    							29 | P a g e    © Adam Equipment Company 2009 
    5.4.4 F
    FF
    F4
    44
    4
    - Weight Calibration 
     
    To  calibrate  the  scales  it  is  necessary  to  have  a  w eight  suitable  to  the 
    capacity of the scale. It should have a mass near t o the capacity of the scale, 
    without exceeding it. For example use 20kg mass to  calibrate the CCEU. The 
    calibration  weight  should  be  accurate  to  assure  acc uracy  of  all  weighing. 
    Weights of OIML Class M1 or better should be used.    Other values of  mass 
    can be used, see below.  
    The  user  calibration  must  verify  the  calibration  is   within  10%  of  this 
    calibration.    This  prevents  calibration  with  no  wei ght  on  the  scale  or  with 
    grossly wrong weights.  
     
       When the display is showing: 
    “   F4   ”    “   CAL     ”        “   SEt  ” 
     
      Press the  [Total / 
    
    
    ] key.  The display will show the following: 
    “   CAL   ”   “   0.000   ” “  12345 ” 
    The number in the TOTAL window is the number of int ernal 
    counts representing the zero weight,  from the anal og to digital 
    converter within the scale.   
       Make sure no other  weights are on the scale. Press  the [Total  /  
    
    
    ]  key 
    to set the zero point.  The display will show: 
     
    “   LoAd   ”    “      20”  “  KiLoS ” 
     
       If a different mass is required from the one shown  press the [CE] key to 
    clear  the  value,  then  enter  the  value  of  the  calibr ation  mass,  i.e.  10 
    kilograms.     
       The display will show: 
     
    “   LoAd   ”    “      10”  “  KiLoS ” 
        
    						
    							30 | P a g e    © Adam Equipment Company 2009 
     Place the calibration weight on the scale.  
     
     
    Press the  [Total / 
    
    
    ] key to show the ADC values for this weight. 
     
    “   LoAd   ”    “      10  ”   “  345689” 
     
      
    Press  the  [Total  / 
    
    
    ]  key  to  show  the  ADC  values  for  this  weight.  The 
    display will then show:
     
     
     
    “   SPAN   ”   “  PASS ”   “               ” 
       
       After few seconds it will show the next function. 
     
    “     F5  ”         “  ACCES ”           “  CAL ” 
     
      Remove the weight from the scale. 
      
    If  [CE/ESp]  is pressed during calibration return to normal oper ation without 
    doing the calibration.     
    5.4.5  F5- Access Code Number
     
      
    To  change  the  user  access  code  number  used  to  enter   the  section  of  the 
    program  for  changing  coin  weights  (see  Section  5.2)   follow  this  procedure. 
    This  parameter can  only  be  accessed by  skipping  the   F3 Calibration  function 
    by using the  [MC/
    
    
    ] key. 
      When the display is showing: 
     
    “   F5   ”     “  ACCES ”    “ SEt     ” 
     
     Press the  [Total / 
    
    
    ] key.  The display will show:  
    						
    							31 | P a g e    © Adam Equipment Company 2009 
     
    “  Pin 1   ”   “             ”  “             ” 
                Enter the new  access code 
    “  Pin 1   ”   “   - - - -”  “             ” 
     press the  [Total/ 
    
    
    ]  key to confirm and store the value, The scale 
    will ask you to repeat the number again for securit y. 
    “  Pin 2   ”   “             ”  “             ” 
                Enter the new  access code 
    “  Pin 2   ”   “   - - - -”  “             ” 
     Press  [CE  /Esc]  to  escape  without  changes  before  it  is  stored  or 
    press the  [Total/ 
    
    
    ] key to confirm and store the value.   
     
    “   F5   ”     “  ACCES ”    “ Set     ” 
     To return to normal operation press the  [CE /Esc] key. 
      
