    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 3-4 - Completing the Installation 3-13
    3-4-2 Optional Peripherals/Peripheral Connection (cont’d)
    Check if the fixing belts are loosen. If it is, tighten the belts again to secure the peripheral(s). If 
    necessary, remove the peripheral from the device room to tighten the belts.
    3-4-2-1 Approved on-board peripherals Refer to Section 5-3 Peripheral Compatibility .
    Figure 3-6   How to Hold Peripherals
    Belts
    Belts BeltsConnect the related cables 
    between the peripheral and 
    LOGIQ 7 if they are NOT con-
    nected each other. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    3-14Section 3-4 - Completing the Installation
    3-4-2-1 Approved on-board peripherals  (cont’d)
    Connecting Cables
    3-4-2-2 Reference off-board peripherals and options None.
    CAUTIONEquipment damage possibility. Be sure to use the following recommended connecting cables to connect 
    recording devices and a network with LOGIQ™ 7 console.
    Table 3-4   List of Connecting Cables
    NamePart No.FigureNOTE
    Power Supply Cable P9509EE Connected to power
    USB Serial Bridge  Cable 2304621 For converting the signal of 
    RS232C cable to USB cable:
    connected to VCR1 on the Rear 
    Panel
    AV Cable 2119874 Connected to Video-In/Out on the 
    Rear Panel
    The followings are the cables fo r BT04 ore lower system ONLY.
    Mini-Plug Cable P9509BE Shutter control signals:
    connected to B/W Printer
    RS232C Cable  Cross 2305550
    For control signals:
    connected to Serial Bridge Cable
    RS232C Cable  Straight 2305549
    For control signals:
    connected to Serial Bridge Cable
    BNC Cable 2297053 For control signal:
    connected to Composite B/W
    USB Cable 2324360 Connected to USB port.
    SCSI cables
    (UP-D50 ONLY) 2375479
    Connected toSCSI port and SCSI 
    Cable Connector. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 3-4 - Completing the Installation 3-15
    3-4-3 Available Probes
    See in specification in the LOGIQ™ 7 Reference Manual for Probes and intended use.
    See Chapter 9 - Renewal Parts for Part Numbers to be used when ordering new or replacement probes.
    Table 3-5   List of Transducers supported (All models)
    Probe Name
    Material of 
    Headshell
    Area of UsingTYPECatalog NumberPart Number
    3C PESABDOMINAL CONVEX H79802P
    H40412LB 2286353
    2286354
    5C PESABDOMINAL CONVEX H79822P
    H40412LA 2294515
    2294516
    8C PBT NEONATAL
    PEDIATRICS MICRO-CONVEX
    H79792P
    H40412LJ 2348093
    2348094
    E8C NORYL
    PBT TRANSVAGINAL MICRO-CONVEX H79852P
    H40412LE 2294640
    2294641
    M7C PBTABDOMINAL CONVEXH79832P
    H40412LC 2294513
    2294514
    M12L PBTSMALL PARTS LINEARH79842P
    H40412LD 2294510
    2294511
    7L NORYL ABDOMINAL
    SUPERFICIAL LINEAR
    H79862P
    H40412LF 2294520
    2294521
    10L NORYL SUPERFICIAL LINEARH79872P
    H40412LG 2294522
    2294523
    10S ABSPEDIATRIC SECTORH79922P
    H4901PC 2309478
    2298589
    3.5C NORYL ABDOMINAL CONVEXH79812P
    H4901PE 2303215
    2050357
    3.5CS NORYL ABDOMINAL CONVEXH78042P
    H40412LK 2380854
    2051858
    3S NORYL CARDIOLOGY SECTORH79632P
    H4701SZ 2348878
    2323337
    i12L ABSINTRAOPERATIVE LINEARH79322P
    H4012L  2270556
    2264883
    M3S PBTCARDIOLOGY SECTORH79892P
    H45011SZ 2293726
    2378099
    6T-OR-TEE PU: PolyUrethane TRASOPHAGEAL FOR 
    ADULT
    CARDIOLOGY SECTOR
    H79932P
    H45521DX 5131947
    KN100068
    7S PBTCARDIOLOGY SECTORH78082P
    H40422LB 2355698
    2347471
    BE9C PBTTRANSRECTAL MICRO-CONVEX
    H40412LW2389381
    2389382
    T739 NORYLINTRAOPERATIVE LINEARH76572SR
    H40212LM2259245
    2259246
    4D3C_L PBTABDOMINAL CONVEX Volume H44801G 5121652
    KTZ195893 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    3-16Section 3-4 - Completing the Installation
    NOTE: PES: Polyethersulfone NORYL: Modified Polyphenylene Oxide PU: Polyurethane
    PBT: Polybutylene Terephthalate ABS: Acrylonitrile Butadiene Styrene
    NOTE: Some probes indicated on the table above have two different part numbers. The upper row  shows the part numbers of probes for Japan. The lower row shows the part numbers of probes 
    for regions other than Japan. Probes which have only one part number are not available in 
    Japan.
