    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-6 - BW Printer Setting / Adjustment 6-29
    6-6-2 Parameters for UP-D895
    1.) When system started, Go to Utility > System >Peripherals, then hit Printers button in the Setup 
    group.
    2.) Click  SONY UP-D895  in the Printer List, then right click and select Properties .
    3.) Select  General tab, then hit  Printing Preferences ... 
    4.) Set  Landscape  into Orientation, then click on Advanced...
    5.) Set  1920x1280  into PaperFigure 6-37   Landscape
    Figure 6-38   Paper 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-30Section 6-6 - BW Printer Setting / Adjustment
    6-6-2 Parameters for UP-D895 (cont’d)
    6.) Click on  Scaling..., then check   off into Enlarge to Paper and set  Bilinear into Interpolation 
    Method.
    7.) Click on  Density Adjust... , set TONE1  into GAMMA and set 7  into Sharpness.
    6-6-3 Service Tips - Print Quality
    When printer is set to “Nearest Neighbor” interepolation method, comments and body pattern may 
    become hard to read.  Re-adjust interpolation method to “Bi-Linear” to avoid this situation. Figure 6-39   Scaling
    Figure 6-40   Density Adjust
    Bilinear 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-6 - BW Printer Setting / Adjustment 6-31
    6-6-4 Service Tips - Print Speed
    1.) Make sure that the UP-D897 is switched ON. (The back light of the LCD light is in green.) If not, 
    press the ON/OFF switch.
    2.) Confirm that “READY” is displayed on the LCD, then press
     the jog dial.
    The UP-D897 enters the menu mode. When you perform a menu operation for the first time 
    after you purchase the unit , the top item of the “BEEP” menu will be displayed. If you already 
    have performed a menu operation, the items that you set last is displayed.
    NOTICEProblem:
    Compared to the UP-D895, it takes a few seconds longer for the UP-D897 to receive data from L7 
    system. All that time, an user cannot operate the L7 system and must wait until the UP-D897 finished 
    receiving data from the L7.
    Cause:
    This is due to driver software difference between UP-D895 and UP-D897. It seems the UP-D897 driver 
    software has been modified from the UP-D895.
    Solution:
    The following actions must be required: • UP-D897 printer: Change from UP-D897 to UP-D895 on “COMD” menu using the printer jog  dial. This solution allows the UP-D897 to be configured to use the UP-D895 interface and driver. 
    When the “COMD” is selected to “CO:895”, the response time becomes the same as the UP-
    D895. Perform this section.
    • LOGIQ7: Change from UP-D897 to UD-D895 using  Connectivity > Service > BW Printer to 
    use UP-D895 driver software.
    Jog Dial
    (PRESS)
    READY 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-32Section 6-6 - BW Printer Setting / Adjustment
    6-6-4 Service Tips - Print Speed (cont’d)
    3.) Turn the jog dial up or down until “ COMD” is displayed on the LCD.
    4.) Press the jog dial. “ CO:897” is displayed.
    The UP-D897 enters the mode in which you can select the driver setting. CO:897, which is 
    the driver setting for D897, is displayed on the LCD. You need to change to CO:895 which 
    is the driver setting for the D895.
    5.) Turn the jog dial up or down until “CO:895” is displayed on the LCD.
    6.) Press the jog dial. The LCD display changes to “COMD”. The UP-D897 completes to be set to the  UP-D895 mode.
    Notice:
    With setting the UP-D897 COMD to UP-D895, the following items cannot be selected or operated from 
    the jog dial.
    Item No.ItemsDescriptions
    1GAMMATONE4(VERY SOFT) cannot be selected.
    2GAMMA
    Cannot be selected from jog dial.  It can change only from Properties of UP-D895 menu.
    Corresponding type of GAMMA between UP-D895 and UP-D897 is as follows:
    UP-D897 -> UP-D895
    TONE1 -> SOFT (TONE3)
    TONE2 -> NORMAL (TONE2)
    TONE3 -> HARD (TONE1)
    TONE4 (VERY SOFT) -> Not Supported
    3QTYCannot be selected from jog dial.  It can change only from Properties of UP-D895 menu.
