    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-13
    Section 7-6
    Diagnostics
    As described in Chapter 5 - Components and Functions (Theory), the service platform uses a web-
    based user interface (UI) to provide access to common service components. A service dongle (part 
    number 066E0703) is required to access areas proprietary to GE.
    The Diagnostic home page displays a listing of test categories consisting of Acquisition Diagnostics, I/
    O Devices and Common Diagnostics. Expand the desired main category to select groups or individual 
    tests.
    Figure 7-13   Global Service User Interface Diagnostic Screen 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-14 Section 7-7 - Acquisition Diagnostics
    Section 7-7
    Acquisition Diagnostics
    Acquisition Diagnostics is one of three main headings under the GSUI Diagnostics Tab. Figure 7-14 
    shows the top level menu choices under Acquisition Diagnostics. The subheadings and their respective 
    diagnostic/troubleshooting menu choices are called out and described in the sections that follow.
    Figure 7-14   GE Service Acquisition Diagnostic Menu Selections 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-15
    Section 7-7      Acquisition Diagnostics (cont’d)
    Individual selections run several diagnostics in a  Bottom-up Confidence-Building Order and return 
    status information identifying any problems. The Acquisition Diagnostic menu is structured as follows:
    7-7-1 Automatic Troubleshooting 
    -  Runs  all diagnostics and analyzes  the results  to isolate to  the  faulty  FRU.
    - This is the  BEST way to diagnose a problem with the system.
    - Takes less than  6 minutes for all tests to run.
    - Disconnect all probes from the scanner before executing the Automatic Troubleshooting  sequence.
    The choices for Automatic Troubleshooting, shown in  Figure 7-16 on page 7-16 , allows for running all 
    acquisition diagnostics under the selection Troubleshoot All Acquisition Hardware. Figure 7-15  Customer Acquisition Diagnostics 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-16 Section 7-7 - Acquisition Diagnostics
    7-7-1Automatic Troubleshooting (cont’d)
    Figure 7-16  Automatic Troubleshooting Selections
    Figure 7-17  User Acquisition Diagnostics 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-17
    When you select a test to run, like  Troubleshoot all Acquisition Hardware , the display changes as shown 
    in  Figure 7-18 on page 7-18  .
    1.)  Instructions  Frame
    - Displays either a test-specific text or the default instructions.
    2.)  Status  Frame
    - Initially displays the last known status for a selected diagnostic. Once the diagnostic starts,  the frame displays the “current” status of all test results.
    - The Status Frame also contains the user interface elements used for Diagnostic Control and  Operator Feedback.
    3.)  Execute  Button
    - This switch has two modes - each with appropriate text:
    - Execute - to start the diagnostic
    - Abort - to stop a diagnostic 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-18 Section 7-7 - Acquisition Diagnostics
    7-7-1Automatic Troubleshooting (cont’d)
    Figure 7-18  Automatic Diagnostics Selected
    Figure 7-19  Diagnostic Selected
    1
    2
    3
    45
    6
    7
    1
    2
    345
    6
    7 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-19
    4.)  Loop Count
    - This is an editable text field that will only accept numeric values of 4 digits or less. When the  switch is configured as an “execute” switch and pressed, the loop count field will be queried 
    to determine the number of times to execute the diagnostic.
    5.)  Progress  Indicator
    - Displays a graphical progress indicator for the user.
    6.)  Short  Text  Message
    - Displays brief messages about the test’s progress during execution.
    7.) Status  Frame  Background Color
    - Initially gray, the Status Frame background color changes upon completion of a diagnostic to  indicate completion status.
    - Code Status Fail = Red
    - Code Status Pass = Green
    - Code Status neither Pass nor Fail (Example: Aborted) = Gray (default color).
    7-7-2 Reliability Testing
    Overnight relialibility test.
    - Runs all diagnostics. Logs  ALL
     results. Results accumulate in the log each time the tests are 
    run. The log file continues to grow.
    NOTE: This is not intended for field service use. Typically a manufacturing test. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-20 Section 7-7 - Acquisition Diagnostics
    7-7-3 Diagnostic Listings
    Table 7-2     Acquisition Diagnostics and Utility Listing 
    Diagnostic/UtilityFunction
    Troubleshoot All Acquisition 
    DiagnosticsRuns all Acquisition Hardware diagnostics and isolates faults to a FRU.
