    Chapter 8 Replacement Procedures 8-3
    8-1-2 Returning/Shipping Probes and Repair Parts
    Equipment being returned must be clean and free of blood and other infectious substances.
    GEMS policy states that body fluids must be properly removed from any part or equipment prior to 
    shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment 
    have been properly decontaminated prior to shipment. Under no circumstance should a part or 
    equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or 
    an ultrasound probe). The purpose of the regulation is to protect employees in the transportation 
    industry, as well as the people who will receive or open this package.
    NOTE: The US Department of Transportation (DOT) has ruled that “items that were saturated and/or 
    dripping with human blood that are now caked with dried blood; or which were used or intended 
    for use in patient care” are “regulated medical waste” for transportation purposes and must be 
    transported as a hazardous material. 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-4 Section 8-2 - Plastic Covers and Bumpers
    Section 8-2 
    Plastic Covers and Bumpers
    8-2-1 Overview of Covers and Bumpers
    Table 8-2 
    LEFT-SIDE COVER RIGHT-SIDE COVER
    FRONT COVERFRONT BUMPER
    UPPER REAR COVERLOWER REAR COVER
    TOP COVER
    Click on the movie camera icon 
    to view the video of the Cover 
    removal. 2 minutes and 33 
    seconds 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-5
    8-2-2 Side Covers (with Bumper) Replacement Procedure
    8-2-2-1 Manpower
    One person, 5 minutes
    8-2-2-2 Tools
    Large coin and a large screwdriver
    8-2-2-3 Preparations
    Power Off / Shutdown the system as described in Section 4-3-2 on page 4-3.
    8-2-2-4 Side Cover Removal Procedure
    1.) Use a large coin to turn the two recessed fasteners at the bottom of the side cover a quarter turn to 
    the horizontal position.
    2.) Take hold under the lower front side of cover and pull out gently. At the same time support the rear 
    side of the cover with the other hand.
    NOTE: The two fasteners hold the left-side cover to the scanner at the bottom of the card cage cover.
    The two fasteners hold the right-side cover to the scanner at the bottom of the AC power supply wall.
    3.) When the cover is loose in front, pull the rear out and move upwards.
    4.) Lift the side cover gently upwards until the three hooks at the top of the side cover are free from the 
    three retainer slots in the top cover as shown in Figure 8-1. 
    5.) To Remove Left side cover, repeat this procedure.
    NOTE:Figure 8-1 shows the right-side cover being removed. Repeat the same steps listed above to remove 
    the left-side cover.
    Select the movie camera icon to view the video of the Cover 
    removal procedure.
    Figure 8-1   Removal of Side Covers
    Coin Recessed1/4 turn fasteners
    Pull front side loose first 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-6 Section 8-2 - Plastic Covers and Bumpers
    8-2-2-5 Side Cover Installation Procedure
    (click here to view the cover installation video)
    1.) Guide the top edge of the side cover so the hooks rest in the top cover retainer slots.
    2.) Push back side in first position so it is locked in Lower Rear Cover.
    3.) Push the front side in position so it is locked in Front Cover.
    4.) With a large coin, turn the two bottom fasteners to the vertical position.Now, using a large 
    screwdriver, push in on these fasteners to lock them in place.
    5.) If required, Power On/Boot-up the system as described in Power On / Boot Up on page 4-2. Figure 8-2  Installation of Side Covers
    Top Cover slots Side Cover hooks
    1
    23
    4 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-7
    8-2-3 Upper Rear Cover Replacement Procedure
    8-2-3-1 Manpower
    One Person, 10 minutes
    8-2-3-2 Tools
    Phillips screwdriver
    8-2-3-3 Preparations
    Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
    8-2-3-4 Filter Cover Removal Procedure
    Take hold of the bottom side of the Filter Cover. Push up, pull out and unhook it from the top side and 
    remove it.
    8-2-3-5 Filter Cover Installation Procedure
    (click here to view the cover installation video)
    Align the cover at the top side. Push in and down to secure it along the bottom.
    Select the movie camera icon to view the video of the Cover 
    removal procedure.
    Figure 8-3   Filter Cover
    Installation starts at the upper side
    Filter Cover
    Removal starts from the lower side 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-8 Section 8-2 - Plastic Covers and Bumpers
    8-2-4 Top Cover Replacement Procedure
    8-2-4-1 Manpower
    One person, 30 minutes
    8-2-4-2 Tools
    Phillips Screwdriver
    8-2-4-3 Preparations
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3
    2.) Remove both left- and right-side covers as described in Section 8-2-2-4 on page 8-5.
    3.) Remove upper rear cover as described in Section 8-2-3 on page 8-7.
    8-2-4-4  Removal Procedure
    1.) Move the Operarttor I/O up and forwards to its end positions.
    2.) Unscrew the 10 screws shown in Figure 8-4onpage8-8 .
    3.) Pull the cover gently backwards and upwards to free the cover from the hoist area and allow it to 
    pass freely past the rear handle.
    4.) Lift the top cover away from the scanner chassis.
    8-2-4-5  Installation procedure
    (click here to view the cover installation video)
    1.) Position the top cover in place and fasten it with the 10 screws.
    2.) Mount the side (Section 8-2-2 on page 8-5) and upper rear (Section 8-2-3 on page 8-7) covers.
    3.) If required, Power On/Boot-up the system as described in Section 4-3-1 on page 4-2
    Select the movie camera icon to view the video of the Cover 
    removal procedure.
    Figure 8-4   Top Cover
    Top Cover
    Screws (10)
    Chassis Shelf is 
    Printer Positioning Shelf 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-9
    8-2-5 Front Cover Replacement Procedure
    8-2-5-1 Manpower
    1 person, 20 minutes
    8-2-5-2 Tools
    Phillips Screwdriver
    8-2-5-3 Preparations
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
    2.) Disconnect all Probes and I/O cabling. See User manual. 
    3.) Remove left- and right-side covers as described in Section 8-2-2 on page 8-5.
    4.) Remove any probes attached to the system.
    8-2-5-4 Removal procedure
    1.) Unscrew the 5 Phillips screws that fasten the front cover to the chassis.
    2.) Pull it outwards to free it from the top cover and the front Bumper as shown in Figure 8-5.
    8-2-5-5 Installation Procedure
    (click here to view the cover installation video)
    1.) Position the front cover so it fits between the chassis and the bumper as shown Figure 8-5.
    2.) Position the front cover so it links together with the top cover. You may have to loosen the two 
    Phillips screws on the top cover that are closest to the front cover.
    3.) Fasten the front cover to the chassis with the five Phillips screws.
    4.) Tighten the two Phillips screws in the top cover if they were loosened.
    5.) Install the left- and right-side covers as described in Section 8-2-2 on page 8-5.
    Select the movie camera icon to view the video of the Cover 
    removal procedure.
    Figure 8-5   Removal/Installation of Front Cover
    Screws (5) 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-10 Section 8-2 - Plastic Covers and Bumpers
    8-2-6 Lower Rear Cover Replacement Procedure
    8-2-6-1 Manpower
    One person, 15 minutes
    8-2-6-2 Tools
    Phillips Screwdriver
    8-2-6-3 Preparations
    1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
    2.) Remove both left- and right-side covers as shown in Section 8-2-2 on page 8-5.
    3.) Remove the upper rear cover as shown in Section 8-2-3 on page 8-7.
    8-2-6-4 Lower Rear Removal Procedure
    1.) Unscrew the six Phillips screws that fasten the lower rear cover to the chassis shown in Figure 8-6.
    8-2-6-5 Lower Rear Cover Installation Procedure
    (click here to view the cover installation video)
    1.) Align the lower rear cover with the chassis and fasten it with the six Phillips screws.
    2.) Install both left- and right-side covers as shown in Section 8-2-2 on page 8-5.
    3.) Install the upper rear covers shown in Section 8-2-3 on page 8-7.
    4.) If required, Power On/Boot-up the system as described in Section 4-3-1 on page 4-2.
    Select the movie camera icon to view the mpeg video of the Cover 
    removal and replacement procedure.
    Figure 8-6   Lower Rear Cover with Bumper
    Screws
    Lower Rear Cover
    Screws
    Magnetic lock
    5 - 6 washers 3 - 4 washers 
    						
