Home > Philips > Medical Equipment > Philips Fm2 Service Manual

Philips Fm2 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Philips Fm2 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 619 Philips manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Introduction
    Chapter 8 - Test and Inspection Matrix
    79
    Caution
    General-purpose personal computers and modems are not designed to 
    meet the electrical safety requirements of medical devices. The RS-232 and 
    parallel connectors on the monitor are electrically isolated to permit safe 
    connections to non-medical devices, which should be connected with a 
    cable of sufficient length to prevent the non-medical equipment from 
    contacting the patient.
    If  a non-medical device is connected to the serial port, a recorder or 
    printer connected at the same time to the parallel port should be located 
    outside the patient vicinity, as the two interfaces are not isolated against 
    each other.
    Introduction
    This chapter contains maintenance and safety information for the FM-2 
    antepartum fetal monitor and its accessories.
    All checks that require the instrument to be opened must be made by qualified 
    service personnel. Please contact your local Agilent representative when you wish 
    safety and maintenance checks to be carried out.
    To ensure that your monitor functions properly and safely, you must adhere to 
    the standards stated in this manual. 
    						
    							When to Perform Tests Blocks
    80Chapter 8 - Test and Inspection Matrix
    When to Perform Tests Blocks
    Table 20 shows which test blocks to perform after repair and preventive 
    maintenance tasks. 
    Table 20 FM-2: When to Perform  Test Blocks
    Service EventTest Block(s) Required
    (See Table 21 on page 81)
    Installation
    The product is customer installed.
    Refer to the Instructions for Use.Visual and Power On
    Preventive
    Maintenance
    Preventive maintenance is limited 
    to exchanging the batteryPo we r  O n
    Safety Tests
    Repairs
    Unit Exchange Visual and Power On
    Unit opened Visual and Power On
    Front End Connector Board 
    exchangedPo we r  O n
    Performance Tests for Trans-
    ducers
    Safety Tests
    Rear Connector Board exchanged Power On
    Safety Tests
    Power Supply exchanged Power On
    Battery exchanged Power On
    Safety Tests 
    						
    							When to Perform Tests Blocks
    Chapter 8 - Test and Inspection Matrix
    81
    Table 21 Test and Inspection Matrix
    Test Block 
    NameTest Expected ResultsWhat to 
    Record
    Visual Test
    (See 20)Inspect exterior of fetal 
    monitor for damage.No Visual Damage V:P or V:F
    Pow e r - O n  Se l f -
    Te s t
    (See 20)Power on the monitor Displays Normal Monitoring 
    Screen Configuration and 
    emits tonesPO:P or PO:F
    Safety Tests:
    • Safety Test 1
    (See 86)Enclosure Leakage 
    Current - S.F.C. Open 
    SupplyAll Safety Test results are in 
    range of Table 22 on page 87S:P or S:F
     Safety Test 2
    (See 88)Patient Leakage Cur-
    rent - S.F.C. Open 
    SupplyAll Safety Test results are in 
    range of Table 23 on page 89S:P or S:F
    Key to Table:P = Pass F = Fail 
    						
    							Battery Charging and Maintenance
    82Chapter 8 - Test and Inspection Matrix
    Battery Charging and Maintenance
    Before attempting to perform any tests, it is important to verify that the battery 
    is charged and functional.
    Using the Battery
    You can run the monitor for approximately six hours on battery power at room 
    temperature of 25o C from a fully charged battery (with a minimum of two 
    hours monitoring at 10o C from a fully charged battery). The age of the battery, 
    operating temperature and number of recharge cycles affects this. A battery life 
    cycle is approximately 200 charge/discharge cycles.
    Low Power Warning
    A low power warning occurs when the battery power is running low. The battery 
    fuel gauge flashes, and a repetitive audible tone sounds. Remaining monitoring 
    time is typically 30 minutes with a new battery at room temperature of 25
    o C. In 
    an older monitor, in which the battery has had a number of recharge cycles, it 
    may occur when fewer minutes remain.
    The warning continues until the low power condition disappears (when you 
    connect the monitor to an AC power supply) or until the system stops 
    functioning because of the low power.You cannot silence it, and you cannot 
    adjust its volume.
    The monitor retains any stored fetal traces, even if it shuts down because of lack 
    of battery power. 
    						
    							Preventive Maintenance
    Chapter 8 - Test and Inspection Matrix
    83
    Recharging the Battery
    Use the power supply to recharge the battery from an AC power source. This 
    takes a maximum of 14 hours when monitoring simultaneously, or eight hours if 
    no monitoring occurs. Typical times are 11 hours if you are monitoring 
    simultaneously, or 4.75 hours if no monitoring occurs.
    Storing
    A stored battery loses capacity over time. To minimize loss of capacity, recharge 
    the battery at least every six months. A battery typically has three to five years 
    shelf life.
    Preventive Maintenance
    The only preventive maintenance task required for the monitor is battery 
    exchange. Life expectancy of a battery depends on the frequency and duration of 
    use. The battery should be replaced every two years.
    See “Removing the Battery” on page 43, and “Replacing the Battery” on page 
    44. 
    						
    							Safety Testing
    84Chapter 8 - Test and Inspection Matrix
    Safety Testing
    This section defines the test and inspection procedures applicable to the FM-2 
    fetal monitors. Use the tables in the following section to determine what test and 
    inspection results must be reported after a repair has been carried out. 
    Test Blocks in Table 20 tells you when to carry out the tests blocks
    Test and Inspection Matrix in Table 21 tells you the expected results of the 
    tests and what to record.
    Warning
    Apply the safety test limits set by the local standards and statutes applicable 
    to the country of installation, such as IEC 601-1, UL2601, CAN/CSA-
    C22.2 No. 601.1.M90.
    The safety tests defined in this chapter are derived from local and 
    international standards but may not be equivalent. They are NOT a 
    substitute for local safety testing.
    Caution
    Successful completion of the patient safety tests does not ensure the correct 
    and accurate functioning of the equipment. 
    						
