    							Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
    19
                                              Document no. M1031520-02
    2.10.2 Serial data interface
    The serial interface is an asynchronous RS-232 connection, which uses hardware flow control. The 
    communication speed is 115.2 kbps. The communication uses the following parameters: Parity 
    None, Data bits 8 and Stop bits 1.
    2.10.3 Digital and analog outputs
    Digital outputs
     There are separate digital outputs. Both signals use TTL level. The outputs are: Defibrillation Sync. 
    and Nurse Call.
    The Defibrillation Synchronization Signal is obtained from Synchronization connector X8, pin 2.
    The Nurse call is obtained from Coding element connector X4, pin 8.
    Defibrillation Sync indication is generated by ECG. When active, the signal is in state 1. After 10 ms 
    the signal is reset to state 0. New Defibrillation Sync is not generated before the indication is 
    deactivated. The delay from the R wave peak to the start of the signal is maximally 35 ms.
    Nurse Call indication is generated by red and yellow alarms. When activated, it is set to state 1 and 
    remains at that state until the alarm situation is over or 
    SILENCE ALARM key is pressed.
    The range of state 0 is from 0 to 0.8 V, and the range of state 1 is 2.8 to 5 V.
    Analog outputs
    Compact Anesthesia/Critical Care Monitor produces two analog real-time signals. The Direct ECG 
    signal is available in Coding element connector X4 pin 7 and Synchronization connector X8, pin 3. 
    The other signal is Invasive pressure output. It is available in Synchronization connector X8 pin 4.
    NOTE:  When the source of the selected analog output becomes invalid (for example, transducer is 
    disconnected), the last valid waveform information remains on the output and the numerical 
    information is no longer valid.
    Direct ECG 
    Delay (max.): 15 ms
    Gain ECG (out)/ECG (in): 1 V/1 mV 
    NOTE: The direct ECG out signal is not available with modules M-ESTP rev. 01, M-EST rev. 00 and 
    M-ETP rev. 00.
    Invasive pressure signal: P1
    Delay (max.): 25 ms
    Gain Signal (out) / Pressure (in): 1V/100 mmHg
    Pressure area 0…300 mmHg
    The Invasive pressure output signal is 1 V/100 mmHg, originally ranging from 0 to 300 mmHg, and 
    with a delay of approximately 25 ms. The signal requires an input impedance of 100 k.
    2.10.4 Printer connection
    For connecting a printer to the monitor, the serial data interface is used. When parallel printing is 
    desired, a serial to parallel converter (ref. 78030) is needed.
    For further information see Part I, Interfacing. 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    20
    Document no. M1031520-02                                                                           
    3 Service procedures
    3.1 General service information
    The field service of the S/5 Compact Anesthesia Monitor and S/5 Compact Critical Care Monitor is 
    limited to replacing faulty PC boards and mechanical parts. Return the faulty PC boards to GE 
    Healthcare for repair.
    GE Healthcare is always available for service advice. Please provide the unit serial number, full type 
    designation and a detailed fault description. 
    CAUTION Only trained personnel with appropriate tools and equipment shall perform the tests 
    and repairs outlined in this section. Unauthorized service may void the warranty of the 
    unit.
    3.2 Service check
    S/5 Compact Anesthesia Monitor, F-CM(REC)1
    S/5 Compact Critical Care Monitor, F-CMC(REC)1 
    Display Unit
    These instructions include complete procedures for a service check. The service should be performed 
    after any service repair. Additionally, the service check procedures can also be used for determining 
    possible failures.
    The procedures should be performed in ascending order.
    The instructions include a check form (“APPENDIX A”), which should be filled in when performing the 
    procedures.
    The symbol 
     in the instructions means that the check list should be signed after performing the 
    procedure.
    3.2.1 Recommended tools
    Tool Order No. Notes
    Power cord
    Screwdrivers
    MemCard - Menu for MemCard Option
    MemCard - Data for MemCard Option
    Anaesthesia Keyboard, K-ARKB
    K-ARK interface cable 881154
    Hemodynamic Module 
    						
    							Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
    21
                                              Document no. M1031520-02
    3.2.2 Recommended parts
    3.2.3 Before beginning
    •
    Make sure that the monitor is switched to standby.
    •Disconnect the power cord and press the service reset switch until the LED next to it turns off.
    •If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable from 
    the Frame Unit. Take out any memory cards, if installed.
    •Detach the Display Unit from the Frame Unit.
    •Open the Display Unit by detaching the Front Unit.
    NOTE: Wear a static control wrist strap when handling PC boards. Electrostatic discharge may 
    damage components on the board.
    3.2.4 Visual inspection
    1. DC/DC power supply board
    Disconnect the DC/DC power supply board. Check the Display Unit – Frame Unit connector 
    under the board:
    • screws are tightened properly
    • the connector is clean and intact
    
    2. Central Processing Board
    Check the Central Processing Board in the Display Unit:
    •screws are tightened properly
    •cables are connected properly
    •all ICs that are on sockets are attached properly
    
