Home > Philips > Medical Equipment > Philips Fm2 Service Manual

Philips Fm2 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Philips Fm2 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 619 Philips manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Service Menu
    Chapter 4 - Power-up Default Settings and the Service Menu
    29
    System A/D Values
    For the end user or service personnel, the system A/D values require no 
    interpretation.
    The A/D screen displays the current value of each analog-to-digital
    (A/D) channel in volts. The A/D channels read DC voltages, for example, power 
    supply voltages and battery voltage. The A/D values displayed are static, being 
    acquired when the screen is first displayed.
    None of the displayed values may be changed or reset in this screen.When in the 
    A/D screen, the 
    Return item is always highlighted; a press of the wheel will 
    return the user to the Service Menu. Rotating the wheel while in the A/D screen 
    will have no effect. The A/D Values screen appears as follows:
    Figure 3 System A/D Values
    Note—The  screenshot above shows untypical system A/D values that are 
    not representative of normal operation. 
    						
    							Service Menu
    30Chapter 4 - Power-up Default Settings and the Service Menu 
    						
    							Chapter 5 - Troubleshooting31
    5
    Troubleshooting
    General
    This section provides a guide for qualified service personnel in locating the 
    possible cause of an apparent malfunction of the monitor or its accessories. Use 
    of built-in diagnostics, flow charts, and step-by-step procedures are provided as 
    aids in the troubleshooting process.
    The chapter leads you to a probable field replaceable assembly or component. 
    Items that are serviceable only at the factory are also identified, as well as 
    instructions for their return for repair/replacement.
    The initial analysis determines whether the apparent malfunction is caused by a 
    replaceable accessory, or by a replaceable assembly within the monitor case.  
    						
    							General
    32Chapter 5 - Troubleshooting
    Initial Trouble Diagnosis
    The first step in diagnosing a possible malfunction is to duplicate, if possible, 
    the symptoms of the reported malfunction. Thereafter, it is important to 
    categorize the problem as either one of two types, as illustrated in Figure 4. They 
    are:
     Totally inoperative condition, or
     Degraded operation.
    The totally inoperative condition is indicated by the failure of the monitor to 
    respond to the power ON front-panel action, or by complete loss of monitoring 
    functions during operation. Contact your local Agilent Response Center.
    Degraded operation performance might be indicated by the appearance of 
    artifacts in the display, or by gaps in the pixel display, by intermittent actions, or 
    by obviously inaccurate measures of fetal activity, etc.
    Figure 4 Diagnosing Malfunction - Initial Diagnosis 
    						
    							Power Problem Diagnosis
    Chapter 5 - Troubleshooting
    33
    Power Problem Diagnosis
    Diagnostic Flow Chart
    The flow chart of Figure 5 shows a sequence of actions and decisions you can use 
    after the initial diagnosis of a “total failure” as a power supply or display 
    problem.
    Note—If the apparent “total failure” is caused by a malfunctioning display 
    backlight, you may see a faintly lit display with the appropriate Power-On 
    Self Test message. The troubleshooting under these circumstances is 
    described in this section.
    Figure 5 Power Diagnosis - Power Supply or Display  
    						
    							Power Problem Diagnosis
    34Chapter 5 - Troubleshooting
    Diagnostic/Repair Actions
    Examine
    Display
    IntenselyReduce background room illumination. If only the back light fluorescent tube 
    or its high voltage driver has failed, the display may be faintly visible.
    Refer to page 50 for removal procedures.
    Replace
    BatteryRefer to page 43.
    Exchange the
    MonitorIf the display is not at all visible, and there is total failure, exchange the monitor.  
    						
    							Monitoring Degradation Problems
    Chapter 5 - Troubleshooting
    35
    Monitoring Degradation Problems
    Diagnostic Flow Chart
    Figure 6 illustrates a logical sequence of troubleshooting diagnostics when the 
    monitor appears to be operative, but the displayed results are suspect for any 
    number of reasons. The resulting diagnosis will result in replacing either the 
    monitor or the transducers.
    Figure 6 Monitoring Degradation, Diagnostic Flow Chart 
    						
    							Monitoring Degradation Problems
    36Chapter 5 - Troubleshooting
    Diagnostic Repair Actions
    Describe
    Degraded
    PerformanceExamine the display after an apparently successful power up sequence and self-
    test.
    Consider, first, the appearance of artifacts, gaps, or general quality of the display.
    Note the values being reported in the numerical frames (FHR(s)) and TOCO). 
    Are they clearly unreasonable?
    Swap
    TransducersUse a spare transducer that is free of defects and meets factory specifications. At 
    facilities where more than one monitor is installed, it may be convenient to 
    borrow a transducer that is working satisfactorily with another monitor.
    Exchange the
    MonitorIf the monitoring problems persist, or there is total failure, exchange the 
    monitor. 
    						
    							Chapter 6 - Disassembly/Reassembly Guide37
    6
     Disassembly/Reassembly Guide
    Warning
    Performance verification: do not place the monitor into operation after 
    repair or maintenance has been performed, until all performance tests and 
    safety tests listed in Chapter 8 of this service manual have been performed. 
    Failure to perform all tests could result in erroneous monitor readings, or 
    patient/operator injury.
    Warning
    Energized circuits are accessible with the covers open. Do not work on the 
    monitor with the covers open. Only qualified service personnel should 
    open or disassemble the monitor.
    Introduction
    Warning
    Before attempting to open or disassemble the monitor, disconnect the 
    power supply from the monitor.
    Warning
    High voltage is generated by the LCD backlight driver. Do not operate the 
    monitor with the covers open. 
    						
    							How to Use this Chapter
    38Chapter 6 - Disassembly/Reassembly Guide
    Caution
    Observe ESD (electrostatic discharge) precautions when working within 
    the unit.
    Caution
    Important for disassembly! When disassembling the monitor, make sure 
    that the battery leads are disconnected. Otherwise, it is likely that circuit 
    damage will occur.
    The monitor can be disassembled down to all major component parts, 
    including:
    PCBs
    battery
    cables
    function buttons
    chassis enclosures
    Remember to store all screws, nuts, washers and other parts in a safe place for 
    later refitting.
    How to Use this Chapter
    The step-by-step procedures that are used to access replaceable parts of the 
    monitor are illustrated in the Disassembly sections which follow. These describe 
    and photographically illustrate procedures for disassembling the monitor to 
    enable removal and replacement of suspected defective assemblies and 
    components.
    The monitor consists of two main assemblies:
    The front case assembly
    The rear case assembly
    The main PCB assembly is separable from the front case assembly. 
    						
    All Philips manuals Comments (0)