Home > GE > Medical Equipment > GE Logiq 7 Service Manual

GE Logiq 7 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 394
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 4-2 - General Procedure 4-7
    4-2-3 Power Shutdown
    Purpose: This is a description on how to Shutdown the system.
    4-2-3-1 Complete Power Down 1.) Press the ON/OFF
     key at the front of the System for about two (2) seconds. Refer to Figure 4-2.
    2.) Switch OFF the Main Circuit Breaker at the rear of the system Refer to Figure 4-1.
    3.) Disconnect the Main Power Cable if needed. Refer to Figure 4-1.
    4-2-4 System Stand-by
    LOGIQ 7 is available with Stand-by option.  The option is effective in reducing the scanner power up 
    time, when used as portable device.
    Users must follow the procedures in order to use the option.
    NOTE: Do not use more than 20 Stand-by mode continuously without shutting down the system.  The  scanner must be powered off daily.
    1.) Press the POWER switch once.
    2.) Select Standby .At this point, both the touch panel and the main monitor will go blank.  Keyboard 
    lights also turns dark.  Power switch button will be illuminated.
    3.) Wait more than 20seconds.
    4.) Plug off the power cord from wall outlet.
    5.) Transport the scanner as required, but re-connect to power outlet within 20 minutes.  If not powered  system may shut down due to loss of sustained battery power.
    6.) Connect the scanner to a wall electrical outlet. Figure 4-10   Selecting Standby
    NOTICEDo not disconnect the power cable immediately.
    NOTICEDo not press the Power button without connecting the power cable.
    Standby 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    4-8Section 4-2 - General Procedure
    4-2-4 System Stand-by (cont’d)
    7.) To power up the scanner, tap POWER button momentary, and wait for power up.  It will  approximately 45seconds for the scanner to power up.
    Stand-by Sequence
    NOTICEDo not press the Power button long time, as it will cause the system to power down.
    Figure 4-11   Stand-by Sequence
    Press POWER button
    Select "Standby"
    Ch e c k Mo n ito rOFFWait
    Check TouchPanel OFF
    Ch e c k POWER button
    illu min ate d
    Wait more than 20 seconds
    Disconnect Power Cord
    Connect to electrical outlet (Power Cord)
    Pre ss POWER bu tton
    Wait 5  se c o n ds
    Scanner Start up completionin 45 seconds
    S t an dby may be
    faile d
    Stand- by
    Se
    quence
    ON
    ON OFF
    OFF or Blinking
    (Note 1)
    Note1 : BT09 Release with BEP4 Hardware have Blinking Power Button 
    capability to indicate the console in Stand-by condition.  For BEP3 and earlier, 
    Power button must be OFF before disconnecting the power cord. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 4-2 - General Procedure 4-9
    4-2-5 Using CD-R/MOD/DVD Drive
    4-2-5-1 Using CD-R or DVD Drive
    1.) Push the EJECT
     button, the disk tray will appear.
    2.) Put the disk onto the disk tray.
    3.) Press the EJECT
     button to insert the disk into the CD-R or DVD device.
    4.) There are a number of methods to eject a disk from the CD-R or DVD. Ejection is automatic in some  cases. Manual ejection methods, listed in preferred order of use, are:
    a.) Press  EJECT
     button on the CD-R or DVD while system is ON.
    b.) Press and hold  EJECT
     button while the system is booting.
    c.) Mechanical ejection. Insert the end of a paper clip into the hole while system power is OFF.
    NOTE: Be careful not to scratch the disk when wiping it off for cleaning.
    NOTE: Keeping your CD-R or DVD disc in an original CD-R or DVD case or caddy all the time will  prevent it from becoming dirty or damaged.
    NOTE: Media Requirement:
    Please be aware of the following regarding saving data to DVD-R on R6.0.x.:
    Use x8 or x16 speed compatible DVD-R. When it is the upgraded R6.0.x system from R4.x.x. 
    or prior version, some systems have a number 4 or 5 printed on the disc tray of the media 
    drive and others dont. If there is no number printed on the disc tray, use x4 or x16 speed 
    compatible DVD-R.