    3. DC/DC power board
    Reconnect and check the DC/DC Power Board:
    •screws are tightened properly
    •all connectors are clean and intact
    •the cables are attached properly
    
    4. Net Unit and Synchronization Unit
    Check that: Part Order No. Notes
    Display Unit Fan filter  897010
    Recorder paper 74205 for recorder unit 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    22
    Document no. M1031520-02                                                                           
    •screws are tightened properly
    •the connector pins are intact
    
    5. Front Unit
    Check that:
    •the Connector Board screws are tightened properly and the cable connectors are clean 
    and intact
    •the Inverter Board screws are tightened properly and the cable connectors are clean and 
    intact
    •the Backlight cables are intact and attached properly
    •the keyboard foil cables are intact and attached properly to connectors on the Connector 
    Board
    •the Connector Board foil cable is intact and attached properly to the connector on the 
    Central Processing Board
    •the video signal foil cable is intact and attached properly to its connector on the Central 
    Processing Board
    Reattach the Front Unit.
    
    6. External parts
    Check external parts of the Display Unit:
    •all stickers are intact
    •the LCD display screen is intact
    •the ComWheel cover is intact and attached properly
    •all mechanical parts are functioning properly
    Connect the Display Unit to the Frame Unit.
    
    7.  Inserting module 
    Check that the module used goes in smoothly and locks up properly in both module slots.
    Leave the module connected. 
    
    8. Fan filter
    Clean or replace the Display Unit fan filter.
     
    						
    							Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
    23
                                              Document no. M1031520-02
    3.2.5 Functional inspection
    9. Stand-by LED
    Connect the power cord.
    Check that the standby LED starts blinking or is on continuously, if the batteries do not need 
    charging.
    
    10. Start-up
    Switch the monitor on.
    Check that the monitor starts up properly, i.e. the alarm LEDs turn blank, the start-up sound is 
    heard from the loudspeaker and the normal monitoring screen appears.
    No error messages should appear on the screen. 
    If the monitor contains the recorder unit, two lines of start-up information should be recorded.
    
    11. Monitor screen configuration
    Configure the monitor screen so that all the needed parameter information is shown:
    Monitor Setup
    Screen 1 Setup
    Waveform Fields
    Digit Fields
    Check that the selected parameter information is shown on the screen.
    NOTE: InvBP waveforms are not shown without a patient simulator. Do no connect the simulator to the 
    monitor at this stage.
    
    12. Mark Event or Take Snapshot
    Push the menu keys Help and Mark Event, or Take Snapshot.
    Check that the picture on the LCD display screen is straight, clear and stable.
    
    13. Time and date
    Check that the time and date are correct, adjust them if necessary.
    Monitor Setup
    Set Time and Date
    NOTE: The selection SET TIME AND DATE cannot be accessed if a case is started, or a patient has 
    been admitted.
    NOTE: If the clock shows time 0:00 continuously (at successive start-ups), the SRAM/TIMEKEEPER 
    battery needs to be replaced. 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    24
    Document no. M1031520-02                                                                           
    The FACTORY RESET should be performed after the battery replacement, see “Performing 
    Factory Reset” .
    
    14. Loudspeaker
    Check the loudspeaker by setting the alarm sound:
    Alarms Setup
    Alarm Volume
    Test the whole volume scale from 1 to 10 by turning the ComWheel and check that the alarm 
    volume changes correspondingly. The alarm sound should be clear and audible with all the 
    settings.
    
    15. Display brightness
    Check the display brightness control: 
    Monitor Setup
    Display Brightness
    Press the ComWheel and check that the display brightness follows the selected brightness.
    
    16. Monitor software
    Enter the service menu:
    Monitor Setup
    Install Service (Password 16-4-34)
    Service (Password 26-23-8)
    Take down the information regarding Monitor and Power Management Board software.
    
    17. Content of service log
    Select Service Log from the menu. 
    Check the contents of the Service Log for possible problems. If the monitor contains the 
    recorder unit, record the Service Log on the recorder by selecting Record Log.
    
    18. Power supply output voltages
    Check the power supply unit output voltages through the service menu:
    Monitor Setup
    Install Service (Password 16-4-34)
    Service (Password 26-23-8)
    Frame
    Power Supply
    The Display Unit voltages should meet the following ranges: 
    						
    							Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
    25
                                              Document no. M1031520-02
    
    19. Watchdog circuitry
    Test the watchdog circuitry:
    Monitor Setup
    Install Service (Password 16-4-34)
    Service (Password 26-23-8)
    Set/Test
    Perform the tests WATCHDOG and WD BY OVERLOAD. Check that the monitor restarts in each 
    case.
    NOTE: When selecting WD BY OVERLOAD, restarting should take place approximately after
    15 seconds.
     
