Page 121

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 2 - Site Preparations 2 - 11
2-3-2 Required Facility Needs
NOTE: GE Healthcare requires a dedicated power and  ground for the proper operation of its Ultrasound 
equipment. This dedicated power shall originate at  the last distribution panel before the system. 
Sites with a mains power system with defined Neutral and Live: 
The dedicated line shall consist of  one phase, a neutral (not shared with any other circuit), and a...

Page 122

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
2 - 12 Section 2-3 - Facility Needs
2-3-3 Desirable Features
• Door is at least 92 cm (3 ft.) wide
• Circuit breaker for dedicated po wer outlet is easily accessible
• Sink with hot and cold water
• Receptacle for bio–hazardous waste, like used probe sheaths
• Emergency oxygen supply
• Storage for linens and equipment
• Nearby waiting room, lavatory, and dressing room
• Dual level lighting (bright and dim)
• Lockable cabinet...

Page 123

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 2 - Site Preparations 2 - 13
2-3-5 Networking   Setup Requirements
2-3-5-1 Stand Alone Scanner (without Network Connection)
None.
2-3-5-2 Scanner Connected to Hospital’s Network Supported networks:
DICOM Network (option)
2-3-5-3 Purpose of the DICOM Network Function DICOM services provide the operator  with clinically useful features for moving images and patient 
information over a hospital network. Examples of DI COM...

Page 124

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
2 - 14 Section 2-3 - Facility Needs
2-3-5-5 Pre-quote Worksheet Figure 2-2   Pre-quote Worksheet for DICOM Network Information
Vivid 7
Host Name
AE Title
Local PortIP Address
Net Mask.
..
.
..
ROUTING INFORMATION
ROUTER2
ROUTER3 ROUTER1
.
..
.
..
.
..
.
..
.
..
.
..
.
.. Destination 
IP Addresses
GATEWAY IP Addresses
Default
DICOM APPLICATION INFORMATION NAME
Store 2 Store 3 Store 1
MAKE/REVISION IP ADDRESSES PORT
AE TITLE
Store 5...

Page 125

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 1
Chapter 3
System Setup
Section 3-1
Overview
3-1-1 Purpose of Chapter 3
This chapter contains information needed to install  Vivid 7. Included is a procedure that describes how 
to receive and unpack the equipment and  how to file a damage or loss claim. 
How to prepare the facility and unit  of the actual installation, and how to check a nd test the unit, probes, 
and external peripherals for...

Page 126

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 2 Section 3-2 - Installation Reminders
Section 3-2
Installation Reminders
3-2-1 Average Installation Time
3-2-2 Installation Warnings
Table 3-2    Average Installation Time
Description
Average 
Installation 
Time
Comments
Unpacking the scanner 0.5 hour
Install Scanner wo/options 4 hours Dependent on the configuration
DICOM Network Configuration 2 hours  Dependent on the configuration
DANGERWHEN USING ANY TEST  INSTRUMENT THAT...

Page 127

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 3
3-2-2 Installation  Warnings (cont’d)
CAUTIONDO NOT OPERATE THIS UNIT UNLESS ALL  BOARD COVERS AND FRAME PANELS ARE 
SECURELY IN PLACE. SYSTEM PERFORMANCE AND COOLING REQUIRE THIS.
CAUTIONOPERATOR MANUAL(S) 
THE USER MANUAL(S) SHOULD BE FULLY READ AND UNDERSTOOD BEFORE OPERATING 
THE Vivid 7 AND KEPT NEAR THE UNIT FOR QUICK REFERENCE.
CAUTIONACOUSTIC OUTPUT HAZARD 
ALTHOUGH THE ULTRASOUND ENERGY...

Page 128

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 4 Section 3-3 - Receiving  and Unpacking the Equipment
Section 3-3
Receiving and Unpacking the Equipment
3-3-1 Overview
3-3-1-1 Purpose of this section 
This section describes how to receive and unpack Vivid 7.
Table 3-4    Contents in this section
SectionDescriptionPage Number
3-3-1 Overview 3-4
3-3-2 The Post Delivery Checklist 3-5
3-3-3 The Tilt & Shock Indicators 3-6
3-3-4Receiving the Vivid 73-7
3-3-5Unpacking Vivid 73-10...

Page 129

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 5
3-3-2 The Post Delivery Checklist
3-3-2-1 Introduction Before shipment from the  factory, the Vivid 7 has been thoroughly  tested and visually inspected. Vivid 7 
ultrasound scanners are fine tuned electronic inst ruments and should be treated properly during 
transportation.
To learn about any issues that are discovered at th e reception of the package or when unpacking and 
installing the Vivid 7,...

Page 130

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 6 Section 3-3 - Receiving  and Unpacking the Equipment
3-3-3 The Tilt & Shock Indicators
3-3-3-1 Overview
Improper handling during transportation may harm the equipment inside the package even if the 
package itself is undamaged. 
To make it easier to detection if the handling duri ng transportation has been improper, a set of Tilt & 
Shock indicators have been attached to the transportation box.
3-3-3-2 Position of the Tilt &...