    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-143
    Figure 8-150  Example of Backplane Text File with Hints 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-144 Section 8-35 - Verify and Update Vital Product Data
    Figure 8-151  Backplane Text File 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-145
    8-35-3-4Modify VPD Text Files (cont’d)
    NOTE: The file defaults values will have ?????? in the bar code field, however this must be changed 
    as only alpha-numberic characters are valid for this field when updating the bar code. Some 
    boards may already have data present (no ??????). If this is the case simply verify the data 
    present and correct if necessary.
    3.) Open the *.txt files for the Equalization Board and Backplane
    4.) In each file:
    NOTE: All Bar Code and revision entries MUST be in all uppercase letters.
    * Verify the Product Name Field as the correct circuit board name. If blank enter the data 
    from Table 8-9 on page 8-135.
    * Verify that the Design Group is MKE. If blank enter MKE.
    * Verify the Part Number Field as the correct circuit board part number. If blank enter the 
    data from Table 8-9 on page 8-135.
    * Enter the Functional Revision (PWA Label) of the board (Uppercase alpha character).
    * Enter the Bar Code Serial Number of the board (minumum six character/alpha numeric 
    digits).
    * Enter the Date of Manufacture from the console rating plate: YYYYMMDD format 
    200_ _ _01(Only numeric characters are valid, so this field must be changed). The Day 
    (DD) is always 01.
    * Power On Hours can be left as is or enter 0 if it is blank.
    * Enter the Fab Revision (PWB Label) of the board. Only alpha characters can be entered 
    so convert any numbers to letters (0=A, 1=B, 2=C, 3=D, 4=E, 5=F, 6=G, etc...).
    * Enter the Date of Installation from the console rating plate: YYYYMMDD format 
    200_ _ _01(Only numeric characters are valid, so this field must be changed). The Day 
    (DD) is always 01.
    The Backplane text file actually contains 3 sections.
    - Backplane PWA Information
    - Console Information
    - Back End Processor (BEP) Information
    NOTE: If the Console Serial Number is not correct here, the customer options will NOT work.
    5.) After file changes have been made, SAVE
     and close all text files.
    NOTE: At this time, none of these fields are write protected. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-146 Section 8-35 - Verify and Update Vital Product Data
    8-35-3-5 Acquisition Diagnostics - Write New VPD Data to EEPROM
    Once the *.txt files have been updated with the appropriate information, go back to the Netscape 
    Browser and navigate to the following menu location:
    1.) Acquisition Diagnostics>Diagnostic Groups>Diagnostic Utilities>Acquisition Hardware 
    Information Utilities>Acquisition Hardware Information Write Utilities.
    2.) Separately select the Equalization circuit board and Backplane menu items to write the new 
    information to the EEPROMs. You can check data by reading back the files as performed in 
    Section 8-35-3-3 on page 8-138. Figure 8-152   Acquisition Hardware Information Write Utilities 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-147
    Section 8-36
    Rear Caster Replacement Procedure
    8-36-1 Manpower
    1 person, 30 minutes (one caster)
    8-36-2 Tools
    #2 Phillips screwdriver
    5mm Hexagon wrench
    Bevel Edged Board and Wooden Wedge (both included in package with replacement Casters)
    8-36-3 Preparations (click here to view the cover removal video)
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
    2.) Remove Upper Rear Cover as described in Section 8-2-3 on page 8-7.
    3.) Remove the Lower Rear Cover as described on Section 8-2-6 on page 8-10 and Section 8-2-3 on 
    page 8-7. 
    4.) Remove the Side covers as described in Section 8-2-2 on page 8-5.
    Select the movie camera icon to view the video of the Rear Caster 
    adjustment procedures. This will illustrate a good way to elevate 
    the scanner to replace the casters.
    Figure 8-153   Bevel Edged Board and Wooden Wedge.
    Wooden Wedge
    Bevel Edged Board 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-148 Section 8-36 - Rear Caster Replacement Procedure
    8-36-4 Rear Caster Removal Procedure 
    1.) Turn Front Casters as shown in Figure 8-154.
    2.) Activate Direction Lock.
    3.) Put the bevel-edged board on the floor.
      
