS Crane Scales SHSSHSaREVB User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 35 Adam Equipment manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							© Adam Equipment Company 2012 
    Adam Eq\bipme\ft 
    SHS/SHSa SER\bES 
    (P.N. 7.00.6.6.0202, Revisio\f B November 2012) 
    Software versio\f S1.1/SA1.1   
    						
    							   1          © Adam Equipment Company 2012 
      
    1.0  CONTENTS 
      
    1.0
     CONTENTS ........................................... ................................................1 
    2.1 FEATURES ........................................................................................2 
    2.2 TECHNICAL SPECIFICATIONS........................... ..............................3 
    2.3 SAFETY NOTICE ...............................................................................4 
    2.4 KEY AND PANEL DESCRIPTION .......................... ............................5 
    2.5 DISPLAY INDICATIONS WITH LED...................................................6 
    3.0 OPERATION .........................................................................................6 
    3.1 SAFETY INSTRUCTIONS ................................ ..................................6 
    3.2 BATTERY OPERATION .....................................................................7 
    3.3 BASIC OPERATIONS ........................................................................8 
    3.4 REMOTE CONTROL UNIT ................................ ................................9 
    3.5 USER PARAMETERS ........................................................................9 
    4.0 CALIBRATION..................................................................................10 
    5.0 PARTS LIST ........................................................................................11 
    WARRANTY INFORMATION ........................................ ..............................13 
      
    						
    							   2          © Adam Equipment Company 2012 
    2.0  INTRODUCTION  
    2.1  FEATURES   
    The SHS/SHSa Series are a range of heavy-duty lower  capacity crane scales  
    The main features include:     ·   Easy Operation & Rugged construction 
    ·   15mm bright red LED display enabling a good viewing  angle. 
    ·   360 degree rotary safety hook and support ring desi gn 
    ·   Overload safety up to 150% of model capacity 
    ·   Selectable weighing units include kg, Lbs and Newto ns (Newtons not 
    available for “a” models) 
    ·   Standard and Peak Hold feature 
    ·   Internal lithium battery fitted as standard 
    ·   Alternative power from AA or NiMH  batteries 
    ·   Auto power off after 30 minutes of non use 
    ·   8.4Vdc 800ma battery charger 
    ·   Strong ABS plastic housing  
    ·   Infrared remote control 
     
      
    						
    							   3          © Adam Equipment Company 2012 
    2.2 TECHNICAL SPECIFICATIONS 
       
    Model No  Capacity  Division  Model No  Capacity  Divisi on 
    SHS 50 50kg  0.01kg SHS 100a 100lb/50kg  0.02lb/0.01kg 
    SHS 150 150kg  0.02kg SHS 300a 300lb/150kg  0.05lb/0.02kg 
    SHS 300 300kg  0.05kg SHS 600a 600lb/300kg  0.1lb/0.05kg 
        
       
    						
    							   4          © Adam Equipment Company 2012 
    2.3  SAFETY NOTICE 
      
    The limit stated below is the Ultimate Overload Lim its of the scales. 
    Adam  Equipment  Co.  Ltd.  Its  directors,  owners  and  a ffiliated  companies  and 
    organisations cannot be held liable for any resulti ng damage or injury sustained as a 
    result  of  the  machines  being  used  in  excess  of  thei r  stated  capacities  or  ultimate 
    overload limits.      
    Model No.  SHS 50/SHS 100a  SHS 150/SHS 300a  SHS 300/ SHS 600a 
    Max  
    capacity   
    50kg/100lb  150kg/300lb  300kg/600lb 
    Overload  
    capacity 
    100kg/200lb  300kg/600lb  450kg/900lb 
                     
        
         
    NOTICE:  
    Do not use the scales if any parts are bent or loos e or show signs of damage or wear. 
    Refer to the user manual.    
    						
    							   5          © Adam Equipment Company 2012 
    2.4  KEY AND PANEL DESCRIPTION  
     
     
    [ON/OFF] The  power  switch  is  located  on  the  rear  housing.  Us e  this  key  to 
    turn the scale ON and OFF. 
    [Zero] Press the [Zero] key to return the reading to zero.  
    [Unit] 
    Press  the  [Unit]  key  to  change the  weighing  unit  being displayed 
    between  Kg,  Lbs  and  Newtons  (Newtons  not  available  for  “a” 
    models) 
    [Hold] 
    Use  this  key  to  activate  standard  Hold.  When  pressed  the  Hold 
    LED will light at the top left side of the display and the reading on 
    the  display  will  lock  and  flash.  Press  the [Hold] key  again  to 
    cancel the  [Hold] function. 
    *When  Peak  hold  is  selected  from  the  parameter  sect ion  the 
    press  of  the  Hold  key  will  activate  Peak  Hold.  The Hold  LED  will 
    flash  and  the  display  will  also  flash  and  lock  show ing  the 
    maximum weight registered of a single weighing process. 
    Press the [Hold] key again to cancel the  [Hold] function. 
     
