Home > GE > Medical Equipment > GE Aestiva 5 Service Manual

GE Aestiva 5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Aestiva 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							1006-0939-0004-1
    4-1wWW W W
    AA A A
    RR R R
    NN N N
    III I
    NN N N
    GG G G
    TO PREVENT FIRES:
    •Use lubricants approved for anesthesia or O
    2 equipment, such as Krytox®1.
    •Do not use lubricants that contain oil or grease. They burn or explode 
    in high O
    2 concentrations.
    •All covers used on the system must be made from antistatic (conductive) 
    materials. Static electricity can cause fires.
    wWW W W
    AA A A
    RR R R
    NN N N
    III I
    NN N N
    GG G G
    Obey infection control and safety procedures. Used equipment may 
    contain blood and body fluids.
    In this sectionRepair policy  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
    Manifold maintenance  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
    Expiratory valve maintenance   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
    Receiver maintenance (active gas scavenging only)  . . . . . . . . . . . . . . . . . . . . . . . . 4-8
    Flow sensor maintenance   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
    Breathing circuit maintenance   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
    Bellows maintenance  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
    Bellows tests  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
    O
    2 sensor calibration - 21% O2  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
    O
    2 sensor calibration - 100% O2  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
    Flow sensor calibration   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
    How to prevent water build-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
    1.® Krytox is a registered trademark of Dupont de Nemours E.I. & Company Inc
    4 User Maintenance 
    						
    							Aestiva
    4-21006-0939-000
    4-2
    Repair policy
    Do not use malfunctioning equipment. Make all necessary repairs or have the 
    equipment serviced by an authorized Datex-Ohmeda service representative. After repair, 
    test the equipment to ensure that it is functioning properly, in accordance with the 
    manufacturer’s published specifications.
    To ensure full reliability, have all repairs and service done by an authorized Datex-
    Ohmeda service representative. If this cannot be done, replacement and maintenance 
    of those parts listed in this manual may be undertaken by a competent, trained 
    individual having experience in the repair of devices of this nature.
    wCC C C
    AA A A
    UU U U
    TT T T
    III I
    OO O O
    NN N N
    No repair should ever be attempted by anyone not having experience in the 
    repair of devices of this nature.
    Replace damaged parts with components manufactured or sold by Datex-Ohmeda. 
    Then test the unit to ascertain that it complies with the manufacturer’s published 
    specifications.
    Contact the Datex-Ohmeda Customer Service Center for service assistance. In all cases, 
    other than where Datex-Ohmeda’s warranty is applicable, repairs will be made at Datex-
    Ohmeda’s current list price for the replacement part(s) plus a reasonable labor charge. 
    						
    							4 User Maintenance
    1006-0939-0004-34-3
    Manifold maintenance
    Step 1
    Complete the basic 
    disassembly (Refer to the 
    Cleaning and Sterilization 
    section).
    Step 2 
    (APL valve)
    1
    Remove the diaphragm. 
    Replace it if:
    •The disk, cage, or 
    diaphragm is damaged.
    •The diaphragm looks 
    worn.
    •The diaphragm fits over 
    the rim.
    Step 3
    (Seals)
    Loosen the thumbscrew and 
    remove the plate.
    • Replace the seal if 
    damaged or worn.
    AB.23.094
    AB.23.069
    AB.23.070 
    						
    							Aestiva
    4-41006-0939-000
    4-4
    Step 4
    (Seals con.)
    Loosen the thumbscrews.
    Step 5
    (Seals con.)
    Open the manifold (top 
    middle, and bottom parts).
    Step 6
    (Seals con.)
    Replace the parts if they are 
    damaged or worn.
    AB.23.073
    AB.23.074
    Side
    TopBottom
    AB.23.071 
    						
    							4 User Maintenance
    1006-0939-0004-54-5
    Step 7
    (Bag/Vent)2
    Replace the parts if they are 
    damaged or worn.
    Step 8
    Put the parts together 
    •Make sure all gaskets are 
    fully seated.
    •Fully tighten the 
    thumbscrews.
    1.When you turn the APL knob, you change the pressure that is necessary to push the disk up and permit gas 
    to exit.
    2. The Bag/Vent switch pushed down plungers to close the other path. In the Bag position, the vent path is 
    closed, etc.
    Before you use the system, complete the preoperative test procedure. Refer to 
    "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
    AB.23.077
    AB.23.075 
    						
    							Aestiva
    4-61006-0939-000
    4-6
    Expiratory valve maintenance
    Disassemble the manifold and examine the seals at 12 autoclave cycles. 
    These conditions indicate a possible leak in the exhalation valve:
    •A decrease in airway pressure during the inspiratory pause in the volume control 
    mode of ventilation.
    •Alarms for low tidal volume.
    •The ventilator circuit has a leak, but the bellows pass the bellows tests in this 
    section.
    Step 1
    Remove the exhalation valve 
    (Cleaning and Sterilization 
    section).
    Step 2
    Remove the top cover.
    AB.23.089
    AB.23.093 
    						
    							4 User Maintenance
    1006-0939-0004-74-7
    Before you use the system, complete the preoperative test procedure. Refer to 
    "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
    Step 3
    Unscrew the lock ring (one 
    quarter turn) and remove the 
    cap.
    Step 4
    Put the parts together 
    •Align the dimples in the 
    cap with the valve seat.
    •To install the top, align it 
    with the base and push 
    down.
    AB.23.072
    Dimples
    AB.23.021 
    						
    							Aestiva
    4-81006-0939-000
    4-8
    Receiver maintenance (active gas scavenging only)
    A blocked filter decreases flow through the scavenging system. In extreme cases this can 
    open the high-pressure relief valve and vent to room air.
    Note: Passive gas scavenging does not require maintenance.
    w WW W W
    AA A A
    RR R R
    NN N N
    III I
    NN N N
    GG G G
    This system operates correctly at the electrical interference levels of IEC 
    601-1-2. Higher levels can cause nuisance alarms that may stop 
    mechanical ventilation.
    •Always make sure that the rear cover is installed.
    •Completely tighten the thumbscrew.
    Before you use the system, complete the preoperative test procedure. Refer to 
    "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
    Step 1
    Remove the receiver:
    •Loosen the thumbscrew.
    •Remove the rear cover. 
    •Pull out and up on the 
    receiver.
    Step 2
    Replace the filter (friction fit 
    with gasket).
    AB.23.091
    AB.23.092
    Filter 
    						
    							4 User Maintenance
    1006-0939-0004-94-9
    Flow sensor maintenance
    Before you use the system, complete the preoperative test procedure. Refer to 
    "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
    Step 1
    Remove the flow sensor 
    module (Cleaning and 
    Sterilization section).
    Push in the tab and remove 
    the cover.
    Step 2
    Install the new sensor with 
    the arrows up. 
    Keep the tubes straight.
    Step 3
    Slide on the cover. Do not 
    pinch the tubes.
    AA.96p.037
    AA.96p.039 AA.96p.038 
    						
    							Aestiva
    4-101006-0939-000
    4-10
    Breathing circuit maintenance
    Disassemble the patient circuit module to replace check valves or seals. 
    Step 1
    Remove the patient circuit 
    module (Cleaning and 
    Sterilization section).
    Step 2
    Hold in the tab.
    Step 3
    Remove the rear cover.
    Note: The rear seal is keyed 
    to fit a specific module.
    AA.96p.017
    AA.96p.041
    AA.96p.042 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Aestiva 5 Service Manual