    NOTICENever move the unit with a disk in the CD-R or DVD because the drive actuator will not be 
    locked and the CD-R or DVD could break. 
    NOTICEAvoid mechanical ejection whenever possible. Mechanical ejection leaves the actuator 
    unlocked and the MOD susceptible to damage if moved. If forced to use this method, reboot 
    the system, then insert and eject a known good disk using one of the other methods.
    Figure 4-12 CD-R or DVD drive
    CD-R Drive Eject Button
    Mechanical Eject Hole
    DVD Drive 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    4-10Section 4-2 - General Procedure
    4-2-5-2 Using MOD Drive
    1.) Before installing an MO disk in the MOD, check the MO disk for loose hardware or damaged labels which could jam inside the MO Drive. Also ensure that the slide switch in one corner of the disk is 
    set so that the disk is write enabled (disk hole closed).
    2.) Insert the disk into the MOD with the label facing up.
    3.) The are a number of methods to eject a disk from the MOD. Ejection is automatic in some cases.  Manual ejection methods, listed in preferred order of use, are:
    a.) Press  EJECT
     button on the MOD while system is ON.
    b.) Press and hold  EJECT
     button while the system is booting.
    c.) Mechanical eject. Insert the end of a paper clip into the hole while system power is OFF.
    NOTICENever move the unit with a disk in the MOD because the drive actuator will not be locked and 
    the MOD could break. 
    NOTICEAvoid mechanical ejection whenever possible. Mechanical ejection leaves the actuator 
    unlocked and the MOD susceptible to damage if moved. If forced to use this method, reboot 
    the system, then insert and eject a known good disk using one of the other methods.
    Figure 4-13 MOD drive
    MO Drive Eject Button
    Mechanical 
    Eject Hole
    MO Drive 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 4-2 - General Procedure 4-11
    4-2-6 Archiving and Loading Presets for BT07 (Including R7.5.x)
    NOTE: Always save presets before any software reload. This ensures the presets loaded after the software 
    reload are as up–to–date as possible.
    All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an CD-
    R/MO/DVD-R disk for reloading on the system.
    4-2-6-1 Regional Preset - General (Supported from BT07) This feature has the capability to have factory default preset defined by the following regions; Americas, 
    Europe, Asia, or Japan.
    NOTICEPresets should NOT be saved on the same CD-R/MO/DVD-R disk as images. The Archive Menu 
    lists the images but does NOT list the presets stored on a CD-R/MO/DVD-R disk.
    NOTICEDo not attempt to change/use Regional Preset buttons for Upgraded BT07.  For Upgrade BT07 always 
    use Factory Default in order to avoid preset conflict. Contact application specialist for details.
    Table 4-2 
    Presets Unique to RegionsPresets NOT unique to Regions
    System Imaging (System>System Imaging page)
    System Measure (System->System Measure page)
    Imaging Settings (Imaging page)
    Comments (Comments page)
    Body Patterns (Body Patterns page)
    Application (Application page)
    3D/4D (3D/4D page)
    Measure (Measure page) :  Utility/Measure Advanced and 
    Doppler preset System General (System>General page)
    Connectivity (Connectivity page)
    Reports (Report page)
    Measure (Measure page) : Utility/Measure M&A and OB 
    Table preset 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    4-12Section 4-2 - General Procedure
    4-2-6-2 Using Regional Preset
    Regional Preset is selected during Application Load (Refer to 8-8-13 - Installing R7.x.x Application 
    Software for details).
    Current Regional Preset is displayed under  Utility > System > About .
    Factory Default Regional Preset can be re-loaded from  Utility > Systme > About > Additional About.
    Figure 4-14   Current Regional Preset
    NOTICEFor Upgraded BT07, this field always shows None as Region Preset is feature available for pure BT07 
    and onward.
