Chapter 2 Pre-Installation 2-1
Chapter 2
Pre-Installation
Section 2-1
Overview
2-1-1 Purpose of Chapter 2
This chapter provides the information required to plan and prepare for the installation of a LOGIQ™ 9. 
Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit.
Table 2-1    Contents in Chapter 2
SectionDescriptionPage Number
2-1
Overview2-1
2-2
General...

Page 52

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
2-2 Section 2-2 - General Console Requirements
Section 2-2
General Console Requirements
2-2-1 Console Environmental Requirements
2-2-1-1 Cooling
The cooling requirement for the LOGIQ™ 9 is 3500 BTU/hr. This figure does not include cooling needed 
for lights, people, or other equipment in the room. Each person in the room places an additional 300 
BTU/hr. demand on the cooling system.
2-2-1-2 Lighting...

Page 53

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 2 Pre-Installation 2-3
2-2-2-3 Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be readily accessible.
2-2-2-4 Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension cords. Other adequate 
outlets for the external peripherals, medical and test equipment needed to support this unit must also 
be present within 1 m (3.2...

Page 54

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
2-4 Section 2-2 - General Console Requirements
2-2-3 EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, 
magnetic fields, and transients in the air or wiring. Ultrasound machines also generate EMI. The 
LOGIQ™ 9 complies with limits as stated on the EMC label. However, there is no guarantee that 
interference will not occur in a particular...

Page 55

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 2 Pre-Installation 2-5
2-2-4 Scan Probe Environmental Requirements
Operation:10° to 40° C
Storage:-20° to 50° C
NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32).
2-2-5 Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six 
weeks before delivery, for enough time to make necessary...

Page 56

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
2-6 Section 2-3 - Facility Needs
Section 2-3
Facility Needs
2-3-1 Purchaser Responsibilities
The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, 
confusion, and waste of manpower can be avoided by completing pre installation work before delivery. 
Use the Pre installation checklist to verify that all needed steps have been taken. Purchaser 
responsibility...

Page 57

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 2 Pre-Installation 2-7
2-3-2 Required Features
NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its 
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before 
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other...

Page 58

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
2-8 Section 2-3 - Facility Needs
2-3-4 Minimal Floor Plan Suggestion
Figure 2-1   Minimal Floor Plan  

Page 59

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 2 Pre-Installation 2-9
2-3-5 Networking Pre-installation Requirements
2-3-5-1 Purpose of the DICOM Network Function
DICOM services provide the operator with clinically useful features for moving images and patient 
information over a hospital network. Examples of DICOM services include the transfer of images to 
workstations for viewing or transferring images to remote printers. As an added...

Page 60

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
2-10 Section 2-3 - Facility Needs
This page was intentionally left blank. 
Start reading GE Logiq 9 Service Manual

Related Manuals for GE Logiq 9 Service Manual

All GE manuals