Home > GE > Medical Equipment > GE Logiq 7 Service Manual

GE Logiq 7 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 394
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 2-2 - General Console Requirements 2-5
    2-2-3  EMI Limitations
    Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, 
    magnetic fields, and transient in the air wiring. They also generate EMI. The LOGIQ™ 7 complies with 
    limits as stated on the EMC label. However there is no guarantee that interface will not occur in a 
    particular installation.
    Possible EMI sources should be identified before the unit is installed.
    Electrical and electronic equipment may produce EMI unintentionally as the result of defect.
    These sources include:
    • medical lasers,
    • scanners,
    • cauterizing guns,
    •computers,
    •monitors,
    •fans,
    • gel warmers,
    • microwave ovens,
    • light dimmers,
    • portable phones.
    The presence of broadcast station or broadcast van may also cause interference. See for EMI 
    Prevention tips.
    Table 2-5   EMI Prevention/abatement
    EMI RuleDetails
    Be aware of RF sourcesKeep the unit at least 5 meters or 15 feet away
     from other EMI sources. Special shielding 
    may be required to eliminate interference problems caused by high frequency, high 
    powered radio or video broadcast signals.
    Ground the unit Poor grounding is the most likely reason 
    a unit will have noisy images. Check grounding 
    of the power cord and power outlet.
    Replace all screws, RF 
    gaskets, covers, cores After you finish repairing or updating the system, replace all covers and tighten all 
    screws. Any cable with an external connecti
    on requires a magnet wrap at each end. 
    Install the shield over the front of card cage. Loose or missing covers or RF gaskets allow 
    radio frequencies to interface with the ultrasound signals.
    Replace broken RF gasketsIf more than 20% or a pair of fingers on the RF gaskets are broken, replace the gaskets. 
    Do not turn on the unit until any loose metallic part is removed.
    Do not place labels where 
    RF gaskets touch metalNever place a label where RF gaskets meet the unit. Otherwise, the gap created will 
    permit RF leakage. Or, if a label has been found in such a position, move the label.
    Use GE specified 
    harnesses and peripheralsThe interconnect cables are grounded and r equire ferrite beads and other shielding. 
    Also, cable length, material, and routing are all important; do not change from what is 
    specified.
    Take care with cellular 
    phones Cellular phones may transmit a 5 V/m signal; that could cause image artifacts.
    Properly dress peripheral 
    cables Do not allow cables to lie across the top of the card cage or hang out of the peripheral 
    bays. Loop the excess length for peripheral cables inside the peripheral bays. Attach the 
    monitor cables to the frame. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    2-6Section 2-2 - General Console Requirements
    2-2-4 Probes Environmental Requirements
    Table 2-6   Operation and storage Temperatures for 2D Probes
    OperationalStorageTransport
    Temperature
    10 - 40 oC
    50 - 104 oF -10 - 60 
    oC
    14 - 140oF -40 - 60 
    oC
    -40 - 140 oF
    Humidity30 - 85%
    non-condensing 30 - 90%
    non-condensing30 - 90%
    non-condensing
    Pressure700 - 1060hPa 700 - 1060hPa 700 - 1060hPa
    Table 2-7   Operation and storage Temperatures for 4D Probes
    OperationalStorageTransport
    Temperature
    18 - 30 oC
    64- 86 oF -10 - 50 
    oC
    14 - 122 oF -10 - 50 
    oC
    14 - 122 oF
    HumidityMax. 70%
    non-condensing Max. 90%
    non-condensingMax. 90%
    non-condensing
    Pressure700 - 1060hPa 700 - 1060hPa 700 - 1060hPa 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 2-3 - Facility Needs 2-7
    Section 2-3
    Facility Needs
    2-3-1 Purchaser Responsibilities
    The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, 
    confusion, and waste of manpower can be avoided by completing pre installation work before delivery. 
    User the Pre Installation checklist to verify that all needed steps have been taken,
    Purchaser reasonability includes:
    • Procuring the materials required.