    5.4.6  F6
    F6F6
    F6
    - Language
     
      
    The  scale  can  be  set  to  print  results  in  English,  G erman,  French  or  Spanish. 
    To change the language,  follow this procedure.   
      When the display is showing: 
    “   F6   ”   “  LAnG ”     “  SEt ” 
       
     Press the  [Total/ 
    
    
    ]  key. 
    “   F6   ”     “  LAnG ”  “   ENG ” 
                                      Current setting,   English 
        
    						
    							32 | P a g e    © Adam Equipment Company 2009 
     For a new setting , use the  [MC/
    
    
    ] key to cycle through the options,: 
    ENG = English  
    DEU = German  
    FRA = French 
    ESP = Spanish 
     
      Press   [CE  /Esc]  to  escape  without changes  before  it  is  stored  or p ress 
    the  [Total /  
    
    
    ] key to confirm and store the new value.   
     
    “   F1  ”   “  baud ”        “  SEt ” 
     Press  [CE /Esc]  to  return to normal operation. 
      
    						
    							33 | P a g e    © Adam Equipment Company 2009 
      
    6.0  SECURE
     PARAMETERS 
    6.1  SETTING UP SECURED PARAMETERS  
      
    The  following  parameters  can  only  be  accessed  when  the  service  access 
    code 9999 is used.  
    Turn the scale on while holding the  [Total/ 
    
    
    ]  key down and then release the 
    key.  
    Key in the Service Access Code as described in sect ion 5.1.1. 
    The first function shown is  F1, the Baud Rate Setting.   
    “   F1  ”   “   bAud ”     “  SEt   ” 
    This procedure gives access to the following parame ters: 
      F1  Baud Rate Setting   
      F2  Parity setting 
      F3  Sleep 
    F4  Weight Calibration 
      F5  Access Code Number Setting  F6  Language 
      
      F7  Enable the   [€/£] key function. 
      F8  Denomination Setting 
      F9  Linearity Calibration  
      F10  Factory Calibration Values   
    See Section 5.4 for details of Functions F1 to F6. 
    These  parameters  are 
    set as in section 5.4. 
    These  parameters  are 
    only  accessible  when 
    the service access code 
    is used.  
    						
    							34 | P a g e    © Adam Equipment Company 2009 
     
    6.1.1 F7 - ENABLE EURO / STERLING CHANGE  USING THE  [€/£] KEY 
      
    The  scale  can  be  set  to  permanently  use  only  Euro  o r  Sterling  coins. 
    Disabling the function of the [€/£] key as describe d in section 2.4. 
     
      When the display is showing: 
    “   F7  ”   “  COIn ”     “  SEt ” 
       
     Press the  [Total/ 
    
    
    ]  key. 
    “   F7  ”  “  COIn ”  “       USEr ” 
                                      Current setting,   User select, [€/£] key enabled. 
       
      For  a  new  value,  use  the  [MC/
    
    
    ]  key  to  cycle  through  the  options.,  
    Select User, Euro or Ster (Sterling)  
       
      Press   [CE  /Esc]  to  escape  without changes  before  it  is  stored  or p ress 
    the  [Total /  
    
    
    ] key to confirm and store the new value.   
     
    “   F8   ”   “  dEm ”      “    Set ” 
     
    6.1.2  F8 - Denomination Values 
     
    The  denomination  keys  for  Europe  have  values  pre-as signed,  from  0.01  to 
    2.00.    Plus  the  3 token  keys  can  be  reassigned  also .  In  total  11 keys  can  be 
    used  for  denominations  or  tokens.  This  procedure  wi ll  allow  you  to  change 
    the  denomination  values  associated  with  each  key.    It  may  be  necessary  to 
    change keypad designs for other denominations.   
     
     When the display is showing: 
      
    						
    							35 | P a g e    © Adam Equipment Company 2009 
        “  F8  ”    “  dEM ”      “  SEt  ” 
      
       Press the  [Total/ 
    
    
    ]  key.  The display will show the decimal point 
    position: 
     
        “  Point ”   “AddES ”  “     0.00     ” 
                  Decimal point current position,     
        Press the  [MC/
    
    
    ]   key. To select either 0.00, 0.0 or 0 
     
      Press  [Total/  
    
    
    ]  to set the decimal point.  
     