    Figure 4-15   Factory Default Regional Preset 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 4-2 - General Procedure 4-13
    4-2-6-3 Cautions Using Regional Preset
    • Do not attempt to modify Regional Preset on Upgraded BT07.
    • Backup/Restore function should be used between the same region systems.  Unexpected setting may result if you restore the preset files to another region setting system.
    4-2-6-4 Formatting CD-R/DVD-R Disk 1.) Insert an empty (blank) CD-R/DVD-R disk into the CD-R/DVD-R device.
    2.) Access to the Utility  Menu on the Touch Panel, and select  Connectivity>Removable Media .
    3.) Select the removable media from media list.
    4.) Type a name for the removable media in Label field.
    5.) Select Format button.
    CAUTIONWhen software is upgraded (from BT07 and on) re-loaded or upgraded (from BT07and on), make sure 
    to select the factory default.  Because the preset region information in globalconfig.res file shall not be 
    upgraded, if different region presets are loaded, then it may cause conflict between system setting and 
    region presets. 
    Figure 4-16   Selecting Format Button 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    4-14Section 4-2 - General Procedure
    4-2-6-5 Archiving Presets to an CD-R/DVD-R Disk
    1.) Insert an empty (blank) formatted CD-R/DVD-R disk into the CD-R/DVD-R device.
    2.) Access to the Utility  Menu on the Touch Panel, and select  System>Backup/Restore . The Backup 
    screen will be shown on the monitor.
    3.) Select the item to back up either from Resource Files.
    4.) Select the media to locate the items.
    5.) Click on  Backup. The backup status for each item is displayed on the Result column. 
    6.) Make sure “Finished OK” is displayed on the Result column. Figure 4-17   Clicking on Backup 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 4-2 - General Procedure 4-15
    4-2-6-6 Loading Presets from an CD-R/DVD-R disk
    1.) Insert the CD-R/DVD-R disk with the archived Presets into the CD-R/DVD-R.
    2.) Access to the Utility  Menu on the Touch Panel, and select  System>Backup/Restore. The Restore 
    sheet will be shown on the monitor.
    3.) Select the item to restore either from resource Files.
    4.) Click on  Restore. A message to make sure the restore process is displayed on the monitor. Click 
    OK . The restore status for each item is displayed on the Result column.
    5.) Make sure “Finished OK” is displayed on the Result column.
    4-2-6-7 Regional Preset - Files Service Tip: The system shall have each region preset files into:
    • Americas  : C:/Pegasus/target/resources/Pegasus/ Americas/userdefs/
    • Europe: C:/Pegasus/target/resources/Pegasus/Europe/userdefs/
    • Asia: C:/Pegasus/target/resources/Pegasus/ Asia/userdefs/
    • Japan: C:/Pegasus/target/resources/Pegasus/Japan/userdefs/
    When region is selected, Region preset files are copied to:
    • C:/Pegasus/target/resources/idunn/userdefs/ Figure 4-18   Clicking on Restore 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    4-16Section 4-2 - General Procedure
    4-2-7 Archiving and Loading Presets for BT06/V65x or lower
    NOTE: Always save presets before any software reload. This ensures the presets loaded after the software 
    reload are as up–to–date as possible.
    All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an CD-
    R/MO/DVD-R disk for reloading on the system.
    4-2-7-1 Formatting CD-R/MO/DVD-R Disk 1.) Insert an empty (blank) CD-R/MO/DVD-R disk into the CD-R/MO/DVD-R device.
    2.) Access to the Utility  Menu on the Touch Panel, and select  Connectivity>Removable Media or 
    Tools . 
    NOTICEPresets should NOT be saved on the same CD-R/MO/DVD-R disk as images. The Archive Menu 
    lists the images but does NOT list the presets stored on a CD-R/MO/DVD-R disk.
    Figure 4-19 Formatting Removable Media1 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Logiq 7 Service Manual