    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-163
    Section 8-43
    Rear Crash Bumper Replacement Procedure
    8-43-1 Manpower
    One person, 30 minutes
    8-43-2 Tools
    No special tool are needed
    8-43-3 Rear Crash Bumper Replacement Procedure
    1.) Tear off the bumper.Be aware of the risk of squeezing fingers in Frogleg Assembly.
    2.) Remove old glue from the console.
    3.) Install new bumper. Use glue adhesive to paint and rubber (Araldite 2026).Figure 8-169   Bumper Crash Rear
    CAUTIONBe aware of the risk of squeezing fingers in Frogleg Assembly.
    Rear Crash Bumper 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-164 Section 8-44 - Inner Boss Bumper Removal Procedure
    Section 8-44
    Inner Boss Bumper Removal Procedure
    8-44-1 Manpower
    8-44-2 Tools
    Hexagon key 2,5 mm
    8-44-3 Inner Boss Bumper Replacement Procedure
    1.) Move Top Console in front position. 
    2.) Unscrew 8 hexagon screws.
    3.) Replace the bumper.
    CAUTIONBe aware of the risk of squeezing fingers in Frogleg Assembly.
    Figure 8-170   Frogleg base
    Hexagon screwsHexagon screws Inner Boss Bumper 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-165
    Section 8-45
    Inner Boss Crash Bumper Replacement Procedure
    8-45-1 Manpower
    8-45-2 Tools
    Hexagon key 2,5 mm
    8-45-3 Inner Boss Crash Bumper Replacement Procedure
    1.) Move Top Console in front. 
    2.) Unscrew 4 hexagon screws.
    3.) Remove Bumper brackets.
    4.) Replace the bumpers.
    CAUTIONBe aware of the risk of squeezing fingers in Frogleg Assembly.
    Figure 8-171   Frogleg base
    Hexagon screws
    Bumper brackets
    Bumpers 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-166 Section 8-46 - Upper and Lower Column Cover Replacement Procedure
    Section 8-46
    Upper and Lower Column Cover Replacement Procedure
    8-46-1 Manpower
    2 persons, 3 hours
    8-46-2 Tools
    • Hexagon key, 10 mm, with torque/momentum controller
    • Hexagon keys, 4,5 and 6 mm
    • Phillips screwdriver, size 1 and 3
    8-46-3 Preparations
    1.) Shut down the system as described in Power Off / Shutdown on page 4-3.
    2.) Move Top Console to down position.
    3.) Unscrew 4 screws on the Upper Column Cover.
    4.) Remove monitor as described in Monitor Assembly Replacement Procedure on page 8-92.
    5.) Remove Frogleg as described in Frogleg Replacement Procedure on page 8-159.
    8-46-4 Upper and Lower Column Cover Removal Procedure
    1.) Unscrew 2 Phillips screws on each side of Upper Column Cover.
    2.) Lift Frogleg off its axle.
    3.) Remove column covers.
    8-46-5 Upper and Lower Column Cover Installation Procedure
    1.) Thread on replacement cover(s) in correct order.
    2.) Install Frogleg described in Frogleg Installation Procedure on page 8-161.
    3.) Fasten 4 screws on Upper Column Cover.
    4.) Install Monitor as described in Monitor Replacement on page 8-94. Figure 8-172   Upper and Lower Column Cover
    Screws
    Upper Column Cover
    Lower Column Cover 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-167
    Section 8-47
    Color Video Printer Replacement Procedure
    8-47-1 Manpower
    One person, 0,5 hour + travel.
    8-47-2 Tools
    Long Phillips driver.
    8-47-3 Preparations
    (click here to view the cover removal video)
    Before replacing the UP-D21MD printer: 
    - Go to Utility > Peripherals > Printers
    - Left click on the printer icon and delete the appropriate icon.
    NOTE: Do not delete the UP-D50 icon.
    - After deleting the icons, power-off the system and replace the printer.
    - Power up the system and the printer will work.
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
    2.) Remove the Left and Right Side Covers as shown on Section 8-2-2 on page 8-5.
    3.) Remove the Front Cover as described on Section 8-2-5 on page 8-9.
