Page 61

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-1
Chapter 3
Installation
Section 3-1
Overview
3-1-1 The Purpose of Chapter 3
This chapter contains information needed to install the unit. Included are procedures to receive, unpack 
and configure the equipment.
Table 3-1    Contents in Chapter 3
SectionDescriptionPage Number
3-1
Overview3-1
3-2
Installation Reminders3-2
3-3
Receiving and Unpacking the Equipment3-4
3-4
Preparing...

Page 62

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-2 Section 3-2 - Installation Reminders
Section 3-2
Installation Reminders
3-2-1 Average Installation Time
 
3-2-2 Installation Warnings
1.) Since the LOGIQ™ 9 weighs approximately 225 kg (500 lb.) without options, preferably two people 
should unpack it. Two people are also preferable for installing any additional items in excess of 35 
pounds.
2.) There are no operator serviceable components. To...

Page 63

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-3
3-2-3 Safety Reminders
DANGERWHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE 
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH 
THE UNIT!
CAUTIONTwo people should unpack the unit because of its weight. Two people are required whenever 
a part weighing 19kg (35 lb.) or more must be lifted.
CAUTIONIf the unit is very cold or hot, do not turn on its...

Page 64

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-4 Section 3-3 - Receiving and Unpacking the Equipment
Section 3-3
Receiving and Unpacking the Equipment 
 
CAUTIONHave two people available to unpack the LOGIQ™ 9. Attempts to move the unit considerable 
distances or on an incline by one person could result in injury or damage or both.
Table 3-4    Unpacking Procedure
StepTask
1.
Open the four hinges on each door
2.
Remove the doors at both ends of the...

Page 65

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-5
Section 3-4
Preparing for Installation
3-4-1 Verify Customer Order
Compare items received by the customer to that which is listed on the delivery order. Report any items 
that are missing, back ordered or damaged.
3-4-2 Physical Inspection
3-4-2-1 System Voltage Settings
Verify that the scanner is set to the correct voltage.
The Voltage settings for the LOGIQ™ 9 Scanner is found...

Page 66

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-6 Section 3-5 - Completing the Installation
Section 3-5
Completing the Installation
3-5-1 Power On / Boot Up
NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be 
able to boot if power is recycled too quickly.
3-5-1-1 Scanner Power On
1.) Connect the Power Cable to the back of the system.
2.) Ensure the Cable Clip slips securely over the...

Page 67

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-7
3-5-1-2 Back-end Processor Boot Up
Press the Standby switch on the Control Panel once.
When the Standby switch on the Control Panel is pressed once, the Back-end Processor starts and the 
software code is distributed to initiate the scanner.
No status messages are displayed during this process.
3-5-2 Power Off/ Shutdown
NOTE: After turning off a system, wait at least ten seconds...

Page 68

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-8 Section 3-5 - Completing the Installation
3-5-2-2 Scanner Shutdown
1.) Switch OFF the Circuit Breaker at the back of the system.
2.) Disconnect the Mains Power Cable is necessary. For example: Relocating the scanner.
3-5-3 Transducer Connection
1.) Connect a transducer to one of the three rightmost transducer receptacle as follows:
A.) Ensure that the transducer twist lock lever to the horizontal...

Page 69

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-9
Section 3-6
System Configuration
3-6-1 System Specifications
3-6-1-1 Physical Dimensions Physical Dimensions of 
The physical dimensions of the LOGIQ™ 9 unit are summarized in Ta b l e 3 - 1 and Ta b l e 3 - 2.
Table 3-1 lists the size of LOGIQ™ 9, with monitor and on-board peripherals.
3-6-1-2 Weight without Monitor and Peripherals
3-6-1-3 Acoustic Noise Output:
Less than...

Page 70

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-10 Section 3-6 - System Configuration
3-6-4 External I/O Connector Panel
.i.External I/O Connector Panel;
 
Figure 3-7   External I/O Panel Labels
Figure 3-8   External I/O Panel Connectors
123467
910111213148
5 