    							Appendix A, Service check form, S/5 Compact Anesthesia Monitor, S/5 Compact Critical Care Monitor
    A-3
    Document no. M1031520-02
    MemCard Option S/N
          OK          N.A.         Fail     OK          N.A.         Fail
    34. Memory option 
    recognition                           35. Memories and PCMCIA 
    controller test                           
    36. Interface status
                               37. Menu Card recognition
                               
    38. Card type recognition
                               39. Data Card recognition
                               
    40. Save and Load Modes
                               41. Data Card functions
                               
    Notes
     
     
    Network Option S/N
          OK          N.A.         Fail     OK          N.A.         Fail
    42. Connection to Network
                               43. Ethernet address ___________________
    44. Data counters
                               45. Data error counters
                               
    46. Hardware error counters
                               47. Disconnection 
    recognition                           
    48. ID-plug recognition
                                                        
    Notes
     
     
    Wireless LAN option  S/N
          OK          N.A.         Fail     OK          N.A.         Fail
    49. Signal strength
                               50. Configure access point  ___________________
    51. Connection to the 
    Network                                                    
    Notes
     
      
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    A-4
    Document no. M1031520-02
    General 
     
         OK          N.A.         Fail     OK          N.A.         Fail
    52. Trend retaining
                               53. Service reset switch 
    check                           
    54. Battery capacity
                               55. Service Log reset
                               
    56. Electrical safety check
                               57. Final cleaning
                               
    Notes
     
     
     
     
     
     
     
     
    Used spare parts 
    Signature   
    						
    							All specifications are subject to change without notice.
    CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
    Outside the USA, check local laws for any restriction that may apply.
    Document no. M1044470-01
    March, 2006
    Datex-Ohmeda Inc.
    P.O. Box 7550
    Madison, WI 53707-7550, USA
    Tel: +1 608 221 1551
    Fax: +1 608 222 9147
    GE Healthcare Finland Oy
    Helsinki, Finland
    P.O. Box 900
    FI-00031 GE, FINLAND
    Tel: +358 10 39411
    Fax: +358 9 1463310
     www.gehealthcare.com
    Copyright  2006 General Electric Company
    Datex-Ohmeda
    S/5
    TM Compact Anesthesia Monitor,
    F-CM1 (Rev. 04), F-CMREC1 (Rev. 01); Frame Unit
    AC/DC Power Supply Unit
    S/5
    TM Compact Critical Care Monitor,
    F-CMC1 (Rev. 04), F-CMCREC1 (Rev.01); Frame Unit
    AC/DC Power Supply Unit
    Technical Reference Manual 
    						
    							Table of contents
    i
                                                                                      Document no. M1044470-01
    Table of contents
    Table of contents i
     Introduction 1
    1 Specifications 2
    2 Functional description 3
    2.1 PMB connector board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
    2.2 Connector board PMB-REC  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
    2.3 PMB (Power Management Board) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
    2.4 Module motherboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
    2.5 AC/DC Power Supply Unit  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
    2.6 Batteries  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
    2.7 Frame Unit Fan  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
    2.8 In-built Recorder Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
    2.9 Connectors and signals  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
    2.9.1 External Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
    2.9.2 Serial Data Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
    3 Service procedures 11
    3.1 General service information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
    3.2 Service check  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
    3.2.1 Recommended tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
    3.2.2 Recommended parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
    3.2.3 Before beginning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
    3.2.4 Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
    3.2.5 Functional inspection  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
    3.2.6 Recorder Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
    3.2.7 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
    3.3 Disassembly and reassembly  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
    3.3.1 To remove the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
    3.3.2 To remove the AC/DC Unit  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
    3.3.3 To disassemble the Side Frame Unit  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
    3.3.4 To detach/remove the PMB Board (Power Managemet Board) . . . . . . . . . . . . . . . . . . . . . . .19
    3.3.5 To detach/remove the Module Mother Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
    3.3.6 To remove the Recorder Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
    3.3.7 Adjustment and Calibrations  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
    3.4 Internal battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
    3.4.1 Battery time  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
    3.4.2 Refreshing the batteries  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
    3.4.3 Replacing the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
    4 Troubleshooting 23
    4.1 General  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
    4.2 Recorder Unit  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
    4.2.1 Troubleshooting chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
    5 Earlier revisions 26 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    ii
    Document no. M1044470-01                                                                    
    Appendix A: Service check form, S/5 Compact Anesthesia Monitor, 
    S/5 Compact Critical Care Monitor A-1 
    						