       
    						
    							   6          © Adam Equipment Company 2012 
    2.5  DISPLAY INDICATIONS WITH LED 
    Kg lb N Weighing unit Indicator 
     Low battery Indicator 
    Hold Hold Indicator 
    Stab Stable  Indicator 
    3.0  OPERATION 
    3.1  SAFETY INSTRUCTIONS 
    Read this manual carefully before you operate the S HS Crane Scale. 
    ·   Unpack the Crane Scale carefully from the packaging .  
    ·   Attach it to your hoist or equivalent existing equi pment securely at the 
    ground level to the support ring in the upright pos ition. Make sure that 
    the equipment to which the scale is attached is in  proper condition. The 
    operator should be trained for its safe operation. 
    ·   Carry out regular visual inspection for any damage  or wear and tear on 
    parts. The split pins in particular at the hook and  eyebolts, as well as the 
    inner wall of the hook and all joints should be che cked regularly. If any 
    damage  or  wear  has  taken  place,  stop  using  the  scal e  and  replace  the 
    parts.   
    ·   Do all inspection and maintenance work when the sca le is not in use. 
    ·   Do  not  expose  the  scale  to  extreme  temperatures.  Th e  operating 
    temperature range is 0° C to 40° C. 
    ·   Secure  the  sample  to  be  weighed  in  the  hook  of  the  scale.  Ensure  that 
    the  sprung  Safety  Lever  returns  to  its  resting  posi tion  which  prevents 
    the sample from slipping out of the hook. 
    ·   Do not exceed the maximum capacity of the Scale . 
    ·   Clean and protect your scale after use.  
    						
    							   7          © Adam Equipment Company 2012 
    Overload Warning 
    Please  do  not  add  any  item  that  exceeds  the  maximum   capacity  of  the  scale. 
    When  the  display  shows  ”- - - - - ”  remove  the  item  to  avoid  damage  to  the  load 
    cell and scale. 
     
    3.2  BATTERY OPERATION 
     
    ·  The  lithium  battery  will  operate  continuously  for  2 0  hours  before 
    requiring  recharging.  The  scale  will  alert  you  that   the  battery  is  low  on 
    charge when the Indicator LED above the battery sym bol lights up. 
    ·   To charge the  battery, unclip the  battery  cover on  the rear of the scale 
    and  insert  the  connector  into  the  socket  as  marked.   Plug  the  charger 
    into  the  mains  power  supply  and  wait  until  the  LED  on  the  charger 
    shows green. The battery is now fully charged. 
    ·   Charging  time  for  a  completely  discharged  battery  i s  approximately  16 
    hours.  A  full  battery  pack  is  recommended  to  keep  t he  scale 
    continuously operational. 
    ·   Only use the charger as supplied with the unit. 
    ·   The  switch  inside  the  battery  cover  must  be  in  posi tion  A  to  make  use 
    of the lithium battery. 
    ·   If using AA or NiMH  batteries the switch must be i n position B 
    ·   The scale will automatically power down after 30 mi nutes of non use to 
    further protect the battery. 
    ·   To replace the lithium battery pack please contact  your dealer 
          
    						
    							   8          © Adam Equipment Company 2012 
    3.3 BASIC OPERATIONS 
    Power on and off
      
    Press  the  [ON/OFF]  button  to  turn  the  scale  on  or  off.  The  display  on  the  scale 
    will show the model capacity then countdown from 9  to 0 before displaying zero 
    weight.   
    Zero the scale
     
    Press the  [Zero] key  so that the display shows zero. 
      
    Change the weighing Unit
     
    Press the  [Unit] key to change the weighing unit 
      
    Standard Hold (setting 1)
     
    When  the  reading  may  not  be  perfectly  stable  or  you   cannot  view  a  reading 
    easily you can press the  [Hold] key to lock it. The reading will be locked and also  
    flash at the point you press the  [Hold] button enabling you to view it when you 
    can. The Hold LED will also light up and once you h ave noted the reading a press 
    of the  [Hold]  key will  return the scale to the normal weighing mode. 
      
    Peak Hold (setting 0)
     
    Peak Hold is the highest weight reading registered  by the scale when a weighing 
    takes  place.  Press  the  [Hold]  key  when  the  display  is  at  zero  and  the  Hold  LED 
    will  flash  which  denotes  Peak  Hold  has  been  activat ed.  When  weight/force  is 
    applied  to  the  scale  the  display  will  also  flash  an d  lock  the  highest  recorded 
    weight of that weighing process. This value and the  Hold LED will remain flashing 
    on  the  display  until  the  [Hold]  key  is  pressed  again  at  which  point  the  scale  will  
    return to normal weighing mode displaying the curre nt weight on the scale. 
    Press  [Hold]  again to perform another Peak Hold weighing at any  time.  
    						
    							   9          © Adam Equipment Company 2012 
     
    3.4  REMOTE
     CONTROL UNIT 
     
    The  remote  control  works  via  infra  red  and  can  be  u sed  to  operate  the  SHS  from 
    long distances. It operates using 2 x AAA batteries  and these can be exchanged by 
    unclipping a cover on the rear of the remote contro l unit. 
    3.5  USER
     PARAMETERS 
    Four  USER  settable  parameters  are  available  on  the  SHS,  to  access  these  power 
    up  the  scale  and  whilst  the  display  is  counting  dow n  press  and  then  release  the 
    [Unit]  button. The display will show Bee 0. Press the  [Hold] button to change the 
    setting, and the  [Zero] button to move to the next parameter.  
    Parameter setting table. 
    FUNCTION  DESCRIPTION  DEFAULT 
    BEE Sets buzzer  0: Disable    
    1: Enable   
    1 
    HOD Sets the Hold function 0: Peak  1: Standard   1 
    CHT 
    Sets weighing unit on power up  0: kg   
    1: Lb’s  
    2: Newtons(Newtons not available for “a”  
    models)
      0 
    YES Sets save: 
    0:  not saved  
    1:  save new settings 0   
    						
    All Adam Equipment manuals Comments (0)