    • Completing the preparations before delivery of the ultrasound system.
    • Paying the costs for any alternations and modifications not specifically provided in the sales 
    contract.
    NOTE: All electrical installation that are preliminary to the positioning of the equipment at the site  prepared for the equipment must be performed by licensed electrical contractors. Other 
    connections between pieces of electrical equipment, products involved (and the accompanying 
    electrical installations) are highly sophisticated and special engineering competence is 
    required. All electrical work on these product must comply with the requirements of applicable 
    electrical codes. The purchaser of GE equipment must only utilize qualified personnel to 
    perform electrical servicing on the equipment.
    The desire to use a non-listed or customer provided product or to place an approved product further 
    from the system than the interface kit allows presents challenges to the installation team. To avoid 
    delays during installation, such variances should be made known to the individuals or group performing 
    the installation at the earliest possible date (preferable prior to purchase).
    The ultrasound suite must be clean proof to delivery of the machine. Carpet is not recommended 
    because it collects dust and creates static. Potential sources of EMI (electromagnetic interference) 
    should also be investigated before delivery. Dirt, static, and EMI can negatively impact system. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    2-8Section 2-3 - Facility Needs
    2-3-2 Required Features
    • Dedicated single branch power outlet of adequate amperage (see  Table 2-3) meeting all local and 
    national codes which is located less than 2.5 m (8 ft.) from the unit’s proposed location
    • Door opening is at least 76 cm (30 in) wide
    • Proposed location for unit is at least 0.3 m (1 ft.) from the wall for cooling
    • Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with  peripheral within 1 m of the unit to connect cables.
    NOTE: The LOGIQ™ 7 has four outlets inside the unit. One is for the monitor and three for on board peripherals.
    • Power outlets for other medical equipment and gel warmer
    • Power outlets for test equipment and modem within 1 m (3.2 ft.) of unit
    • Clean and protected space to store transducers (in their cases or on a rack)
    • Material to safely clean probes (done with a plastic container, never metal)
    2-3-3 Desirable Ultrasound Room Facilities
    • Door is at least 92 cm (3 ft.) wide
    • Circuit breaker for dedicated power outlet is easily accessible
    • Sink with hot and cold water
    • Receptacle for bio–hazardous waste, like used probe sheaths
    • Emergency oxygen supply
    • Storage for linens and equipment
    • Nearby waiting room, lavatory, and dressing room
    • Dual level lighting (bright and dim)
    • Lockable cabinet ordered by GE for its software and proprietary manuals 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 2-3 - Facility Needs 2-9
    2-3-4 Recommended and Alternate Ultrasound Room Layout
    Recommended standard floor plan and a minimal floor plan for ultrasound equipment:
    Figure 2-1   RECOMMENDED ULTRASOUND ROOM  LAYOUT
    LOGIQ
    LOGIQ 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    2-10Section 2-3 - Facility Needs
    2-3-5 Networking Pre-installation Requirements
    2-3-5-1 Purpose of DICOM Network Function
    DICOM services provide the operator with clinically useful features for moving images and patient 
    information over a hospital network. Examples of DICOM services include the transfer of images to 
    workstations for viewing or transferring images to remote printers. As an added benefit, transferring 
    images in this manner frees up the on-board monitor and peripherals, enabling viewing to be done while 
    scanning continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at 
    a lower cost.
    2-3-5-2 DICOM Option Pre-installation Requirements To configure the LOGIQ™ 7 to work with other network connections, the site’s network administrator 
    must provide some necessary information.
    Information must include: 
    • A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ™ 7. 
    • The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION.
    • The host name, IP address, port and AE Title for each device the site wants connected to the 
    LOGIQ™ 7 for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and 
    the revision of the device, is also included. This information may be useful for solving errors. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 2-3 - Facility Needs 2-11
    2-3-5-2 DICOM Option Pre-installation Requirements (cont’d)
    .