    4.) Pull the system backwards up the board incline. This will lift the Rear Wheel on the opposite side 
    of the System from the floor.
    5.) Activate the brakes on all Casters, except the one you intend to replace.
    6.) Turn the Rear Caster that stands on the Bevel Edged Board in the direction as shown in Figure 8-
    155.
    WARNINGWARNINGPrior to elevating scanner, verify that the floating keyboard is locked in its lowest 
    position. 
    WARNINGWARNINGUse extreme caution as long as LOGIQ™ 9 is un-stable, not resting on all four Casters.
    Figure 8-154   Pull system backwards up the board incline 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-149
    8-36-4 Rear Caster Removal Procedure (cont’d)
    7.) The system is now nearly balanced between one Front and one Rear Caster (illustrated in this 
    Procedure; Left Front and Right Rear). 
    8.) Make the System rest on both Front Casters and lift the Rear Caster. Put the Wooden Wedge under 
    the AC Transformer. This will stabilize the System with the Rear Caster free from the floor. Figure 8-155   Activate Rear Brakes
    Figure 8-156   System stabilized with rear Caster free from the floor
    Rear Wheel free 
    from the floor
    Wooden Wedge under 
    AC Transformer 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-150 Section 8-36 - Rear Caster Replacement Procedure
    8-36-4 Rear Caster Removal Procedure (cont’d)
    9.) Remove four (4x) screws fastening Caster.
    10.) Remove Caster
    8-36-5 Rear Caster Installation Procedure
    1.) Replace the Caster and tightly fasten it with four (4x) screws.
     
    2.) Remove the Wooden Wedge.
    3.) Roll the system off the Bevel Edged Board.
    4.) To replace the other Rear Caster, repeat all the steps from Front Caster Replacement Procedure 
    on page 8-151, but now using the Bevel Edged board and the Wooden Wedge on the other side of 
    the system.
    5.) Install the Upper Rear Cover as described in Section 8-2-3 on page 8-7.
    6.) Install the Lower Rear Cover as described on Section 8-2-6 on page 8-10 and Section 8-2-3 on 
    page 8-7. 
    7.) Install the Side Covers as described in Section 8-2-2 on page 8-5. Figure 8-157   Rear Caster fastening
    WARNINGWARNINGUse extreme caution in the next two steps. The scanner is unstable.
    Screws 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-151
    Section 8-37
    Front Caster Replacement Procedure
    8-37-1 Manpower
    1 person, 30 minutes
    8-37-2 Tools
    Torx screwdriver T-20
    Open End wrenches, 10 and 134 mm
    5 mm hexagon key
    Bevel Edged Board and Wooden Wedge (included in package with replacement Casters)
    8-37-3 Preparations
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
    2.) Remove the Front Bumper as described on Section 8-2-7 on page 8-11.
    3.) Turn Front Casters as shown in Figure 8-159.
    4.) Activate Direction Lock.
    5.) Put the Bevel Edged Board on the floor.
      
    6.) Pull the system backwards up the board incline.Figure 8-158   Bevel Edged Board and Wooden Wedge.
    WARNINGWARNINGPrior to elevating scanner, verify that the floating keyboard is locked in its lowest 
    position. 
    WARNINGWARNINGUse extreme caution as long as VIVID 7 is un-stable, not resting on all four Casters.
    Wooden Wedge
    Bevel Edged Board 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-152 Section 8-37 - Front Caster Replacement Procedure
    8-37-3 Preparations (cont’d)
    7.) Turn the Rear Caster on the board in the direction as shown in Figure 8-160.
    8.) Activate both Rear Brakes as shown in Figure 8-160. Figure 8-159   Pull system backwards up the board incline
    Figure 8-160   Activate both rear brakes 
    						