    8-47-4 Removal Procedure
    1.) Disconnect the cables at the rear of the Video Printer. (RGB & Sync, S-Video, Remote and Power)
    2.) Loosen the two straps that hold the printer. Remove the printer from the system
    Click on the movie camera icon to view the video illustrating the 
    removal and replacement of the color printer. 
    2 minutes and 9 seconds 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-168 Section 8-47 - Color Video Printer Replacement Procedure
    8-47-5 Mounting Procedure
    1.) Slide the new printer in place and connect up the cables.
    2.) Position the printer and fasten the two straps to secure it in place.
    3.) Boot up and test the system.
    4.) Ensure that the print keys still function as they did before.
    5.) Replace the covers. (click here to view the cover installation video)
    Figure 8-173   Front and Rear view of Color Video Printer in System 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-169
    8-47-6 Differences in Replacing the Digital Color Printers
    The major differences in the replacement of the Digital Color Printers is the rear panel cabling.
    For the Sony UPD-50 there is the power cable, USB cable from USB#5 (C20) and the USB to SCSI 
    Adapter.
    Dip switches on the back of the UPD-50 are for SCSI termintaion and SCSI ID. 
    - SCSI termination is set to ON
    - SCSI ID should be set to OFF/OFF/ON (SCSI ID #1). SCSI ID can be set to almost any 
    number. Typically you avoid SCSI ID 0 which is usually the SCSI Controller.
    The smaller Sony UPD-21 has only Power and the USB connection from USB#5 (C20).Figure 8-174  Sony UP-D50 Digital Color Printer
    Figure 8-175  Sony UP-D21 Digital Color Printer 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-170 Section 8-47 - Color Video Printer Replacement Procedure
    8-47-6 Differences in Replacing the Digital Color Printers (cont’d)
    1.) To remove the Sony UP-D21, unfasten the two screws in the Cover Box and lift it out of the top of 
    the system.
    2.) Unfasten the Lock Latch and slide the UP-D21 assembly out the left side of the system.
    3.) Turn the Color Video Printer upside down.
    4.) Detach the printer from the mounting plate loosening the four (4) screws supplied as shown in 
    Figure 4-2.Figure 8-176   Cover Box
    Figure 8-177   Printer Assembly
    Screws
    Latch Lock 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-171
    8-47-6 Differences in Replacing the Digital Color Printers (cont’d)
    8-47-7 UP-D21 Installation Procedure
    1.) Install the mounting plate on the new printer.
    2.) On the left side of the System: Slide the UP-D21MD Printer Assembly into the color printer opening.
    3.) The Mounting Plate has six Latch Clamps under and one Latch Lock at the rear end. These fit in 
    square holes in the system’s frame and locks the UP-D21MD when it’s pushed in position.
    4.) Install the Cover Box and secure with two (2) screws shown in Figure 4-4. Figure 8-178   Sony UP-D21MD Mounting Plate
    Screws
    Latch Clamps 
    Three more 
    under the 
    Printer 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-172 Section 8-48 - Black and White Video Printer Replacement Procedure
    Section 8-48
    Black and White Video Printer Replacement Procedure
    8-48-1 Manpower
    One person, 30 minutes
    8-48-2 Tools
    No special tools needed
    8-48-3 Preparations
    (click here to view the cover removal video)
    Before replacing the UP-D895MD printer: 
    - Go to Utility > Peripherals > Printers
    - Left click on the printer icon and delete the appropriate icon.
    NOTE: Do not delete the UP-D50 icon.
    - After deleting the icons, power-off the system and replace the printer.
    - Power up the system and the printer will work.
    1.) Shut down the system as described in Power Off / Shutdown on page 4-3.
    2.) Remove Left Side Cover as described in Side Covers (with Bumper) Replacement Procedure on 
    page 8-5.
    NOTE: If more working room is desired, remove the right side, filter, top and front covers for improved 
    access.
    8-48-4 Black and White Video Printer Removal procedure
    1.) Loosen the strap that holds the printer in place.
    2.) Disconnect the cables at the rear of the Video Printer and remove it from the system.
    Click on the movie camera icon to view the video illustrating the 
    removal and replacement of the black and whiter printer.
    2 minutes and 1 second 
    						
    All GE manuals Comments (0)