      The displays will show: 
     
        “   F8  ”    “  dEM ”      “   kEy  ” 
     
     Press  the  denomination  key  to  be  set.      Use  the  key s  for  the 
    tokens or for the denominations as needed.    
    The displays will show:   
        “   F8  ”    “  K01 ”      “   1.00  ” 
     
    Where  the  K01  is  the  key  number  for  the  key  being  u sed.    The 
    display also shows the denomination currently selec ted.  
     
        “   F8  ”    “   K08 ”     “     0.10  ” 
      
    						
    							36 | P a g e    © Adam Equipment Company 2009 
        For example the number 8 key has a value of 0.10.   
     
     To  enter  a  new  value for  this key  position simple u se the  numeric 
    keys and press  [Total/ 
    
    
    ] to save the new value.    
     
      To  set  a  key  position  for  a  token  press  the  [MC/
    
    
    ]  to  set  T1,  T2 
    etc. up to T8.   
     
      The display will show “ SAVE” and then display: 
     
        “   F8  ”    “  dEM ”  “   kEy  ” 
     
    You can select another key if necessary.  
    When  all  changes  have  been  made  press  the  [CE  /Esc]  key  to 
    return  to  the  parameters.  Press  [MC/
    
    
    ]  to  advance  to  the  next 
    parameter. 
      
    6.1.3  F9- Linearity Calibration
     
      
        The display will show:   
        “     F9”    “ LinEr”  “   SEt  ” 
     
    The  linearity  of  the  scale  is    set  by  calibrating  w ith  2  masses, 
    1/2capacity (10kg)  and full capacity (20kg).     
     
      Press the  [Total/ 
    
    
    ] key to enter this function. 
     
      The display will show the following:  
    						
    							37 | P a g e    © Adam Equipment Company 2009 
    “  CAL   ”    “   0.000  ” “  12345 ” 
    The number in the TOTAL window is the number of int ernal 
    counts representing the zero weight,  from the anal og to digital 
    converter within the scale.   
     Make sure no other weights are on the scale. Press  the [Total /  
    
    
    ] 
    key to set the zero point.  The display will show: 
     
    “   LoAd 1  ”   “      10”  “  23456 ” 
     Place 10kg mass on the scale and then press the  [Total/ 
    
    
    ] key                                                                  
     
     The display will then show: 
     
    “   LoAd   ”    “      20”  “  23456 ” 
     Place 20kg mass on the scale and then press the  [Total/ 
    
    
    ] key 
     
      The display will then show: 
     
     
    “   SPAN   ”   “  PASS ”   “               ” 
       
      After few seconds it will show the next function. 
     
    “     F10  ”         “  FACt ”           “  SEt ”  
    						
    							38 | P a g e    © Adam Equipment Company 2009 
     
    6.1.4 F10- Factory Calibration Value 
      
    This calibration is stored in a separate location f rom the user calibration (F4).  
    If a scale has been disrupted it is possible to rec all this calibration in order to 
    have a scale that has the original calibration rest ored.   
     
    This  calibration  can  be  done  at  any  time  without  ch ecking  for  being  within 
    10%  of  the  previous  factory  calibrations.    This  val ue  will  also  overwrite  the 
    user  calibration  value  as  set  in  F3.      When  the  use r  calibration  is  performed 
    the  results  must  agree  within  10%  of  the  factory  ca libration.    This  prevents 
    calibration with no weight on the scale or with gro ssly wrong weights. 
     
      When the display is showing: 
     
        “   F10  ”    “ FACt”   “  Set ” 
      
        Press the  [Total / 
    
    
    ] key.  The display will show: 
        “  0.000 ”   “         ”  “ xxxxxx” 
         
     Make sure no other weights are on the scale. 
     
      Press the  [Total / 
    
    
    ] key. 
     
      The displays will show the calibration weight requi red, equal to the 
    full scale capacity. 
        “ 20.000 ”   “         ”  “ xxxxxx” 
            
    						