Home > GE > Ultrasound System > GE Logiq 3 Instruction Manual

GE Logiq 3 Instruction Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 3 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Acoustic Output
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-65
    Direction 5122542-100 Rev. 2
    Probe Name:7S
    Table 1-51:  Transducer Mode 7S Operating Mode: B Mode
    Index LabelMI
    TISTIB
    TIC
    scannon-scannon-
    scan
    Aaprt1
    Global Maximum: Index Value 1.061.3
    IECFDAUnits
    pra pr.3 (MPa) 2.76
    PWo (mW)89.9
    min of [P
    α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
    zs z1(cm)
    zbp zbp(cm)
    zb zsp(cm) 2.52
    z  at max. I pi,α zsp(cm)
    deq(zb) deq(zsp) (cm)
    ƒawffc(MHz) 3.993.99
    Dim of A
    aprtX(cm)1.0
    Y(cm)
    0.7
    t
    dPD (µsec) 0.6
    prr PRF (Hz) 4192.9
    pr at max. Ipi   pr@PIImax (MPa) 3.35
    deq at max. Ipi deq@PIImax (cm)
    Focal Length FLX(cm)4.8
    FL
    Y(cm)8.0
    I
    pa, αat max. MI IPA . 3@MImax 
    (W/cm2) 344.35
    Frequency (MHz)
    T4T4
    Image Depth (cm) 14.0
    10.0
    Other Conditions
    ARM_Rev_2.book  Page 65  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-66LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Table 1-52:  Transducer Mode 7S
     Operating Mode: M Mode (Inc. B-Mode)
    Index LabelMI
    TISTIB
    TIC
    scannon-scannon-
    scan
    Aaprt1
    Global Maximum: Index Value 1.481.1
    IECFDAUnits
    pra pr.3 (MPa) 2.32
    PWo (mW)79.4
    min of [P
    α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
    zs z1(cm)
    zbp zbp(cm)
    zb zsp(cm) 2.33
    z  at max. I pi,α zsp(cm)
    deq(zb) deq(zsp) (cm)
    ƒawffc(MHz) 3.973.97
    Dim of A
    aprtX(cm)1
    Y(cm)
    0.7
    t
    dPD (µsec) 0.61
    prr PRF (Hz) 4192.8
    7
    pr at max. Ipi   pr@PIImax (MPa) 3.35
    deq at max. Ipi deq@PIImax (cm)
    Focal Length FLX(cm)
    FLY(cm)
    Ipa,αat max. 
    MI IPA . 3@MIma
    x  (W/
    cm2)362.72
    Frequency (MHz)
    T4T4
    Image Depth (cm) 14.0
    14
    ARM_Rev_2.book  Page 66  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-67
    Direction 5122542-100 Rev. 2 Table 1-53:  Transducer Mode 7S Operating Mode: Color Flow Mode (Inc. M-Mode)
    Index LabelMI
    TISTIB
    TIC
    scannon-scannon-
    scan
    Aaprt1
    Global Maximum: Index Value 1.42.1 1.7
    IECFDAUnits
    pra pr. 3 (MPa) 3.4
    PWo (mW)44.9 105.0
    min of [P
    α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
    zs z1(cm)
    zbp zbp(cm)
    zb zsp(cm) 2.262.26
    z  at max. I pi,α zsp(cm)
    deq(zb) deq(zsp) (cm)2.84
    ƒawffc(MHz) 3.463.46 3.46
    Dim of A
    aprtX(cm)1.0 1
    Y(cm)
    0.7 0.7
    t
    dPD (µsec) 095
    prr PRF (Hz) 4251.7
    pr at max. Ipi   pr@PIImax (MPa) 2.5
    deq at max. Ipi deq@PIImax (cm)2.84
    Focal Length FLX(cm)3.5
    FL
    Y(cm)8.0
    I
    pa, αat max. MI IPA . 3@MImax 
    (W/cm2) 171.65
    Frequency (MHz)
    3.3
    Image Depth (cm)2.4
    Vel Scale (kHz)4.3
    Line Density
    Penet OFF
    ROI6010
    PS 10
    10 10
    ARM_Rev_2.book  Page 67  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-68LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Table 1-54:  Transducer Mode: 7S Operating Mode: Pulsed Doppler (inc. B-Mode)
    Index LabelMI
    TISTIB
    TIC
    scannon-scannon-
    scan
    Aaprt1
    Global Maximum: Index Value 1.32.1
    IECFDAUnits
    pra pr.3 (MPa) 2.4
    PWo (mW)8.244.9
    min of [Pα(zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
    zs z1(cm)
    zbp zbp(cm)
    zb zsp(cm) 2.32.33
    z  at max. I pi,α zsp(cm)
    deq(zb) deq(zsp) (cm)0.24
    ƒawffc(MHz) 3.4 5.13.4
    Dim of AaprtX(cm)0.31.8
    Y(cm)0.30.7
    tdPD (µsec) 1.17
    prr PRF (Hz) 1736.1
    1
    pr at max. Ipi   pr@PIImax (MPa) 3.13
    deq at max. Ipi deq@PIImax (cm)0.24
    Focal Length FLX(cm)0.8
    FLY(cm)1.5
    Ipa,αat max. MI IPA . 3@MImax 
    (W/cm2) 384.41
    Frequency (MHz)
    3.3 5.03.3
    Image Depth (cm) 3.5 4.0
    Vel Scale (kHz)1.71.7
    B Pause(on/off)
    SV 22
    PenetOFFOFF
    ARM_Rev_2.book  Page 68  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-69
    Direction 5122542-100 Rev. 2 Table 1-55:   Transducer Mode:7S Operating Mode :CW 
    Doppler
    Index LabelMI
    TISTIB
    TIC
    scannon-scannon-
    scan
    Aaprt1
    Global Maximum: Index Value 0.66 3.034.29 0.0
    IECFDAUnits
    pra pr.3 (MPa) 3.1
    PWo (mW)158.73158.7
    3158.7
    3
    min of [P
    α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
    zs z1(cm)
    zbp zbp(cm)
    zb zsp(cm) 2.842.84
    z  at max. I pi,α zsp(cm) 
    deq(zb) deq(zsp) (cm)0.70
    ƒawffc(MHz) 4.014.014.01 4.01
    Dim of A
    aprtX(cm)0.930.93 0.93
    Y(cm)
    0.70.7 0.7
    t
    dPD (µsec)
    prr PRF (Hz)
    pr at max. Ipi   pr@PIImax (MPa) 1.16
    deq at max. Ipi deq@PIImax (cm)0.70
    Focal Length FLX(cm)5.765.76
    FL
    Y(cm)3.503.50
    I
    pa, αat max. MI IPA . 3@MImax 
    (W/cm2) 
    Frequency (MHz) D4.0D4.0D4.0
    Image Depth (cm)
    Other conditions
    ARM_Rev_2.book  Page 69  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-70LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    ARM_Rev_2.book  Page 70  Monday, March 15, 2010  2:19 PM 
    						
