    Chapter 8 Replacement Procedures 8-23
     
    12.) The power supply can now be removed and set on the floor while the last power cable is accessed.
    13.) Disconnect the motherboard power cable, as shown in Figure 8-17 on page 8-23 , by pushing down 
    on the back of the locking key in the center of the connector and pulling the connector out.
    The power supply is now removed from the BEP. Set it aside for disposal.Figure 8-16  Four Power Supply Screws in the Back
    Figure 8-17  Disconnect Motherboard Power Cable
    Motherboard Connector 
    Locking Key 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-24 Section 8-4 - Back End Processor Power Supply Replacement
    8-4-5 Installing the New BEP Power Supply
    Ensure that the On/Off switch on the back of the new power supply is set to the ON position.
    Ensure that the Voltage Select Switch is set to 220 Volts.
    Refer to Figure 8-19 on page 8-24 . If the power supply is going into a BEP1.0 ensure that the +5 Volt 
    standby plug is installed in the new supply power cable. If the power supply is going to be installed in a 
    BEP2.x ensure that this +5 Volt Standby plug is removed.Figure 8-18  Power Supply Voltage Selector Location
    Figure 8-19  +5 Volt Standby Connector
    Voltage Select
    +5 Volt Standby 
    Connector 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-25
    8-4-5 Installing the New BEP Power Supply (cont’d)
    Reverse the removal procedure to install the new BEP Power Supply.
    1.) Install the UPS Control Connector Panel to the back of the BEP (Figure 8-15 on page 8-22 ).
    2.) Attach the motherboard power connector from the power supply to the motherboard (Figure 8-
    17 on page 8-23 ). For BEP2.1, plug the purple +5Volt Standby connector into the cable coming 
    from the DGVIVC Tray.
    3.) Attach the Power Supply to the back of the case using the 4 screws removed in Step #11 on page 
    22, Figure 8-16 on page 8-23 .
    4.) Replace the angle bracket securing the end of the power supply. Loosely fasten the side screw 
    removed in Step #7 on page 21 then replaced and tighten the 2 nuts removed in Step #6 on page 
    21. Finally tighten the side screw.
    NOTE: It is normal to get a spark when this connection is made.
    5.) Attach the 5 power connectors, removed in Step #4 on page 20, to the:
    -CD
    -MO
    -PC2IO
    - HDD
    - PC Bird Card (If system has option)
    NOTE: All five power connectors are exactly alike. Reference the old power supply for cable routing 
    purposes.
    6.) If necessary, neatly bundle the power cables together and ty-wrap them to the power supply as 
    shown in Figure 8-11 on page 8-20 .
    7.) Plug the power cable in the BEP power supply.
    8.) Attach the UPS Battery to the Power Supply UPS Battery Connector removed in Step #5 on page 
    21. 
    9.) Replace the BEP side cover.
    10.) Install all covers removed in the process. (click here to view the cover installation video). 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-26 Section 8-5 - BEP HDD (Hard Disk Drive) Replacement Procedures
    Section 8-5
    BEP HDD (Hard Disk Drive) Replacement Procedures
    8-5-1 Manpower
    One person, 1 hour + travel
    8-5-2 Tools
    Phillips screwdriver
    5.5 mm wrench or nut driver
    8-5-3 Preparations (click here to view the cover removal video)
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
    2.) Remove the left side cover as shown in Section 8-2-2 on page 8-5.
    8-5-4 BEP 1 HDD Replacement (2366650)
    1.) Remove the BEP cover by turning the retaining screw counter-clockwise (Figure 8-20). Pull the 
    back of the cover out and away from the BEP chassis.
    2.) On the bottom right side of the BEP is the HDD. Remove the single screw shown in Figure 8-21. Figure 8-20  BEP Side Cover Removal 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-27
    8-5-4 BEP 1 HDD Replacement (2366650) (cont’d)
    NOTE: This is a little more difficult if the BEP has the 3D Option installed with the 3D card extending 
    over the HDD.
    3.) Grab a hold of the HDD on each side and firmly pull towards you (Figure 8-22).
    4.) The HDD tray will be released from the tabs holding it to the BEP chassis (Figure 8-23). Figure 8-21  Loosen HDD Assembly
    Figure 8-22  Pull HDD Assembly 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-28 Section 8-5 - BEP HDD (Hard Disk Drive) Replacement Procedures
    8-5-4 BEP 1 HDD Replacement (2366650) (cont’d)
    5.) Disconnect the Power and IDE Cables from the back of the HDD.
    6.) Remove the four screws that attach the HDD to the tray.Figure 8-23  Lift Up HDD Assembly
    Figure 8-24  HDD and Tray 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-29
    8-5-4 BEP 1 HDD Replacement (2366650) (cont’d)
    7.) Check the jumper on the back of the new HDD to ensure it is the same as the old one (Figure 8-25).
    The only jumper on the back of the Hard Disk Drive (HDD) should be set to Cable Select (CS). Refer to 
    the diagram on the HDD to ensure that the CS jumper is in place. CS will allow Master/Slave 
    determination by the cable.
    8.) Attach the new HDD to the tray with the four screws. See the orientation in Figure 8-24.
    9.) Attach the power and IDE cables to the new HDD Assembly.
    10.) Slide the assembly into the BEP chassis and ensure the chassis tabs engage the slots on the HDD 
    tray(Figure 8-23).
    11.) Secure the HDD assembly with the single screw in the center (Figure 8-21).
    8-5-5 BEP 2.X HDD Replacement (2348186-31)
    1.) Remove the BEP cover by turning the retaining screw counter-clockwise (Figure 8-26). Pull the 
    back of the cover out and away from the BEP chassis.
    2.) Unplug the Fan connected to the cover and set the cover aside.
    3.) Locate the hard drive under the MOD or CD-RW drive tray and disconnect the power and IDE 
    cables from the drive (Figure 8-27). Figure 8-25  HDD EIDE Jumper
    Figure 8-26  BEP 2.X Cover with Fan 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-30 Section 8-5 - BEP HDD (Hard Disk Drive) Replacement Procedures
    8-5-5 BEP 2.X HDD Replacement (2348186-31) (cont’d)
    4.) Remove the two 5.5 mm nuts that secure the HDD to the MOD/CD-RW tray (Figure 8-28).
    5.) Slide the HDD assembly out towards you and slightly down.Figure 8-27  HDD Power and IDE Cables
    Figure 8-28  HDD Assembly Hardware 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-31
    8-5-5 BEP 2.X HDD Replacement (2348186-31) (cont’d)
    6.) The BEP 2.x HDD should come with the mounting hardware attached. If the replacement HDD does 
    not come with a mounting plate, transfer the mounting plate to the ne HDD.
    7.) Check the jumper on the back of the new HDD to ensure it is the same as the old one (Figure 8-25).
    8.) Slide the new HDD assembly on to the mounting tabs on the bottom of the MOD/CD-RW tray 
    (Figure 8-31).Figure 8-29  HDD Assembly with Tray
    Figure 8-30  HDD EIDE Jumper 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-32 Section 8-5 - BEP HDD (Hard Disk Drive) Replacement Procedures
    8-5-5 BEP 2.X HDD Replacement (2348186-31) (cont’d)
    9.) Secure the assembly with the 5.5 mm nuts removed in step 3.
    10.) Attach the Power and IDE to the back of the HDD. Pay attention to the orientation of the guide pin 
    on the IDE Cable connector and the missing pin on the IDE connection of the HDD.Figure 8-31  HDD Mounting Tabs
    Figure 8-32  IDE Ribbon Connector
    IDE Guide Pin 
    						