Page 101

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-41
3-15-6 If the System is NOT on a Network
If the system is connected to a network, skip this section. 
If the system is NOT on a network, a static dummy IP-Address must be install in Windows for DICOM 
and Daignostics to function.
To install a Static IP-Address in windows:
1.) Boot up the system with the service dongle attached to the USB port.
2.) From the Start Applications...

Page 102

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-42 Section 3-15 - Optional Software Procedures
7.) IF the local Area Connections Status Window is displayed, as shown in Figure 3-35, click on 
PROPERTIES
.
8.) Hightlight INTERNET PROTOCOL (TCPIP)
 and select PROPERTIES as shown in Figure 3-
36.
9.) Select Use this following IP address.  Fill in:
* IP-Address: 3.28.28.28
* Subnet Mask : 255.255.252.0Figure 3-35  Connection Status Window
Figure 3-36...

Page 103

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-43
Figure 3-37  Static Network Settings in Windows 

Page 104

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-44 Section 3-16 - Installing USB Line Printer - HP970/990
Section 3-16
Installing 
USB Line Printer - HP970/990
1.) Install the printer after the system has been ghosted, but before application software has been 
installed. Just plugging in the USB cable is enough to install the printer.
2.) If a service key is available, enter maintenance mode and exit to the desktop. Plug in the USB cable 
and the...

Page 105

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-45
Section 3-17
Paperwork
NOTE: During and after installation, the documentation (i.e. User Manuals, Installation Manuals...) for the 
peripheral units must be kept as part of the original system documentation. This will ensure that all 
relevant safety and user information is available during the operation and service of the complete 
system.
3-17-1 Product Locator Installation...

Page 106

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-46 Section 3-17 - Paperwork
This page was intentionally left blank. 

Page 107

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 4 Functional Checks 4-1
Chapter 4
Functional Checks
Section 4-1
Overview
4-1-1 Purpose of Chapter 4
This chapter provides procedures for quickly checking major functions of the LOGIQ™ 9 scanner 
diagnostics by using the built-in service software, and power supply adjustments.
Section 4-2
Required Equipment
• An empty (blank) MO Disk.
• At least one transducer. See “Probes” on page 1-41. for an...

Page 108

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
4-2 Section 4-3 - General Procedure
Section 4-3
General Procedure
4-3-1 Power On / Boot Up
NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be 
able to boot if power is recycled too quickly.
4-3-1-1 Scanner Power On
1.) Connect the Power Cable to the back of the system.
2.) Ensure the Cable Clip slips securely over the shoulders on the molded...

Page 109

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 4 Functional Checks 4-3
4-3-1-2 Back-end Processor Boot Up
Press the Standby switch on the Control Panel once.
When the Standby switch on the Control Panel is pressed once, the Back-end Processor starts and the 
software code is distributed to initiate the scanner.
No status messages are displayed during this process.
4-3-2 Power Off / Shutdown
NOTE: After turning off a system, wait at least ten...

Page 110

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
4-4 Section 4-3 - General Procedure
4-3-2-2 Scanner Shutdown
1.) From the rear of the system. switch OFF the Circuit Breaker.
2.) Disconnect the Mains Power Cable by unfastening the metal wire fastener.Figure 4-4   Circuit Breaker and Power Cable on Back of Scanner
NOTICEDisconnecting the Mains Power Cable before switching OFF the Circuit Breaker will activate the 
uninterruptible power system (UPS) in...