GE Logiq 7 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 394
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 5-8 - Monitor Video Specification 5-45
    5-8-2-4 NTSC EIA
    5-8-3 SVHS and Composite Video
    5-8-3-1 Basic DC Parameters
    GeneralSVHS-lumaSVHS-cromaCompositeRGBB&W
    Line/field: 525/60
    FH: 15734 HzFV: 59.94 Hz
    Bandwidth: Min.  4.2MHz (luma) Signal: 92.5 IRE 
    (from black level)  100% amplitude
    Blanking setup:  7.5 IRE
    Sync: 40 IRE Burst amplitude: 
    +/- 20 IRE, 9 
    cycles
    Fsc: 3.579545  MHz 100% saturation Summed luma 
    and croma Signal: 0-700 
    mV 
    Comp sync/H 
    sync/ V sync: 0.3 
    - 4 Vpp SVHS luma
    ParametersNTSCPAL
    White relative to blank
    714 +/- 7 mV700 +/- 7 mV
    Black relative to blank 54 +/- 7 mV 0
    Sync relative to blank -286 +/- 7 mV-300 +/- 7 mV
    Burst amplitude (nominal, p-p) 286 +/- 7 mV300 +/- 7 mV 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    5-46Section 5-8 - Monitor Video Specification
    This page was intentionally left blank. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-1 - Overview 6-1
    Chapter 6
    Service Adjustments
    Section 6-1
    Overview
    6-1-1 Purpose of this chapter 6
    This section describes how to test and adjust the scanner. These tests are optional. You may use them 
    to check the system for errors.
    Table 6-1   Contents in chapter 
    SectionDescriptionPage Number
    6-1
    Overview 6-1
    6-2 LV Unit Adjustments (For BT03 or lower) 6-2
    6-3 Caster Brake/Swivel Function Adjustments 6-4
    6-4Reloading the Probe Data6-9
    6-5Monitor and LCD Adjustments6-10
    6-6BW Printer Setting / Adjustment6-27
    6-7 Cleaning the Trackball 6-33
    6-8 Jumper and Dip Switch Setting 6-37 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-2Section 6-2 - LV Unit Adjustments (For BT03 or lower)
    Section 6-2
    LV Unit Adjustments (For BT03 or lower)
    This system contains three power supply modules; LV unit. The LV unit for BT04 or later can not be 
    adjusted.
    6-2-1 Access to Adjustments
    1.) Remove the LV unit:
    a.) Remove the left side cover.
    b.) Remove the LV unit cover.
    c.) Remove the bracket, then pull out the LV unit Assy with the cables connected.
    Figure 6-1   LV Unit Assy
    LV Unit Assy 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-2 - LV Unit Adjustments (For BT03 or lower) 6-3
    6-2-2 Adjustments Procedures
    1.) Using the following VRs, DC output can be adjusted.
    Table 6-2   DC Output Specification for LV Unit
    VRSpecifications
    VR1
    3.4V +/- 34mV
    VR25V +/- 50mV
    VR36V +/- 60mV
    VR4-6 V +/- 60mV
    VR512V +/- 120mV
    VR6-12V +/- 120mV
    Figure 6-2   VRs for Adjusting DC Output
    VR4 for -6V VR2 for +5V
    VR3 for +6V
    VR6 for -12V VR1 for +3.4V
    VR5 for +12V
    OR
    -6V
    +12V
    +5V -12V +3.3V +6V 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-4Section 6-3 - Caster Brake/Swivel Function Adjustments
    Section 6-3
    Caster Brake/Swivel Function Adjustments
    6-3-1 Brake Function Adjustment
    The LOGIQ 7 contains front and rear brake lock adjusters. They are separately adjusted using the 
    similar method.
    Figure 6-3   Locations of Brake Lock Adjusters
    Front Brake Lock Adjuster Rear Brake Lock Adjuster 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-3 - Caster Brake/Swivel Function Adjustments 6-5
    6-3-1 Brake Function Adjustment (cont’d)
    1.) Locate the adjuster nut on the caster linkage joint.
    2.) Rotate the adjuster until the caster lock lever comes center as shown. Figure 6-4   Adjuster Nut Location
    Figure 6-5   Rotating Adjuster
    Adjuster Nut
    Center lock lever 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-6Section 6-3 - Caster Brake/Swivel Function Adjustments
    6-3-1 Brake Function Adjustment (cont’d)
    3.) After completion of adjustment, press on the brake and check its function and location of the boss,  connected to the caster lock arm.
    Figure 6-6   Brake Function Check
    Example: GOOD
    Boss of the front caster in place, and brake 
    properly adjusted. Example: NO GOOD
    Boss of the front caster NOT in place, and brake 
    NOT properly adjusted. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-3 - Caster Brake/Swivel Function Adjustments 6-7
    6-3-1 Brake Function Adjustment (cont’d)
    4.) Adjuster of the rear caster is also located on the caster linkage joint as shown.
    5.) Turn the adjuster so that the caster lock lever is in the ceter position, as same as the front caster,  and the link is securely in place. Figure 6-7   Rear Caster Adjuster Location
    Adjuster Nut
    Link
    Rear Caster
    Lock Arm
    Boss 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-8Section 6-3 - Caster Brake/Swivel Function Adjustments
    6-3-2 Swivel Function Adjustment
    The LOGIQ 7 contains one swivel lock adjuster. It can be adjusted using the similar method as brake 
    function adjustment described above.
    Figure 6-8   Locations of Swivel Lock Adjuster
    Swivel Lock Adjuster 
    						