Home > GE > Medical Equipment > GE Aestiva 5 Service Manual

GE Aestiva 5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Aestiva 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							4 User Maintenance
    1006-0939-0004-114-11
    Do the steps in the opposite order to assemble the module.
    Before you use the system, complete the preoperative test procedure. Refer to 
    "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
    Step 4
    Push in the tabs and remove 
    the top cover.
    Step 5
    Remove the seals.
    Turn and pull to remove the 
    check valves.
    AA.96p.043
    AA.96p.044 AA.96p.045 
    						
    							Aestiva
    4-121006-0939-000
    4-12
    Bellows maintenance
    Step 1
    Remove the bellows from the 
    breathing system.
    Step 2
    Turn the housing 
    counter-clockwise and lift.
    Step 3
    Remove the bottom edge of 
    the bellows from the rim. 
    AA.96p.023 AA.96p.051
    AA.96p.052 
    						
    							4 User Maintenance
    1006-0939-0004-134-13
    wWW W W
    AA A A
    RR R R
    NN N N
    III I
    NN N N
    GG G G
    Do not disassemble the pressure relief valve. This can damage the seat or 
    diaphragm and cause injury to the patient.
    Step 4
    Remove the disk from the 
    bellows.
    Step 5
    Remove the ring from inside 
    the top of the bellows.
    Step 6
    Push the latch toward the 
    center and remove the rim.
    Step 7
    Remove the pressure relief 
    assembly. 
    AA.96p.031
    AA.96p.032
    AA.96p.033
    AA.96p.034 
    						
    							Aestiva
    4-141006-0939-000
    4-14
    Do these steps in the opposite order to assemble the bellows. If you see a dust-like 
    powder on the housing or the bellows, apply a thin layer of KRYTOX lubricant to the ribs 
    of the bellows housing. Make sure the lubricant is applied smoothly and there are no 
    lumps.
    Make sure that:
    •The arrow on the seal points up.
    •You hear a double click when you install the rim.
    •The rim is locked in position.
    •The inner ring is correctly installed inside the top of the bellows.
    •Only the bottom ring of the bellows fits over the rim.
    •The housing is locked in position. You cannot lift it off.
    Step 8
    Push the latch towards the 
    center and remove the 
    locking tabs.
    Step 9
    Remove the seal.
    AA.96p.035
    AA.96p.036 
    						
    							4 User Maintenance
    1006-0939-0004-154-15
    Bellows tests
    wWW W W
    AA A A
    RR R R
    NN N N
    III I
    NN N N
    GG G G
    Objects in the breathing system can stop gas flow to the patient. This can 
    cause injury or death:
    •Do not use a test plug that is small enough to fall into the breathing 
    system.
    •Make sure that there are no test plugs or other objects caught in the 
    breathing system.
    wWW W W
    AA A A
    RR R R
    NN N N
    III I
    NN N N
    GG G G
    The bellows assembly test does not replace the preoperative tests. Always 
    complete the tests in the section Preoperative Tests before you use the 
    system with a patient.
    This test makes sure that all components are correctly assembled. It is not an alternative 
    to a complete system checkout.
    If the bellows operates correctly, install it in the system.
    If there is a problem, disassemble the bellows. Look for and replace damaged parts.
    Step 1
    Hold the bellows assembly 
    vertical and close the 
     ports (A and B).
    A
    B 
    						
    							Aestiva
    4-161006-0939-000
    4-16
    Step 2
    Invert the bellows. They must 
    not fall more than 
    100 mL/min.
    If it does:
    •The ports are not tightly 
    sealed.
    •The bellows is incorrectly 
    installed.
    •The seal inside the 
    bellows is not correctly 
    installed (with its groove 
    pointed up).
    •Parts are damaged.
    Step 3
    Remove the plugs from the 
    ports. Permit the bellows to 
    fully extend.
    Step 4
    Close port C.
    AA.96p.124
    C 
    						
    							4 User Maintenance
    1006-0939-0004-174-17
    If the result for all the bellows tests was “passed,” install it in the system.
    Before you use the system, complete the preoperative test procedure. Refer to 
    "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.
    Step 5
    The bellows must not fall 
    more than 100 mL/min.
     If it does:
    •The port is not tightly 
    sealed.
    •The bellows or the 
    pressure relief valve is 
    not correctly installed.
    •Parts are damaged.
      
    						
    							Aestiva
    4-181006-0939-000
    4-18
    O2 sensor calibration - 21% O2
    Note: The 21% O2 calibration must be done before the 100% O2 calibration. During O2 
    calibration the screen replaces O
    2 data with - - -. Mechanical ventilation must be off to 
    perform calibration.
    This procedure takes three minutes or less.
    Step 1
    Push the menu key.
    Step 2
    Turn the knob to select 
    Setup/Calibration (highlight).
    Step 3
    Push the knob to show the 
    next screen.
    AB.29.013
    AB29.002 AB.90.036
    AB.90.037
    AB.90.091 
    						
    							4 User Maintenance
    1006-0939-0004-194-19
    The screen shows “Calibrating. . .”, followed by the result (“Complete” or “Failure”).
    Step 4
    Turn, then push the knob to 
    select O
    2 Sensor Cal.
    Step 5
    Select 21%. Then, push the 
    knob.
    Step 6
    Complete the steps shown on 
    the screen.
    • Remove the O
    2 sensor from 
    the circuit.
    •Do not twist or stress the 
    cable.
    •Make sure the cable is 
    connected.
    Step 7
    Select Start Cal.
    Then, push the knob.
    AB29.002
    AB.29.012AB.90.048
    AB.90.049
    AB.48.080 
    						
    							Aestiva
    4-201006-0939-000
    4-20
    If the calibration fails:
    •perform the calibration again.
    •do a 100% O
    2 sensor calibration. If this passes, repeat the 21% O2 calibration.
    If the calibration passes, install the O
    2 sensor. If necessary, do the 100% O2 calibration.
    Before you use the system, complete the preoperative test procedure. Refer to 
    "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual. 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Aestiva 5 Service Manual