    							LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-1
    Direction 5122542-100 Rev. 2
    Chapter 2
    Measurement Formulas
    Reference for measurement formula tables for the 
    various applications. 
    ARM_Rev_2.book  Page 1  Monday, March 15, 2010  2:19 PM 
    						
    							Measurement Formulas
    2-2LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Measurement Acronyms
    Cardiac Measurements
    The following table shows tha cardiac measurements available 
    on the LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3
    When you make a measurement,you select the abbreviationfor 
    the measurement on the Touch Panel. Most abbreviations are 
    made using acronyms. The following table lists acronyms used 
    for naming cardiac measurements.
    Table 2-1: Cardiac Measurements
    AbbreviationsDefinitionUnit
    %FS LV Fractional Shortening, 2D %
    %FS LV Fractional Shortening, M-mode %
    %IVS Thck IVS Fractional Shortening, 2D %
    %IVS Thck IVS Fractional Shortening, M-mode %
    %LVPW Thck LV Posterior Wall Fractional Shortening, 2D %
    %LVPW Thck LV Posterior Wall Fractional Shortening, M-mode %
    Ao Arch Diam Aortic Arch Diameter cm
    Ao asc Ascending Arch Diameter cm
    Ao Desc Diam Descending Aortic Diameter cm
    Ao Isthmus Aortic Isthmus cm
    Ao Root Diam Aortic Root Diameter cm
    Ao Root Diam Aortic Root Diameter, M-mode cm
    AR ERO PISA:Regurgitant Orifice Area
    cm
    2
    AR FlowPISA:Regurgitant Flow ml/s
    AR PHT AV Insuf. Pressure Half Time ms
    AP Rad PISA:Radius of Aliased Point cm
    AR RV PISA:Regurgitant Volume Flow ml
    AR Vel PISA:Aliased Velocity m/s
    ARM_Rev_2.book  Page 2  Monday, March 15, 2010  2:19 PM 
    						
