Home > GE > Ultrasound System > GE Logiq 9 Service Manual

GE Logiq 9 Service Manual

Here you can view all the pages of manual GE Logiq 9 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 71

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-11
3-6-4External I/O Connector Panel (cont’d)
 
Table 3-5    External I/O Connector Descriptions
ItemConnector NameTable NumberDescription
1
Audio Output (Left)RCA Phono Jack
2
Audio Output (Right)RCA Phono Jack
3
Trigger OutputBNC Connector
4FootswitchTable 3-12Hirose Electric Co. RM12BRB-5S,
Footswich transport shelf is in the back at the bottom
5RS232 Com1DSUB 9 pin male...

Page 72

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-12 Section 3-6 - System Configuration
3-6-4-1 External I/O Pin Outs
 
 
 
 
Table 3-6    S-Video Connector, 4 Pin
Pin NoOutput SignalDescription
1 SVIDEO OUT/IN YG Y (Luma) GND
2 SVIDEO OUT/IN CG C (Chroma) GND
3 SVIDEO OUT/IN Y Y (Luma) SIGNAL
4SVIDEO OUT/IN CC (Chroma) SIGNAL
Table 3-7    S-VGA Connector, Shrinked Sub-D 15 Pin
Pin NoOutput SignalDescription
1 VGA OUT1 R Red
2  VGA OUT1 G Green
3 VGA...

Page 73

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-13
3-6-4-1External I/O Pin Outs (cont’d)
Table 3-10    Modem, RJ11 Modular 6 Pin
Pin NoOutput SignalDescription
2 TEL L4 Telephone L4
3 TEL L2 Telephone L2
4 TEL L1 Telephone L1
5TEL L3Telephone L3
OthersNCNon-connection
Table 3-11    Remote Expose 1 & 2 Connectors
Pin NoOutput SignalDescription
1 SHUTTER/PRINT
2 Signal GND
Table 3-12    Footswitch Connector
Pin NoOutput...

Page 74

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-14 Section 3-6 - System Configuration
3-6-5 Video Specification
Video specifications may be needed to be able to connect laser cameras or other devices to the 
LOGIQ™ 9.
Table 3-13    Video Specifications
Timing Parameter
1024x768
75Hz768x576
50Hz640x480
60Hz800x600
60Hz800x600
75Hz1024x768
60Hz
Horizontal Rate [kHz] 60.02 31.25 31.47 37.88 46.88 48.36
Horizontal Period [µs]16.66 32.00 31.78 26.40...

Page 75

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-15
Section 3-7
Available Probes
See Specifications in the LOGIQ™ 9 User Reference Manual for probes and intended use.
See Chapter 9 - Renewal Parts, for part numbers to be used when ordering new or replacement 
probes.
Section 3-8
Software/Option Configuration
Refer to the LOGIQ™ 9 Basic User Manual, Chapter 16, Customizing Your System for information on 
configuring items like...

Page 76

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-16 Section 3-9 - Connectivity Installation Worksheet
Section 3-9
Connectivity Installation Worksheet
Site System Information
Services (Destination Devices)TCP/IP Settings
Device Type
Manufacturer
Name
IP AddressPort
AE Title
1
2
3
4
5
6
7
8
9
10
11
12
Name - AE Title:
Dept:
LOGIQ SN:Type:
Floor:
Room:
REV:
Comments:
CONTACT INFORMATION
Name
Title
Phone
E-Mail Address
Site:
IP Settings
IP Address:...

Page 77

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-17
Section 3-10
Insite IP Address Configuration
This information needed to facilitate the OnLine Center engineers in connecting to the system is found 
in the InSite Installation Manual, 2324712-100.
Section 3-11
Before Starting a Software Upgrade
3-11-1 Printers
In R2.1.0 software and greater, if the customer has an HP990, HP970, UP-D50, UP-D895, UP-D21 or 
Talley printers, any...

Page 78

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-18 Section 3-12 - Loading Base Image Software
Section 3-12
Loading Base Image Software
NOTE: While it is believed to be unnecessary, It would not hurt to disconnect the system from the 
network and remove all transducers.
NOTE: If the user has an HP990, HP970, UP-D50, UP-D895, UP-D21 or Talley printers, connect the 
printer and turn it on. If a printer is plugged in at this time it will automatically...

Page 79

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
Chapter 3 Installation 3-19
Section 3-12      Loading Base Image Software (cont’d)
Now is your chance to abort the Base Image load or continue.
5.) Press any key to continue. To abort, remove the CD-ROM from the drive and press CTRL-C
. Figure 3-10  Base Load Warning Message 

Page 80

GE MEDICAL SYSTEMSPROPRIETARY TO GE
D
IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
3-20 Section 3-12 - Loading Base Image Software
Section 3-12      Loading Base Image Software (cont’d)
The first choice is to load the complete hard drive.
1.) Load the complete disk - All existing data will be lost.
NOTE: APPLICATION SOFTWARE WILL NEED TO BE INSTALLED WHEN THIS PROCEDURE 
HAS COMPLETED. ALL PATIENT DATA (IF ANY) WILL BE DESTROYED!
Figure 3-12 is the last warning message before the...
Start reading GE Logiq 9 Service Manual

Related Manuals for GE Logiq 9 Service Manual