Home > GE > Ultrasound System > GE Vivid 7 User Manual

GE Vivid 7 User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Vivid 7 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 536
    							Scanning Modes
    Vivid7/Vivid7 PRO Users Manual117
    FC092326-03
    Using PW/CW Doppler modes
    Controls and opera-
    tions for PW and 
    CW mode are the 
    same unless other-
    wise noted.There are two ways to start PW/CW Doppler:
    Alter native 1:
    1. Press PW or CW. A scanning screen is displayed with a 
    Doppler cursor on the 2D mode image and a Doppler 
    spectrum in the lower part of the screen.
    2. Use the trackball to position the Doppler cursor line and in 
    PW the sample volume location over the area of interest.
    Sample Volume ad-
    justment may affect 
    the Scale, Frame 
    rate and LV rej. set-
    tings.3. In PW, with the Soft menu rocker key, adjust the sample 
    Volume (SV):
    To enlarge the SV, press the Right arrow of the rocker
    To narrow the SV, press the Left arrow of the rocker.
    Alter native 2:
    1. Press CURSOR on the control panel. A cursor line is 
    displayed on the 2D image.
    2. With the trackball adjust the position of the cursor line.
    3. Press 
    PW or CW.
    Optimizing PW/CW Doppler modes
    Refer to page 400 
    about creating pre-
    setsThe use of preset gives optimum performance with minimum 
    adjustment. If necessary, the following controls can be adjusted 
    to further optimize the PW/CW modes display:
    • Adjust the Active mode gain to set the gain in the spectral 
    Doppler area.
    • Adjust Low velocity reject to reduce unwanted low 
    velocity blood flow and tissue movement.
    •  In PW mode, adjust Sample volume to low setting for 
    better resolution, or higher setting to more easily locate the 
    disturbed flows.
    • Adjust the Compress setting to balance the effect of 
    stronger and weaker echoes and obtain the desired 
    intensity display.
    Frequency and 
    Frame rate settings 
    may affect the Low 
    Velocity Reject.• Adjust Frequency to optimize flow display. Higher setting 
    will improve resolution and the lower setting will increase 
    the depth penetration.
    • Adjust Frame rate to a higher setting to improve motion 
    detection, or to a lower setting to improve resolution. 
    						
    							Scanning Modes
    118Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    The Doppler Power 
    setting affects only 
    Doppler operating 
    modes.•  Adjust Power to obtain an acceptable image using the 
    lowest setting possible. This is particularly important in CW 
    mode, as the energy duty cycle is 100% (constant).
    Adjust the following settings to further optimize the display of 
    the image.
    • Use the Horizontal sweep to optimize the sweep speed.
    •  To view signal detail, use the Velocity range to enlarge the 
    vertical spectral Doppler trace.
    • Use Invert to reverse the vertical component of the spectral 
    Doppler area of the display.
    • Use Angle correction to steer the ultrasound beam to the 
    blood flow to be measured (Not typically required during 
    cardiac studies).
    CAUTIONUse all noise reduction controls with care. Excessive 
    application may obscure low level diagnostic information. 
    						
    							Scanning Modes
    Vivid7/Vivid7 PRO Users Manual119
    FC092326-03
    Tissue Velocity Imaging (TVI)
    TVI over view
     Figure 3-10: The TVI Mode screen 
    1. Color sector marker
    2. Status window:
    3. Soft menu
    Controls marked with R 
    are also available in 
    freeze and cine replay. 
    						
    							Scanning Modes
    120Vivid7/Vivid7 PRO Users Manual
    FC092326-03
     Figure 3-11: The TVI Mode controls on the front panel 
    Tissue Velocity Imaging (TVI) calculates and color-codes the 
    velocities in tissue. The tissue velocity information is acquired 
    by sampling of tissue Doppler velocity values at discrete points. 
    The information is stored in a combined format with greyscale 
    imaging during one or several cardiac cycles with high 
    temporal resolution.
    1. Assignable keys:
    • 2D Width
    • Scale
    • Baseline R
    • Invert R
    • Simultaneous R
    • Dual focus
    • Color maps
    • Cineloop (Freeze 
    only)
    • Q-Analysis (Freeze 
    only)
    ALT
    • Curved AMM R
    • AMM R
    • Tissue Tracking R
    • Strain rate R
    • Strain R
    2. Zoom
    • Display zoom
    • HR zoom
    3. Soft menu
    • Frame rate
    • Compress R
    • Reject
    • Frequency
    • Lateral avg.
    • Radial avg.
    • Baseline R
    • Power
    4. Freeze
    5. TVI
    6. 2D
    7. Alternatives
    Controls marked with R are 
    also available in freeze and 
    cine replay
    4
    6
    7
    1
    2
    3
    5 
    						
    							Scanning Modes
    Vivid7/Vivid7 PRO Users Manual121
    FC092326-03
    TVI controls
    TVI assignable controls
    2D width
    Controls the angular width of the 2D image sector.
    Lower scale value 
    allows greater depth 
    and lower Nyquist 
    limit.Scale
    Adjusts the repetition rate of the Doppler pulses transmitted to 
    acquire the data for color mapping.The Scale value influences 
    the Nyquist limit (the ability to detect maximal velocity without 
    color-aliasing).
    Invert
    Enables the color scheme assigned to positive and negative 
    tissue velocities to be inverted. Invert is available in live and 
    cine replay.
    Simultaneous
    Enables simultaneous display of 2D image and 2D image with 
    TVI color.
    Dual focus
    Activates Dual focus mode. To adjust the Dual focus, press 
    ACT. MODE and rotate the FOCUS assignable.
    Cineloop (in Freeze only)
    Starts cineloop acquisition.
    Color maps
    Displays a menu of color map options. Use the trackball to 
    point a color map and press 
    SELECT to activate a desired color 
    map. Each color map is assigning different color hues to 
    different velocities.
    Q-analysis (in Freeze only)
    Starts the Quantitative analysis application (see Chapter 7, 
    ’Quantitative Analysis’ on page 233). 
    						