    4 RSIZE Cannot be selected from jog dial.  It can change only from Properties of UP-D895 menu.
    5 SHARP Cannot be selected from jog dial.  It can change only from Properties of UP-D895 menu.
    Jog Dial
    COMD
    Jog Dial
    CO:895 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-7 - Cleaning the Trackball 6-33
    Section 6-7
    Cleaning the Trackball
    1.) Power OFF the scanner.
    2.) Place your fingers onto the notches of the trackball retainer ring.
    3.) Rotate the retainer ring counterclockwise until it can be removed from the keyboard.
    4.) Lift off the inner retainer and trackball from the keyboard.Figure 6-41   Rotating the Retainer Ring
    Figure 6-42   Removing Inner Retainer and Trackball
    Retainer Ring
    Inner RetainerTrackball 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-34Section 6-7 - Cleaning the Trackball
    Section 6-7      Cleaning the Trackball (cont’d)
    5.) Wipe off any oil or dust from the trackball using a cleaner or dry cloth.
    6.) Wipe off any oil or dust from the trackball housing, rollers, and small ball, using a cleaner or cotton  bud.
    NOTICEWhen cleaning the trackball housing, make sure not to spill or spray any liquid into the trackball housing 
    (Keyboard or system). 
    Avoid organic solvents that may damage the mechanical parts of the trackball assembly.
    Do not apply much force to the small ball.
    Figure 6-43   Cleaning TrackBall and Housing
    Trackbal Housing
    Cotton Bud
    Roller Roller
    Small Ball 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-7 - Cleaning the Trackball 6-35
    Section 6-7      Cleaning the Trackball (cont’d)
    7.) Wipe off any oil or dust from the two rollers using a cleaner or cotton bud.
    NOTICEWhen cleaning the roller, make sure not to spill or spray any liquid into the trackball housing (Keyboard 
    or system). Use either ethanol, isopropyl alcohol or VCR head cleaner to clean the trackball assembly.
    Avoid other solvents that may damage the mechanical parts of the trackball assembly.
    Figure 6-44   Cleaning Rollers
    Rollers
    Cotton Bud 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-36Section 6-7 - Cleaning the Trackball
    Section 6-7      Cleaning the Trackball (cont’d)
    8.) Insert the trackball into the housing.
    9.) Place the trackball and inner retainer into the housing with its stopper facing down.Lift off the inner  retainer and trackball from the keyboard.
    10.) Install the trackball retainer ring onto the inner retainer, then rotate it clockwise until its notches are  set in the horizontal position. Figure 6-45   Removing Inner Retainer and Trackball
    Figure 6-46   Rotating the Retainer Ring
    Inner RetainerTrackball
    Inner RetainerStopper
    Retainer Ring Notches (Set Horizontally) 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-8 - Jumper and Dip Switch Setting 6-37
    Section 6-8
    Jumper and Dip Switch Setting
    6-8-1 Dip Switch Setting
    Normally the dip switches shall not be adjusted. In case of special needs, such as scanner upgrade, 
    refer to each instructions/documentations for proper setting.
    The dip switch must be set properly according to the table below. The value in the table (Dip Switch 
    setting) represents the bit location as shown.
    Figure 6-47   Dip Switch Setting
    Figure 6-48   Bit Setting (1 to 4) 4 (lower 4 bits)
    E (Upper 4 bits)
    Example: “E4 (h)” is set as shown.
    O
    N
    1
    2
    3
    4
    5
    6
    7
    8
    O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8
    0 (h) 1 (h) 2 (h) 3 (h) 4 (h) 5 (h) 6 (h) 7 (h)
    O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8
    8 (h) 9 (h) A (h) B (h) C (h) D (h) E (h) F (h)
    Bit Setting for DIP Switch 1 to 4 (lower 4 bits) 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-38Section 6-8 - Jumper and Dip Switch Setting
    6-8-1 Dip Switch Setting (cont’d)
    Figure 6-49   Bit Settings (5 to 8)
    0 (h)
    1 (h)
    2 (h) 3 (h) 4 (h) 5 (h) 6 (h) 7 (h)
    O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8
    O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8O
    N
    1
    2
    3
    4
    5
    6
    7
    8
    8 (h) 9 (h) A (h) B (h) C (h) D (h) E (h) F (h)
    Bit Setting for DIP Switch 5 to 8 (Upper 4 bits) 
    						