    							Measurement Acronyms
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-3
    Direction 5122542-100 Rev. 2
    AR Vmax Aortic Insuf. Peak Velocity m/s
    AR VTI Aortic Insuf. Velocity Time Integral cm
    ARed max PC Aortic Insuf. End-Diastole Pressure Gradient mm Hg
    ARed Vmax Aortic Insuf. End-Diastole Velocity m/s
    AV Acc Slope Aortic Valve Flow Acceleration
    m/s
    2
    AV Acc TimeAortic Valve Acceration Time ms
    AV AccT/ET AV Acceleration to Ejection Time Ratio
    AV CO Cardiac Output by AOrtic Flow Vmin
    AV Cusp Aortic Valve Cusp Separation, 2D cm
    AV Cusp Aortic Valve Cusp Separation, M-mode cm
    AV Dec Time Aortic Valve Decelation Time  ms
    AV Diam Aortic Diameter, 2D cm
    AV max PG Aortic Valve Peak Pressure Gradient mm Hg
    AV mean PG Aortic Valve Mean Pressure Gradient mm Hg
    AV SV Stroke Volume by Aortic Flow ml
    AV Vmax Aortic Valve Peak Velocity m/s
    AV VTI  Aortic Valve Velocity Time Integral cm
    AVA (Vmax) AV Area by Continuity Equation by Peak V
    cm
    2
    AVA (VTY)AV Area by Continuity Equation VTI
    cm2
    AVA PlanimetryAortic Valve Area
    cm2
    AVETAortic Valve Ejection Time ms
    AVET Aortic Valve Ejection Time, M-mode ms
    CO (A-L A2C) CO 2CH, Single Plane, Area-Length l/min
    CO (A-L A4C) CO 4CH, Single Plane, Area-Length l/min
    CO (Biplane) CO, Bi-Plane, MOD l/min
    CO (Bullet) CO, Bi-Plane, Bullet l/min
    CO (MOD A2C) CO 2CH, Single Plane, MOD (Simpson) l/min
    CO (MOD A4C) CO 4CH, Single Plane, 4CH, MOD (Simpson) l/min
    CO (Cube) Cardiac Output, 2D, Cubic l/min
    Table 2-1: Cardiac Measurements
    AbbreviationsDefinitionUnit
    ARM_Rev_2.book  Page 3  Monday, March 15, 2010  2:19 PM 
    						
    							Measurement Formulas
    2-4LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    CO (Cube)Cardiac Output, M-mode, Cubic l/min
    CO (Tech) Cardiac Output, 2D, Teicholtz l/min
    CO (Tech) Cardiac Outpu, M-mode, Teicholtz l/min
    D-E Excursion MV Anterior Leaflet Excursion cm
    D-E Excursion Mitral Valve D-E Slope cm
    EDV (Bullet) LV Volume, Diastolic, Bi-Plane, Bullet ml
    EDV (Cube) Left Ventricle Volume, Diastolic, 2D, Cubic ml
    EDV (Cube) Left Ventricle Volume, Diastolic, M-mode, Cubic ml
    EDV (Teich) Left Ventricle Volume, Diastolic, 2D, Teicholtz ml
    EDV (Teich) Left Ventricle Volume, Diastolic, M-mode, Teicholtz ml
    E-F (A-L A2C) Ejection Fraction, 2CH, Sigle Plane, Area-Length %
    EF (A-L A4C) Ejection Fraction, 4CH, Sigle Plane, Area-Length %
    EF (Biplane) Ejection Fraction, Bi-Plane, MOD %
    EF (Bullet) Ejection Fraction, 2CH, Bi-Plane, Bullet %
    EF (MOD A2C) Ejection Fraction, 2CH, Single Plane, MOD %
    EF (MOD A4C) Ejection Fraction, 4CH, Single Plane, 4CH, MOD (Simpson) %
    E-F Slope Mitral Valve E-F Slope m/s
    EF (Cube)  Ejection Fraction, 2D, Cubic %
    EF (Cube) Ejection Fraction, M-mode, Cubic %
    EF (Teich) Ejection Fraction, 2D, Teicholtz %
    EF (Teich) Ejection Fraction, M-mode, Teicholtz %
    EPSS E-Point-toSeptum Separation, M-mode cm
    EPSS 2D E-Point-to-Septum Separation, 2D cm
    ERC Effective Regurgitant Orifice
    cm
    2
    ESV (Bullet)LV Volume, Systolic, Bi-Plane, Bullet ml
    ESV (Cube) Left Ventricle Volume, Systolic, 2D, Cubic ml
    ESV (Cube) Left Ventircle Volume, Systolic, M-mode, Cubic ml
    ESV (Teich) Left Ventricle Volume, Systolic, 2D, Teicholtz ml
    ESV (Teich) Left Ventricle Volume, Systolic, M-mode, Teicholtz ml
    HR AV Heart Rate, Dop bpm
    Table 2-1: Cardiac Measurements
    AbbreviationsDefinitionUnit
    ARM_Rev_2.book  Page 4  Monday, March 15, 2010  2:19 PM 
    						
    All GE manuals Comments (0)