    							Scanning Modes
    122Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Alter native assignable controls
    Press ALT to access to the following modes:
    •  Curved Anatomical M-Mode (see page 98)
    •  Tissue Synchronization Imaging (see page 142)
    •  Tissue Tracking (see page 125)
    •  Strain rate (see page 130)
    • Strain (see page 136)
    TVI Soft menu controls
    Frame rate
    Controls the line density. When adjusting frame rate, there is a 
    trade off between spatial and temporal resolution.
    Compress
    Controls the amount of color compression. The color bar is 
    adjusted accordingly.
    Reject (low velocity rejection)
    Adjust the cut-off level for the low velocity of TVI to be 
    discarded when generating the color image.
    Threshold
    Controls the level of greyscale intensity that is used as a 
    threshold for color.
    Transparency
    Controls the degree of transparency of the TVI color.
    Frequency
    Enables the adjustment of the transmission frequency to 
    control the sensitivity or the level of penetration.
    Lateral Averaging
    Use Averaging con-
    trols with caution 
    so as not to obscure 
    significant diagnos-
    tic informationSmooths the image by averaging collected data along the 
    same horizontal line. An increase of the lateral averaging will 
    reduce noise, but this will also reduce the lateral resolution.
    Radial Averaging
    Smooths the image by averaging collected data along the 
    same radial line. An increase of the radial averaging will reduce 
    noise, but this will also reduce the radial resolution. 
    						
    							Scanning Modes
    Vivid7/Vivid7 PRO Users Manual123
    FC092326-03
    Baseline
    Adjusts the color map to emphasize tissue motion either toward 
    or away from the probe. Baseline is available in both Live and 
    Freeze.
    When power is re-
    duced, it reduces 
    the signal-to-noise 
    ratio, so that the im-
    age may appear 
    noisier.Power
    Controls the amount of acoustic power applied in all modes. 
    When power is set to maximum, it is equal to or less than the 
    maximum acoustic power level permitted by regulatory 
    standards. The Thermal Index (TI) and the Mechanical Index 
    (MI) are displayed on the screen. 
    						
    							Scanning Modes
    124Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Using TVI
    1. Select the desired probe.
    2. While in 2D mode press 
    TVI on the control panel.
    Optimizing TVI
    Refer to page 400 
    about creating pre-
    sets.The use of preset gives optimum performance with minimum 
    adjustment. If necessary, the following controls can be adjusted 
    to further optimize the TVI display:
    The Scale value also 
    affects the frame 
    rate. There is a trade 
    off between the 
    frame rate and 
    quantification 
    noise.•  To reduce quantification noise (variance), the Nyquist limit 
    should be as low as possible, without creating aliasing. To 
    reduce the Nyquist limit: Reduce the Scale value from the 
    assignables on the control panel.
    PW will be opti-
    mized for Tissue 
    Velocities when ac-
    tivated from inside 
    TVI.•  TVI provides velocity information only in the beam direction. 
    The apical view typically provides the best window since the 
    beams are then approximately aligned to the longitudinal 
    direction of the myocardium (except near the apex). To 
    obtain radial or circumferential tissue velocities, a 
    parasternal view must be used. However, from this window 
    the beam cannot be aligned to the muscle for all the parts 
    of the ventricle. 
    						
    							Scanning Modes
    Vivid7/Vivid7 PRO Users Manual125
    FC092326-03
    Tissue Tracking
    Tissue Tracking over view
     Figure 3-12: The Tissue Tracking Mode screen 
    1. Color sector marker
    2. Tissue Tracking 
    color bar
    3. Status window
    4. Soft menu
    5. Track start and track 
    end markers
    6. Tracking start and 
    end from R-peak
    Controls marked with R 
    are also available in 
    freeze and cine replay. 
    						
    							Scanning Modes
    126Vivid7/Vivid7 PRO Users Manual
    FC092326-03
     Figure 3-13: The Tissue Tracking controls on the front panel
    Tissue Tracking calculates and color-codes the displacement in 
    the tissue over a given time interval, typically the systole. The 
    displacement is defined as the distance the tissue move during 
    this time interval. The displacement is found as the time 
    integral (sum) of the tissue velocities during this interval.
    Only displacements in the beam direction are found. Only 
    positive (systolic) displacements are mapped into colors, 
    negative displacements are mapped into greyscale.
    1. Assignable keys:
    • Track start R
    • Track end R
    • Track scale R
    • Invert R
    • Simultaneous R
    • Color maps R
    • Cineloop (Freeze 
    only)
    • Q-Analysis (Freeze 
    only)
    MORE
    • 2D width
    ALT
    • Curved AMM R
    • AMM R
    • Tissue Tracking R
    • Strain rate R
    • Strain R
    2. Zoom
    3. Soft menu:
    • Frame rate
    • Transparency R
    • Threshold R
    • Frequency
    • Lateral avg.
    • Radial avg.
    • Power
    • Cine compound 
    (Freeze only)
    4. Freeze
    5. TVI
    6. 2D
    Controls marked with R are 
    also available in freeze and 
    cine replay
    4
    6
    1
    2
    3
    5 
    						
    All GE manuals Comments (0)