    							GE MEDICAL SYSTEMS PROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL 
    Chapter 8 Replacement Procedures 8-11
    8-2-7 Front Bumper Replacement Procedure
    8-2-7-1 Manpower
    One person, 30 minutes
    8-2-7-2 Tools
    Phillips size 2
    Torx screwdriver T-20
    8-2-7-3 Preparations
    1.) Shut down the system as described in Section 4-3-2 on page 4-3.
    2.)  Remove Front Cover as described in Section 8-2-5 on page 8-9.
    8-2-7-4 Front Bumper Removal Procedure
    1.) Unscrew 4 Torx screws under the bumper.
    2.) Lift the bumper off.
    8-2-7-5 Front Bumper Installation Procedure
    1.) Put the new bumper in position.
    2.) Fasten the 4 screws firmly.
    NOTE: Do NOT overtighten these screws.
    3.) Install Front Cover as described in Section 8-2-5 on page 8-9. (click here to view the cover 
    installation video)
    Figure 8-7   System Front Bumper
    Screw holes
    Front Bumper 
    						
    							GE MEDICAL SYSTEMSPROPRIETARY TO GE
    D
    IRECTION 2294854-100, REVISION 3  LOGIQ™ 9 PROPRIETARYMANUAL   
    8-12 Section 8-2 - Plastic Covers and Bumpers
    8-2-8 Rear Handle Replacement Procedure
    8-2-8-1 Manpower
    One person, 15 minutes
    8-2-8-2 Tools
    • Phillips screwdriver size 2
    • Hexagon key 5 mm
    8-2-8-3 Preparations
    1.) Shut down the system as described in 4-3-2 Power Off / Shutdown on page 4-3.
    2.) Remove Side Covers as described in Side Covers (with Bumper) Replacement Procedure on page 
    8-5.
    3.) Remove Top Cover as described in Top Cover Replacement Procedure on page 8-8.
    8-2-8-4 Rear Handle Removal procedure
    1.) Remove the two (2) hex-screws on each side of Rear Handle (See Figure 8-8).
    2.) Pull the handle upwards.
    8-2-8-5 Rear Handle Installation procedure 
    1.) Push Rear Handle down its slides. Fasten four (4) hex-screws.
    2.) Install Top Cover as described in Top Cover Replacement Procedure on page 8-8 and Side Covers 
    as described in Side Covers (with Bumper) Replacement Procedure on page 8-5. (click here to view 
    the cover installation video)
     
    Figure 8-8   Rear Handle
    Hexagon Screws 
    						
    All GE manuals Comments (0)