    Figure 2-2   Worksheet for DICOM Network Information
    LOGIQ™ 7
    Host Name
    AE TitleLocal PortIP Address
    Net Mask. .
    ..
    .
    ..
    ROUTING INFORMATION
    ROUTER2
    ROUTER3 ROUTER1
    .
    ..
    .
    ..
    .
    ..
    .
    ..
    .
    ..
    .
    ..
    .
    .. Destination 
    IP Addresses
    GATEWAY IP Addresses
    Default
    DICOM APPLICATION INFORMATION NAME
    Store 2 Store 3 Store 1
    MAKE/REVISION
    IP ADDRESSESPORT
    AE TITLE
    Store 5 Store 6 Store 4
    Storage 
    Commit MPPS
    Worklist
    .
    ..
    .
    ..
    .
    ..
    .
    ..
    .
    ..
    .
    ..
    .
    ..
    .
    ..
    .
    .. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    2-12Section 2-3 - Facility Needs
    This page was intentionally left blank. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 3-1 - Overview 3-1
    Chapter 3
    Installation
    Section 3-1
    Overview
    3-1-1 Purpose of Chapter 3
    This chapter contains information needed to install the unit. Included are references to a procedure that 
    describes how to receive and unpack the equipment and how to file a damage or loss claim.
    How to prepare the facility and unit of the actual installation, and how to check and test the unit, probes, 
    and external peripherals for electrical safety are included in this procedure. Also included in this section 
    are guidelines for transporting the unit to a new site.
    3-1-2 Average Installation Time
    The LOGIQ™ 7 has been designed to be installed and checked out by an experienced service 
    technician in approximately four hours. LOGIQ™ 7 consoles with optional equipment may take slightly 
    longer. Table 3-1   Contents in Chapter 3
    SectionDescriptionPage Number
    3-1
    Overview 3-1
    3-2 Receiving and Unpacking the Equipment 3-3
    3-3 Preparing for Installation 3-10
    3-4 Completing the Installation 3-11
    3-5 Installation Paperwork 3-18
    Table 3-2   Average Installation Time
    DescriptionAverage Installation TimeComments
    Unpacking the scanner
    0.5 hour
    Scanner wo/options 0.5 hour
    Dependant on the configuration that is required
    DICOM Option 0.5 hour
    Dependant on the amount of configuration
    InSite Option 0.5 hour 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    3-2Section 3-1 - Overview
    3-1-3 Installation Warnings
    1.) Since the LOGIQ™ 7 weighs approximately 225 kg. (496 lb) without options, preferably two people 
    should unpack it. Two people are also preferable for installing any additional bulky items.
    2.) There are no operator serviceable components. To prevent shock, do not remove any covers or  panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service 
    personnel should carry out servicing and troubleshooting.
    3.) After being transported, the unit may be very cold or hot. If this is the case, allow the unit to  acclimate before you turn it on. It requires one hour for each 2.5×C increment its temperature is 
    below 10×C or above 40×C.
    3-1-3-1 Brake Pedal Operation
    CAUTIONEquipment damage possibility. Turning the system on without acclimation after arriving at site 
    may cause the system to be damaged.
    Table 3-3   Time for Settlement
    °C60 55 50 45 40 35 30 25 20 15 10 5 0 -5 -10 -15 -20 -25 -30 -35 -40
    °F140 131 122 113 104 96 86 77 68 59 50 41 32 23 14 5 -4 -13 -22 -31 -40
    hrs864200000002468101214161820
    WARNINGWARNING
    R
    REMEMBER: IF THE FRONT CASTER SWIVEL LOCK IS ENGAGED FOR 
    TRANSPORTATION, PRESSING THE RELEASE PEDAL ONCE EDISENGAGES THE 
    SWICEL LOCK. YOU MUST DEPRESS THE RELEASE PEDAL A SECOND TIME TO 
    ENGAGE THE BRAKE. 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Logiq 7 Service Manual