    The following transducers, additional to ones listed in Table  3-6, are supported by V65x and later LOGIQ 
    7 models.
    The following transducers, additional to ones listed in Table 3-7 , are supported by R7.5.x and later 
    LOGIQ 7 models.
    The following transducers, additional to ones listed in Table 3-8 , are supported by BT09 and later 
    LOGIQ 7 models.
    4D10L PBDSUPERFICIAL LINEAR Volume H44801GB 5121651
    KTZ156836
    4C PBTABDOMINAL CONVEXH4904PC5131944
    5123455
    12L NorylSUPERFICIAL LINEARH40412LH2295375
    2295377
    P2D ABSCARDIOLOGY CWDH4830JETE100024
    P6D ---SUPERFICIAL CWDH4830JGTQ100002
    Table 3-6   List of Transducers supported by V65x and later 
    Probe Name
    Material of 
    Headshell
    Area of UsingTYPECatalog NumberPart Number
    9L PBTVASCULAR SMALL 
    PARTS LINEAR
    H40412LT5131433
    4DE7C PBTOB Gyn Urology CONVEX Volume H44801GA 5121650
    Table 3-7   List of Transducers supported by R7.5.x and later 
    Probe Name
    Material of 
    Headshell
    Area of UsingTYPECatalog NumberPart Number
    3CRF PU: PolyUrethane ABDOMINALMICRO-CONVEX H40442LP 5214820
    5196216
    Table 3-8   List of Transducers supported by BT09 and later 
    Probe Name
    Material of 
    Headshell
    Area of UsingTYPECatalog NumberPart Number
    11L PBTSUPERFICIAL LINEARH78842P
    H40412LY 5251881
    5171885
    Table 3-5   List of Transducers supported (All models)
    Probe Name
    Material of 
    Headshell
    Area of UsingTYPECatalog NumberPart Number 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 3-4 - Completing the Installation 3-17
    3-4-4 Video Specification
    3-4-5 Software Option Configuration
    3-4-5-1 Onsite check and configuration
    Select Utility > Admin > System Admin  and check the option software to be installed.
    Table 3-9   Video Specifications
    CRTLCD
    Timing Parameter
    800x600
    75Hz
    Timing Parameter
    1280 x 1024 60Hz
    Horizontal Rate [kHz] 46.88Horizontal Rate [kHz] 64
    Horizontal Period [ µs] 21.33 Horizontal Period [µs] 15.625
    Pixel Clock [MHz] 49.50Pixel Clock [MHz] 108
    H Blank Width  [µs] 5.17 H Blank Width [µs] 3.778 (408dots)
    H Sync Width  [µs] 1.62 H Sync Width [µs] 1.037 (112dots)
    H Front Porch  [µs] 0.32 H Sync Front Porch [µs] 0.444 (48dots)
    Active Horizontal Period  [µs] 16.16 Active Horizontal Period  [µs] 11.852 
    (1280dots)
    Vertical Rate [Hz] 75.00Vertical Rate [Hz] 60.0
    Vertical Period [ms] 13.33Vertical Period [ms] 16.67 
    (1066lines)
    V Sync Width [lines=ms] 25=0.53V Sync Front Porch [ms] 0.016 (1 line)
    V Front Porch  [lines=µs] 3=64.00 Equalization Porch None
    Equalization Gate  [lines=µs] 1=21.3 Lines: Field/Frame 1066
    Lines: Field/Frame 625Active Lines/Frame 1024
    Active Lines/Frame 600 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    3-18Section 3-5 - Installation Paperwork
    Section 3-5
    Installation Paperwork
    NOTE: During and after installation, the documentation (i.e. Users Manual, Installation Manuals...) for 
    the peripheral units must be kept as part of the original system documentation. This will ensure 
    that all relevant safety and user informations are available during the operation and service of 
    the complete system.
    3-5-1 Peripherals/Accessories Connector Panel
    LOGIQ™ 7 peripherals and accessories can be properly connected using the rear connector panel 
    located behind the rear door, front connector panel located next to the video printer, and Footswitch 
    connector located bottom of the OP panel. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 3-5 - Installation Paperwork 3-19
    3-5-1-1 Rear Panel Connector
    Located on the rear panel are video input and output connectors, audio input and output, camera 
    expose connectors, footswitch connector power connector and control connections for VCR, printer, 
    and service tools. 
    This section indicates the pin assignment for each connector.
