    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-15
    8-2-7 Install Base System (cont’d)
    4.) The following screen appears. Press 
    B then Enter to initialize the C drive
    5.) Check if the message appears as reloading C: partition only, then press Enter  key.
    If the message appears as reloading the entire disk, press  CTRL-ALT-DEL to reboot the sysytem 
    and retry from procedures described in previous page. Figure 8-65   Select B
    CAUTIONIf you select [A], all the archive would be lost.
    Figure 8-66   Enter 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    8-16Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
    8-2-7 Install Base System (cont’d)
    6.) If the following screen appears, verify that  OK buttton is selected, then press  Enter key. (If it is not 
    selected, select OK  button using the arrow keys then press Enter  key.)
    7.) Approx. fifteen minutes later, the following screen appears. Eject the  Base System Software Load 
    Image DVD from the drive.
    8.) To restart the system, press  Ctrl + Alt + DEL key at the same time. 
    Figure 8-67   OK
    Figure 8-68   Ejecting DVD
    NOTICEThe Base System Software Load Image DVD must be stored onto the PC box cover. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-17
    8-2-8 Modify System/Windows Settings
    After Base System is successfully installed, various parameters must be properly set before proceeding to Application 
    load.  The below outlines necessary system/windows setting.
    8-2-8-1 UPS Setting UPS automatically setup by Ghost. You do not need to setup it manually. You can see the message 
    below on the desktop.
    Please choose  OK.
    Table 8-6   System/Windows Settings
    ItemMethodProper Value
    UPS Setting
    Automatic Setting (n/a – just load)
    (Restart System)
    Avance AC97 Audio Load from c:\windows\system32\drivers (n/a – just load)
    (Restart System)
    Drive LetterAutomatic Setting but check desktop icon if not.DVD Drive = G
    HW DeviceMy Computer
        
    Properties
    Various
    LAN Desktop
        
    My Network Places
          \  
    Properties
                 
    Local Area Connection Double Click to activate
    TCP/IP Setting Desktop
        
    My Network Places
          \  
    Properties
              \  
    Local Area Connection
                  \  
    Internet Protocol (TCP/IP)
                      \  
    Properties
                          \  
    Advanced
                              \  
    Options
                                    
    TCP/IP Filter Values in ‘Permit Only’
    Date/Time Start
        
    Settings
          \  
    Control Panel
                 
    Date and Time Proper Time Zone
    Accurate Date / Time
    Figure 8-69   UPS Setting 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    8-18Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
    8-2-8-2 Drive Letter Setting
    If you CANNOT see this dialog at the first boot up, this dialog appears at 2nd boot up.
    Please choose  OK.
    You should check the Drive Letter manually by Windows Explorer for double check. - DVD Drive should be assigned to G:
    If Drive letter is wrong, please double-click Remap Drive Letter  icon on the desktop.
    8-2-8-3 Hardware Device Check: 1.) On the windows desktop, left-click on  My Computer, then right-click on Properties.
    Figure 8-70   OK
    CAUTIONDrive F: (Swap) Letter is removed on R7.0.0EZ. (This is correct.)
    Figure 8-71   Properties 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-19
    2.) The System Properties window appears. Select Hardware > Divice Manager...
    3.) Verify that X or ! is not displayed.
    Figure 8-72   Divice Manager
    NOTICEThe Eusc  (Mlultifunction adapters > Eusc) might contain “ !” mark, but it is no problem. Skip any of 
    device driver installation procedures described below.
    “!” mark on Eusc is no problem. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    8-20Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
    8-2-8-3 Hardware Device Check:  (cont’d)
    4.) If  X or  ! is displayed:
    - Left-click on the device with  x mark, then rigtht-click on Enable.
    - Left-click on the device with  ! mark, then rigtht-click on Properties  to perform the device driver 
    reconfiguration. When some files are required, designate the file in the directories of 
    C:\WINDOWS\SYSTEM32  or C:\WINDOWS\SYSTEM32\DRIVERS . If they are not found, 
    search them using a windows search function.
    5.) Close the window. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-21
    8-2-8-4 LAN Parameter Setting
    1.) On the desktop, left-click on  My Network Places, then right-click on  Properties...
    2.) Verify that two network icons appears. (If “Local Area Connection” icon does not appear, the  hardware failure might occur.)
    3.) Using the right trackball button, double-click on  Local Area Connection.
    4.) When the following window appears, it will be automatically closed later.
    5.) When the following window appears, right-click on  Cancel.
    6.) Close all windows. Figure 8-73   Properties
    Figure 8-74   Local Area Connection Figure 8-75   Cancel 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    8-22Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
    8-2-8-5 TCP/IP Filter Setup
    1.) On the desktop, left-click on  My Network Places, then right-click on  Properties...
    2.) Double-click  Local Area Connection  or Local Area Connection #2 (if you do NOT find Local 
    Area Connection).
    3.)  Select  Internet Protocol (TCP/IP)  , then click Properties button.
    Figure 8-76   Local Area Connection
    Figure 8-77   Internet Protocol (TCP/IP)
     
      
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-23
    8-2-8-5 TCP/IP Filter Setup (cont’d)
    4.) Click Advanced...  button.
    5.)  Click  Options tab.Figure 8-78   Advanced
    Figure 8-79   Options
     
      
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    8-24Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
    8-2-8-5 TCP/IP Filter Setup (cont’d)
    6.) Select TCP/IP Filtering , then click Properties button.
    Figure 8-80   TCP/IP Filtering 
    						