    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 94 
    Z-Mechanism replacement 
    To prepare: 
     Power down the Vivid E9. 
     Disconnect all probes and external I/O cabling. 
     Remove the LCD Rear Cover. 
     Remove the LCD Monitor assembly. 
     Remove the LCD Arm assembly. 
     Remove the Operator Panel, Upper. 
     Remove the Operator Panel, Lower. 
     Remove the Bulkhead Board. 
     Remove the Bulkhead Plate. 
     Remove the Upper Frame Assembly. 
     Remove the Lower Frame Assembly. 
     Remove the XY (Frogleg) Assembly. 
     Remove the XYZ Control Box 
     
     Remove the four screws from the bottom of the Z mechanism: 
     
     Remove these four screws the Z mechanism: 
     
    Z-Mechanism 
      
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 95 
     Lift Z Mechanism away. 
     
    XYZ Control Assembly Replacement 
    To prepare: 
     Power down the Vivid E9. 
     Disconnect the Mains Power Cable from the wall outlet. 
     Disconnect all probes and external I/O cabling. 
     Remove the Left Side Cover. 
     Remove the Rear Cover. 
     Remove the Top Cover. 
    XYZ Control Assembly removal procedure: 
     Disconnect the cables to the XYZ Control Assembly. 
     Unscrew and remove two (2x) screws. 
     
      
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 96 
     Pull the top of the XYZ Control Assembly forwards, up and away 
    from the bracket. 
    XYZ Control Assembly installation procedure: 
     Thread the XYZ Control Assembly into position. 
      Install the three screws at the top, rear. (M6x16 sunken. Torque: 
    9.8 Nm.) 
     Install the two remaining screws on the top. (M4x8, sunken. 
    Torque: 4.9 Nm.) 
     4.) Connect the cables to the XYZ Control Assembly. 
     5.) Install the Top Cover. 
     6.) Install the Rear Cover. 
     7.) Install the Left Side Cover. 
    Main Power Suppy Replacement 
    The Logiq E9 has two types of power supllies the PS Lamba and the 
    Cherokee/Mitra. The mounting is the same for both devices but the 
    connectors are different. It is possible to replace the fan assembly 
    and the 4D Motor Controller (4D MC) motor of the Cherokee/Mitra 
    without replacing the entire MPS unit. 
    The Vivid E9 power supply does not contain any FRUs and must be 
    replaced entirely if it becomes defective. There are several versions 
    so be sure to determine the correct part number for your system 
    before ordering. 
    To prepare for MPS replacement perform the following steps: 
     System Power Down 
     Disconnect all probes and external I/O cabling. 
     Remove both Side Covers. 
     Remove the Top Cover. 
     Remove the Filter Cover. 
      Remove the Rear Cover. 
     Disconnect the battery cable from EPS or CB  
    Vivid E9 Main Power Supply Replacement 
    Follow these steps to remove the MPS: 
      
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 97 
     Disconnect the cables from the top of the power supply. 
     
     Remove the three screws that secure the MPS. 
     Ease the MPS away from the fang bracket. 
     Pull/slide the entire MPS backwards until the connectors to the 
    FEP/Backplane are unseated and lift the MPS away. 
    Logiq E9 Main Power Supply Replacement 
    Follow these steps to remove the MPS: 
    Disconnect the MPS cable and all cables from the top of the MPS: 
     AC output to peripherals, printer 
     Power to BEP 
     USB 
    Remove the three screws that secure the MPS with a 5 mm Allen 
    wrench and ease the MPS away from the fang bracket. 
    Carefully pull the entire MPS backwards until the connectors to the 
    Card Cage/Backplane are unseated. 
    If the 4D Motor Controller option is installed, you will need to 
    disconnect the USB Cable from J28 on the BEP. 
    To replace the new MPS: 
    If needed, remove the 4D 
    motor controller from the 
    PS and install it in the 
    replacement Power Supply 
    before procceding  
    The 4D MC will also have 
    to be transferred to the 
    replacement MPS.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 98 
    Make sure the GFI PCIe cable is routed behind the MPS fang as shown 
    on the left. This is for the GFI Configuration only. MRX systems do not 
    have this cable. 
    Slide the MPS forward to seat the connectors on the card cage. 
    Engage the tab onto the Fang and install the three screws that secure 
    the MPS with a 5 mm Allen wrench (Torque: 10 Nm {7.4 lb-ft}). 
    Replace all cables and covers in reverse order of MPS removal 
    procedure. 
    Back End Processor (BEP) 
    There are several versions of the BEP for both systems. Each BEP 
    version has internal FRU’s such as :DD, I/O board, Battery Pack. We 
    will not cover all permutations in this manual. Please refer to the GE 
    parts replacement documentation for correct parts when preparing 
    to replace BEP or BEP sub-component parts: 
    http://www3.gehealthcare.com/en/Support/Support_Documentation_Library 
     
    Logiq E9 Vivid E9 
    BEP w/ EPS for R1.x.x and R2.x.x: 
     For GFI configurations. 
     For Phase I and BT 2010 
    systems. 
    BEP – MRX w/ EPS for R3.x.x: 
     For MRX configurations. 
     For BT 2011 only systems. 
     Not backward compatible. 
     Requires new I/O board. 
     New PCIe cable for MRX. 
    BEP for BT’09 (4D): 
     Has same 4D video card as V7. 
     Requires SW v108.0.0 and 
    higher. 
    BEP for BT’11: 
     BEP without 4D. 
     Requires SW v110.0.0 or 
    higher. 
     Optional / new 4D video card 
    (NVIDIA FX 1800) to handle 
    the 4D feature. 
     