    Figure 3-7   Rear Pannel Connector (BT04 or later)
    Video In
    Rec/Expose
    VGA
    S-Video
    Out
    S-Video
    In
    CmpstOut
    Insite Audio
    In
    LR
    LR
    AudioOut
    Ethernet
    VCR2 VCR1 Service
    CameraB/W
    Color B/W
    Color
    Video Out
    Power
    AC Line Input
    1200VA
    220-240V
    Circuit
    Breaker
    Circuit
    Breaker
    50 /60Hz
    Power
    AC Line Input
    1200VA
    50 /60Hz
    100-120V
    Circuit
    Breaker
    Circuit
    Breaker
    100-120V  350VA Max
    Rear Pannel Connector for BT04 and later models
    5
    6
    87
    2
    3
    1 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    3-20Section 3-5 - Installation Paperwork
    3-5-1-1 Rear Panel Connector (cont’d)
    NOTE: Each outer (case) ground line of peripheral/accessory connectors are protectively grounded.  Signal ground lines are not isolated, except the Service port (3). All of signal lines (include signal 
    GND) of the Service port are isolated.
    The specified peripherals/accessories only can be connected to the USB ports.Rear Pannel Connector for BT03 or lower
    Figure 3-8 Rear Panel Connector (BT03 or lower)
    
    
    USB1.1 (USB compatible) 
    or (USB + USB Serial Bridge 
    + RS232C)
    (Reserved for future use)
    2
    1
    3
    4
    5
    6
    87
    USB1.1 (USB compatible) 
    or (USB + USB Serial Bridge 
    + RS232C)
    USB1.1 (USB compatible) 
    or (USB + USB Serial Bridge 
    + RS232C) 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 3-5 - Installation Paperwork 3-21
    3-5-1-1 Rear Panel Connector (cont’d)
    Pin Assignment of S-Video Connector Connector: S-Terminal, 4-pin
    Pin Assignment of VGA Connector Connector: Shrank D-Sub, 15-pin
    Pin Assignment of Service/VCR 1/VCR 2 Connector, USB1.1 Connector: 4 pin Table 3-10   Pin Assignment of S-Video Connector
    Pin No Output/Input Signal Description
    1SVIDEO OUT/IN YG Y (Luma) GND
    2 SVIDEO OUT/IN CG C (Chroma) GND
    3 SVIDEO OUT/IN Y Y (Luma) SIGNAL
    4 SVIDEO OUT/IN C C (Chroma) SIGNAL
    Table 3-11   Pin Assignment of VGA Connector
    Pin NoOutput Signal Description
    1IO VGA OUT1 R Red
    2 IO VGA OUT1 G Green
    3 IO VGA OUT1 B Blue
    6 IO VGA OUT1 RG Reg GND
    7 IO VGA OUT1 GG Green GND
    8 IO VGA OUT1 BG Blue GND
    13 IO VGA OUT1 HS H Sync
    14 IO VGA OUT1 VS V Sync
    Others GND GND
    Table 3-12   Pin Assignment of Service/VCR 1/VCR 2 Connector
    Pin NoOutput Signal Description
    1VBUSn Power Supply
    2 Dn Data (-)
    3 Dn Data (+)
    4 GNDn Power Ground
    1
    24
    3
    1
    2
     1
     6
     115 
    10 
    15 
    3
    1  2  3  4 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    3-22Section 3-5 - Installation Paperwork
    3-5-1-1 Rear Panel Connector (cont’d)
    Pin Assignment of Digital Int. Connector Connector: 6-pin
    Pin Assignment of Ethernet Connector: RJ-45 Modular, 8-pin
    Pin Assignment of Insite Connector: RJ-11 Modular, 6-pin Table 3-13   Pin Assignmen t of IEEE-1394 Connector)
    Pin NoOutput Signal Description
    1VP 1394 Power Supply
    2 VG 1394 Power Ground
    3 TPB 1394 Data B (-)
    4 TPB 1394 Data B (+)
    5 TPA 1394 Data A (-)
    6 TPA 1394 Data A (+)
    Table 3-14   Pin Assignment of Ethernet Connector
    Pin NoOutput Signal Description
    1ETHER TD Ethernet TD+
    2 ETHER TD Ethernet TD-
    3 ETHER RD Ethernet RD+
    6 ETHER RD Ethernet RD-
    Others NC Non-connection
    Table 3-15   Pin Assignment of Insite Connector
    Pin NoOutput Signal Description
    2TEL L4 Telephone L4
    3 TEL L2 Telephone L2
    4 TEL L1 Telephone L1
    5 TEL L3 Telephone L3
    Others NC Non-connection
    4
    1  3  5 2  4  6
    5
     8 1
    6
     8 1 
    						
    All GE manuals Comments (0)