    Preset monitor settings: 
    Record/Print
    Record Waveforms
    Waveform 1 --> ECG1
    Waveform 2 --> P1
    Waveform 3 --> P2
    Delay --> Off
    Paper Speed --> 6.25 mm/s
    Length --> 30 s
    20. Alarm LEDs
    Enter the Keyboard service menu:
    Monitor Setup
    Install Service (Password 16-4-34)
    Service (Password 26-23-8)
    Keyboard
    Select the text Upper Led. Check that the red alarm LED(s) is turning on and off when pressing 
    the ComWheel. Check also the yellow alarm LED(s) by selecting Lower Led from the menu.
    
    21. Membrane keys check
    Press the keys on the monitor command board one by one. Check that each key generates a 
    sound from the loudspeaker. Also, the corresponding text in the menu should change from 
    yellow to red.+12V 11.50...12.50 V
    +3.3 V 3.00...3.60 V
    Vbus 10.00…17.00V
    Depends on the power source 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    26
    Document no. M1031520-02                                                                           
    Check also the functioning of the side panel keys.
    
    22. ComWheel check
    Turn the ComWheel clockwise and counter clockwise and check that each step generates a 
    sound from the loudspeaker and the corresponding values at the bottom of the menu increase.
    Select Dummy Press. Press the ComWheel and check that the press generates a sound and the 
    corresponding value in the menu increases.
    
    23. K-ARK connector
    Check that the command board section of the Anesthesia Keyboard functions normally through 
    the K-ARK connector.
    
    24. Module communication
    Connect a patient simulator to the module.
    Enter the ESTP: ECG service menu:
    Monitor Setup
    Install Service (Password 16-4-34)
    Service (Password 26-23-8)
    Parameters
    ESTP: ECG
    Check that the Timeouts, Bad checksums and Bad c-s by mod values do not increase faster 
    than by 5 per second.
    Change the module into the other slot. When the module is recognized, check that the values do 
    not still increase faster than by 5 per second.
    
    25. Battery
    Enter the Battery service menu:
    Monitor Setup
    Install Service (Password 16-4-34)
    Service (Password 26-23-8)
    Frame
    Power Supply
    Battery
    Check that the information on the page is correct.
     
    						
    							Display Unit for F-CM(C)1 (Rev. 04), F-CM(C)REC1 (Rev. 01)
    27
                                              Document no. M1031520-02
    26. Battery voltage
    Check that the shown Battery Voltage value is within 10 – 17V.
    
    27. Monitor temperature
    Check that the shown Monitor Temperature value is within 20 - 40.
    
    28. Battery operation
    Disconnect the power cord (without switching the monitor to standby).
    Check that the monitoring continues normally. The battery indicator should appear in the upper 
    right hand corner of the screen.
    
    29. Reconnecting the power cord
    Reconnect the power cord.
    Check that the monitoring continues without problems. The battery indicator should be 
    replaced by the text CHARGE.
    
    Recorder Unit
    30. Cover open recognition
    Open the recorder unit cover. Check that the message ‘Recorder: Cover open’ appears on the 
    screen, then close the cover.
    
    31. Recording
    Press the key 
    Record Wave and check that the module starts recording the selected 
    waveforms. Press the key 
    Stop on the module to stop recording.
    NOTE:  InvBP waveforms are not recorded without a patient simulator.
    NOTE: If no recording appears, check that the paper roll is installed in a correct way. Only one side of 
    the paper is printable.
    
    32. Paper speed
    Press the key 
    Record Wave again and this time wait until the recording stops. Check that the 
    length of the recorded waveform scale is 18.7 cm (± 1.5 cm).
    Change the paper speed setting to 1 mm/s: 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    28
    Document no. M1031520-02                                                                           
    Record/Print
    Record Waveforms
    Paper Speed --> 1 mm/s
    Press the key 
    RECORD WAVE and wait until the recording stops. Check that the length of the 
    scale is now 3.0 cm (± 0.5 cm).
    
    33. Quality of recording
    Check that the quality of the recording is acceptable.
    
    MemCard option 
    NOTE: The following steps are applicable only when the MemCard option is installed in the monitor.
    34. Memory option recognition
    Enter the Memory Card service menu:
    Monitor Setup
    Install Service (Password 16-4-34)
    Service (Password 26-23-8)
    Frame
    MemCards
    Check that the memory option is recognized properly, i.e. Present and Active state YES.
    
    35. Memories and PCMCIA controller test
    Check that the Memory board memories and the PCMCIA controller have passed their tests. The 
    status for each should be OK.
    
    36. Interface status
    Select Communication. Check that Interface status states ACTIVE continuously and the error 
    counter values at the bottom part of the menu are stable.
    
    37. Menu Card recognition
    Select Module Status. Insert a Memory card labelled Menu in the left hand side memory card 
    slot. Check that the message ‘Menu Card inserted’ appears in the message field and the white 
    menu card symbol in the upper right hand corner of the screen within 1 minute.
    NOTE:  The battery symbol overrides the memory card symbols.
      
    						