    							Frame Unit for F-CM(C)1 (rev. 04), F-CM(C)REC1 (Rev. 01)
    1
                                              Document no. M1044470-01
     Introduction
    This section provides information about the maintenance and service of the following products:
    •S/5 Compact Anesthesia Monitor frame F-CM1 and S/5 Compact Critical Care Monitor frame 
    F-CMC1 (including the AC/DC power unit, battery and Power Management Board). 
    •S/5 Compact Anesthesia Monitor frame F-CMREC1 and S/5 Compact Critical Care Monitor 
    frame F-CMCREC1 (in addition to the above mentioned parts also a built-in recorder).
    All the service menus related to the monitor are described in a separate “Service Menu” slot and the  
    spare part lists in the “Compact Monitors Spare Parts” slot.
    NOTE: Printings on thermal paper may be destroyed when exposed to light, heat, alcohol, etc. Take a 
    photocopy for archive. 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    2
    Document no. M1044470-01                                                                           
    1 Specifications
    Dimensions 335 x 205 x 340 mm /13.2 x 8.1 x 13.4in
    (W x D x H)
    Weight  8.5 kg (19.4 lbs) (w/o modules)
    Electrical Requirements
    Power supply
    Rated voltage range and frequencies 100 to 240 VAC 50/60 Hz 
    (limits 90-264 VAC)
    Stability ±10 % of nominal voltage
    Power consumption 140 VA (max)
    Fuses 2x2A, T/SLOW, 250 V
    Grounding Hospital grade
    Interruptability Data memory and alarm settings are saved up to 15 minutes 
    after monitor standby
    Battery
    Battery type NiMH, 3.5Ah 
    Battery charging time  10 hours to full capacity
    Battery operation up to 1.5 hours typical on full charge at 23 °C / 73 °F
    Recharging  when connected to mains power
    Battery operaton alarm
    early warning ‘Battery low’ note
    final warning  ‘Battery low’ alarm
    Battery state indicator Capacity bar displayed on the screen
    Environmental Requirements
    Operating temperature 10 - 35 °C / 50 - 95 °F
    Storage temperature -10 - 50 °C / 14 - 122 °F
    Atmospheric pressure 500 - 800 mmHg (660 - 1060 mbar, 66 - 106kPa)
    Humidity 10 - 90 % non-condensing
    In-built recorder
    Power consumption 3 W
    Principle Thermal array
    Print resolution
    Vertical 8 dots/mm (200 dots/inch)
    Horizontal 32 dots/mm (800 dots/inch) at a speed of 25 mm/s and 
    slower
    Paper width 50 mm, printing width 48 mm
    Traces Selectable 1, 2, or 3 traces
    Print speed 1, 6.25, 12.5, 25 mm/s 
    						
    							Frame Unit for F-CM(C)1 (rev. 04), F-CM(C)REC1 (Rev. 01)
    3
                                              Document no. M1044470-01
    2 Functional description
    The Frame Unit consists of the following parts:
    •PMB (Power Management Board)  
    •PMB connector board(s) 
    •connector board PMB-REC (in the side frame unit) 
    •two batteries 
    •AC/DC power unit 
    •optionally in-built recorder
    The left side of the Frame Unit has places for two double-width modules or four single-width modules. 
    The network signals are led through their own cables to the Frame Unit. The connectors are placed on 
    the left side of the Frame Unit in frame version F-CM(C)1 (rev.04). In frame versions F-CM(C)(REC) 
    (rev.00...03) the connectors are placed on the back plate of the Frame Unit. 
    The serial data interface is led through the Frame Unit to the AC/DC power unit. The serial connector is 
    located at the side of the AC/DC power unit. 
    						
    							Datex-Ohmeda S/5 Compact Anesthesia and Compact Critical Care Monitors
    4
    Document no. M1044470-01                                                                           
    Figure 1 Frame block diagram
    2.1 PMB connector board
    The PMB connector boards are small boards connected together by a ribbon cable and are both 
    equipped with the D-25 connectors, one for AC/DC unit and the other for the Display Unit. The board 
    with the Display Unit connector has also a 16-pin connector for the PMB board. 
    The PMB connector boards route signals between the AC/DC unit, PMB board and Display Unit. 
    Signals are transferred via a ribbon cable between the D-connector boards (connected to the AC/DC 
    and DU). Power lines are transferred separately through the PMB.
    2.2 Connector board PMB-REC 
    The connector board PMB-REC is the one mounted in the Side Frame unit, and connected to the PMB 
    board via a 15-pin D-connector when inserted into the frame. The connector board PMB-REC routes 
    operating voltage and signals for the frame cooling fan and for the recorder unit (optional).
    PMB board
    Battery #1Battery #2
    Fan
    Sideplate connector board
    Recorder
    25-pin D-connector25-pin D-connector
    CM1 frame
    Module
    motherboard
    Module
    connectors
    CM1_frame_block_diag.vsd 
    						