    							Safety Testing
    Chapter 8 - Test and Inspection Matrix
    85
    Safety Test Procedures
    The test procedures outlined in this section are to be used only for verifying the 
    safe installation or service of the product. “Safety checks at installation refer to 
    safety aspects directly related to the installation and setup activities and not to 
    intrinsic safety features that have already been checked during final acceptance 
    testing at the factory.” 
    [FDA 97-4179 Medical Quality Systems Manual: A Small Entity Compliance 
    Guide (QSR manual). First Edition December 1996. Chapter 14: Storage, 
    Distribution and Installation. This document may be viewed at http://
    www.fda.gov.
    When running safety tests on the peripheral components of the system, refer to 
    the documentation included with those products.
    When performing safety tests, you must use a standard safety tester. You can use 
    testers complying with IEC 601-1 internationally. You may also use testers 
    complying with any local standards and statutes applicable to the country of the 
    installation. For safety test procedures see the operation instructions of the safety 
    tester used.
    If using the Metron safety tester use your local regulations to perform the test, 
    for example
    in Europe: IEC60601-1/IEC60601-1-1
    in the US: UL2601-1
    The Metron Report should print results as detailed in this chapter, along with 
    other data. 
    Note—For information and ordering guides for Metron products contact:
    Metron AS, Travbaneveien 1, N-7044 Trondheim, Norway
    Tel. (+47) 73 82 85 00; Fax (+47) 73 91 70 09; 
    E-mail: [email protected]/[email protected]
    www: http://www.metron-biomed.com
    There are two general categories of safety test:
    Ground integrity
    Electrical Leakage 
    						
    							Safety Testing
    86Chapter 8 - Test and Inspection Matrix
    Ground Integrity
    The FM-2 is a Class 2, Type BF instrument. It has an external power supply, and 
    is constructed with double insulation. The FM-2
     does not require an isolated 
    Earth Ground terminal, neither is one installed. No Protective Ground 
    Continuity check is required.
    Electrical Leakage
    The following tests verify the electrical leakage of the monitor:
    Enclosure leakage Current
    Patient Leakage Current
    Safety Test 1: Enclosure Leakage Current Test
    Normal
    ConditionThis test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied 
    Parts. The test measures leakage current of exposed metal parts of the 
    Instrument Under Test. It tests both normal and reversed polarity. 
    For type BF and CF Applied Parts the test measures AP/GND.
    Safety Test according IEC 601-1 (Clause 19).
    Report the highest value.
    Single Fault
    Condition
    (S.F.C.) Open
    SupplyThis test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied 
    Parts. The test measures leakage current of exposed metal parts of the 
    Instrument Under Test with one supply lead interrupted. It tests both normal 
    and reversed polarity. 
    For type BF and CF Applied Parts the test measures AP/GND.
    Safety Test according IEC 601-1 (Clause 19).
    Report the highest value. 
    						
    							Safety Testing
    Chapter 8 - Test and Inspection Matrix
    87
    Te s t  S t e p s1. Connect the external power supply output cord to the monitor. Then 
    connect the appropriate external power supply input power cord to the 
    analyzer as recommended by the analyzer operating instructions. 
    2. Using the appropriate test cable, connect the analyzer to a fixing nut of 
    the RS-232 connector on the back of the monitor.
    3. Turn the monitor on.
    4. Perform the test as recommended by the analyzer operating instructions.
    The analyzer leakage current indication must not exceed the values listed in 
    Table 22.
    Table 22 Enclosure Leakage Current
    Test Condition PolarityMaximum Leakage Current 
    (µA)
    IEC 601-1 ES1AAMI/ANSI 
    Standard
    NormalNormal 100 µA 100 µA
    Reversed 100 µA 100 µA
    S.F.C.1 (Open Supply)
    1. S.F.C. = Single Fault Condition
    Normal 500 µA 300 µA
    Reversed 500 µA 300 µA 
    						
    							Safety Testing
    88Chapter 8 - Test and Inspection Matrix
    Safety Test 2: Patient Leakage Current Test
    Normal
    ConditionThis test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied 
    Parts. The test measures patient leakage current from any individual patient 
    connection to earth (power ground) in parallel. It tests both normal and reversed 
    polarity.
    Safety Test according IEC 601-1 (Clause 19).
    Report the highest value.
    Single Fault
    Condition
    (S.F.C.) Open
    SupplyThis test is applicable to Class 1 and 2 equipment, type B, BF, and CF Applied 
    Parts. The test measures patient leakage current from any individual patient 
    connection to earth (power ground) in parallel, with one supply lead 
    interrupted. It tests both normal and reversed polarity.
    Safety Test according IEC 601-1 (Clause 19).
    Report the highest value.
    Te s t  S t e p sThis test requires a sample patient cable for each device parameter. These must 
    be configured as recommended by the safety analyzer operating instructions.
    1. Configure the electrical safety analyzer as recommended by the analyzer 
    operating instructions.
    2. Connect the appropriate external power supply input power cord to the 
    analyzer as recommended by the analyzer operating instructions. Connect 
    the external power supply output cord to the monitor.
    3. Turn the monitor on.
    4. Perform the test as recommended by the analyzer operating instructions.
    For each test condition, the measured leakage current must not exceed that 
    indicated in Table 23. 
    						
    All Philips manuals Comments (0)