    Logiq E9 and. Vivid E9 
    BEPs are not compatible 
    with each other.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 99 
     
    Logiq BEP replaceables: 
    Hard Disk drive HDD 
    BEP Power Supply 
    EPS Battery Pack 
    Vivid BEP replaceables: 
    BEP5 power supply 
    Hard Disk drive HDD 
    Graphics adapter 
    Fan 
    If the old BEP is a BEP5 with 4D, and the replacement BEP is a BEP6, a 
    new Graphics Adapter is needed. 
    If the optional DVR Board is present it must be moved over from the 
    old to the new BEP. 
    To prepare for procedure: 
     Record the system’s TCP=P settings and installed Option strings. 
     Back up the Patient Archive images, Report Archive, User-Defined 
    Configuration (Customer Presets), and Service settings. You will 
    need to perform a Restore after replacement of the BEP. 
     Raise the top console to its highest position. 
     Power down the system. 
     Disconnect all probes, and external I/O cabling. 
     Remove the left side cover only, unless the top cover is removed 
    to increase access to BEP top cables then also remove the right 
    side cover. 
     Remove the printer by releasing the two clips holding the printer 
    bracket to the top of the BEP and gently sliding the printer 
    bracket back towards the center of the system to remove the 
    printer bracket. 
    Logiq E9 BEP Replacement 
    This procedure is for Logiq E9 R4.x and later.  
    Patient IO not sold 
    separately.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 100 
    If a new battery does not come installed in the replacement BEP FRU, 
    transfer the battery into the new BEP. 
    1. Disconnect all I/O cables and all cables from the top of the BEP. 
    2. Loosen the thumb screws or screws at the top of the BEP cover, or 
    the three screws that attach the cover to the BEP. Tilt the top of the 
    BEP cover away from the BEP. 
    3. All the connectors to the CB (ChargeBoard) must be “unlocked“ in 
    order to remove them. Slide the locking latch towards you and pull 
    the connector to the left to release. 
    Disconnect: 
     The battery cable from PCN1 on the CB (ChargeBoard). 
     The CB to the PB (PowerBoard) cable from PCN2. 
     
    Place the BEP cover aside. 
    4. Remove the two 5 mm HEX key screws at the inside base of the 
    BEP. 
    5. Release the lock connector for the BEP to Backplane cable, J2 (left) 
    on a MRX Configuration as shown below. 
    Note: BEP replacement 
    requires a software 
    reload. 
    Note: You cannot power 
    up the unit with the cables 
    detached. The system will 
    power up without the 
    battery pack, but the 
    cables on the cover must 
    be attached. 
    When the Charge Board 
    (CB) is exposed first 
    remove the battery cable 
    and the CB to PB from 
    PCN1 on the CB. IF the 
    24V from the battery is 
    exposed and you drop a 
    tool or a screw on the EPS 
    or CB, you could short the 
    battery and cause 
    permanent damage to the 
    system.  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 101 
     
     
    Or, the Backplane cable, J3 (right) and J5 PCIe to the Card Rack on a 
    GFI Configuration, using a BEP6 as shown below. Keep the cable to 
    transfer to the replacement BEP. 
     
    6. Slide the entire BEP away from the system approximately one inch, 
    and finish removing the printer bracket then slide the entire BEP 
    chassis out of the system. 
    7. You will need to remove the following parts from the BEP, to 
    reinstall in the replacement BEP: 
     Option Dongle  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 102 
     Printer Bracket 
     BackPlane to BEP cable 
     Wireless Card and Antennas (if present) 
     Patient I/O (if present) 
     S Video Card (if present) 
    Vivid E9 BEP Replacement 
    This procedure is for Vivid E9 systems. Follow these steps to remove 
    the BEP: 
    1. Disconnect all I/O cables. 
    2. Disconnect the cables from the top of the BEP. 
    3. Unlatch the two latches that clamps the printer bracket to the top 
    of the BEP. 
     
    4. Tilt the front cover forward just until the column cover stop tabs 
    clear and lift the lower column cover. 
    5. Push the printer bracket toward the FEP approximately 1.3 cm 
    (1/2 inch) to free the 3 tabs from the BEP. Push in the direction of 
    the white arrow: 
     
     
    6. Remove the BEP cover:  
    						
    							 GE Logiq E9/Vivid E9 Training Manual 
    © 2016 Conquest Imaging 103 
     
    Loosen the BEP cover’s fixing screw(s): 
     BEP6 has two finger screws on the top of the BEP Cover. 
     Newer versions of the BEP5 have three Phillips screws, one on 
    the top and one on each side of the BEP Cover. 
     Older BEP5 have one finger screw at the top of the BEP cover. 
    Tilt the top of the BEP cover away from the BEP and set the cover 
    aside. 
    7. Remove the Printer Bracket from the BEP. 
    8. Remove the two hex key screws at the inside base of the BEP. 
    9. Reach inside the BEP and disconnect the Boundary Scan Cable 
    from the FEP Backplane Connector. It is available through a cut 
    out in the BEP’s rear cover. 
    10. Slide the entire BEP out of chassis. 
     
    Note: If a DVR broad is installed, remove it from the existing BEP and 
    install it in the replacement BEP. 
    Note: If 4D is installed:  
    If you change from BEP6 to a new BEP6 or from a BEP5 to a new BEP5 
    you can move the graphics adapter  
    If you remove a BEP5, and the new BEP is a BEP6, you will need to 
    order a new BEP6 compatible graphics adapter.  
    						