    Overnight Reliability Test
    Runs all diagnostics and records a running log.
    SCB PCI Interface Diag
    Tests the host’s ability to communicate with the SCB via the secondary PCI bus
    SCB Memory Diag
    Tests all of the on-board memory, including registers.
    SCB DSP Diag
    Tests the Scan Sequencer’s DSP and associated components.
    SCB Scan Bus Diag
    Tests the local Scan Bus and the associated system timing.
    SCB Image Port Diag
    Tests the Image Port’s functionality.
    BMP PCI Interface Diag
    Tests the host’s ability to communicate with the BMP via the PCI bus.
    BMP Memory Diag
    Tests all of the on-board memory, including registers.
    BMP Signal Path Diag
    Tests the entire BMP signal path.
    EQ PCI Interface DiagTests the host’s ability to communicate with the registers and memory on the EQ via 
    the PCI bus.
    EQ Memory Diag
    Tests all of the on-board memory, including registers.
    EQ Signal Path Diag
    Tests the entire EQ signal path.
    TD Memory Diag
    Tests all of the on-board memory, for all installed TD boards, including registers.
    TD Access DiagTests all of the Access Modes utilized for Host/TD communication as well as Channel 
    RAM to Rigel transfers.
    XDIF Interface DiagTests all of the XDIF functionality that is testable without a probe or probe simulator 
    attached to the system, including the IIC communication.
    Comprehensive Signal Path DiagTests the signal path of every channel from the TD Rigel Transmitter Memories to the 
    SCB, via both the EQ and the BMP.
    TD Analog Reference Diag
    Tests the AD_REF and VREF voltages on the TD.
    Analog TGC Diag
    Tests the generation of the analog TGC voltage.
    Power Supply Diag
    Tests the system voltages available on the EQ.
    Temperature Diag
    Tests the temperature of the system, wherever available.
    Noise Floor DiagMeasures the noise floor of the system. The noise floor will be calculated form IQ data 
    received without a signal source.
    Loopback Diag
    Tests the analog front-end of the system.
    Acquisition Hardware Information Utility Writes a file for each Acquisition Hardware FRU containing all Vital Product Data.
    Transmit Utility - Start
    Transmit Utility - Change Channel
    Transmit Utility - Stop
    Receive Utility - Single Frame Receives signal on one channel at a time.
    Receive Utility - 8 Frames
    Visual Channel Utility 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 7 Diagnostics/Troubleshooting 7-21
    High Voltage Reset Utility
    Tests whether all TD boards can detect a high voltage fault.
    Set VRF Utility Set the analog VREF to a specified value.
    Set TGC Utility Set the analog TGC to a specified value.
    Probe Select Utility Select a probe connector to be active.
    Probe ID Read Utility Read the Probe ID of all installed probes.
    Front-End Calibration Utility
    Measures the Front End DC Offset, Gain, and Phase. 
    Table 7-2     Acquisition Diagnostics and Utility Listing (Continued)
    Diagnostic/UtilityFunction 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    7-22 Section 7-7 - Acquisition Diagnostics
    7-7-4 Diagnostic Groups
    NOTE: In the simplified block diagrams for the Diagnostic Groups, all unused system components (boards or 
    signals) are ghosted (drawn in gray).
    7-7-4-1 Board-Level SCB Diagnostics
    •All Board-Level SCB Diags: For this board-level Diagnostic Group, the setup simply includes the 
    SCB and the Host. 
    NOTE: The basic wellness of the i960 Intelligent I/O processor is tested indirectly via system start-up 
    and the diagnostics listed below.Figure 7-20  Diagnostic Group Selections - SCB Tests
    Figure 7-21   Board-Level SCB Block Diagram
    XDIF/RFTD sEQBMP
    Scan Control BoardHostPCI
    PCI
    PCI
    IQ,
    RxSYNCBM
    IQ
    IIC
    RFProbeRF
    TD
    CTRLTxSYNC
    RxSYNC
    TxSYNCTxSYNC 
    						