    							Getting started
    Vivid7/Vivid7 PRO Users Manual47
    FC092326-03
    Connecting and disconnecting probes 
    The connector panel situated in the front of the unit has three 
    imaging probe ports, one parking port (inactive) and one pencil 
    probe port.
    To connect probes:
    1. Hold the probe connector vertically with the cable pointing 
    upward.
    Probes can be con-
    nected or changed 
    any time while the 
    unit is on.2. Turn the connector locking handle to the horizontal 
    position.
    3. Align the connector with the probe port and carefully push 
    into place.
    4. Rotate the locking handle to the full vertical position to lock 
    in place.
    To disconnect probes:
    1. Rotate the lock handle counter-clockwise to the horizontal 
    position to unlock the connector.
    2. Remove the connector.
    Adjusting the display monitor
    The display monitor’s contrast and brightness controls may 
    need periodic adjustment due to changes in ambient light. They 
    are adjusted from the three buttons on the front part of the 
    display monitor.
    To adjust the contrast of the display monitor
    1. Press the Center button on the display monitor once (see 
    Figure 1-11).
    2. Press the Right button to increase contrast.
    Press the Left button to decrease contrast.
    CAUTIONHandle the probes gently while connecting and disconnecting.
    WARNINGDo NOT touch the patient and any of the connectors on the 
    ultrasound unit simultaneously, including ultrasound probe 
    connectors. 
    						
    							Getting started
    48Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    The amount of contrast is shown on a slider on the screen 
    (see Figure 1-11).
    To adjust the brightness of the display monitor
    1. Press the Center button on the display monitor twice (see 
    Figure 1-11).
    2. Press the Right button to increase brightness.
    Press the Left button to decrease brightness.
    The amount of brightness is shown on a slider on the 
    screen (see Figure 1-11).
     Figure 1-11: Contrast and brightness adjustment
    1. Decrease
    2. Contrast (press once), Brightness (press twice)3. Increase
    123 
    						
    							Getting started
    Vivid7/Vivid7 PRO Users Manual49
    FC092326-03
    Star ting an examination
    Beginning an exam consists of three steps:
    •  Creating a new patient record or starting a new examination 
    from an existing patient record (see below)
    •  Selecting Probe and Application (see page 53)
    •  Start scanning (see page 53)
    Creating a new Patient record or 
    star ting an examination from an 
    existing patient record
    Connectivity on the Vivid 7/Vivid 7 PRO ultra-
    sound unit
    Refer to page 423 
    for further informa-
    tion on connectivity 
    setup.The connectivity on the Vivid 7/Vivid 7 PRO Ultrasound unit is 
    based on the Dataflow concept. A Dataflow is a set of pre-
    configured services (e.g. DICOM services like storage, worklist, 
    verify...etc. or other service types like video print, standard print 
    or messaging). When starting an examination, the user selects 
    a pre-configured Dataflow that will automatically customize the 
    ultrasound unit to work according to the services associated to 
    the Dataflow.
    Star ting an examination
    1. Press NEW EXAM.
    To create an opera-
    tor ID, see 
    page 449.If the unit is password protected a Log In window will 
    appear asking for operator ID, and password.
     Figure 1-12: The Operator login window
    1. Data stored only for 
    the duration of the 
    current 
    examination
    2. Select the operator 
    						
    							Getting started
    50Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    2. Press Log on when completed.
    The Search/Create Patient window is displayed (see 
    Figure 1-13).
    3. Type the patient Last Name, and/or ID.
    The unit can be con-
    figured to automati-
    cally generate a 
    patient ID (see 
    page 437).When default configured, the system automatically 
    searches to see if the patient is already in the database. 
    The result of this search is displayed in the Patient List 
    field.
    If the Patient name is on the patient record list:
    To restrain the 
    search to special 
    category of patient 
    record, press More 
    and use the search-
    ing filters.1. Trackball to the actual patient and double-click the 
    Trackball 
    SELECT key (or press SELECT once and then 
    Select patient).
    The unit is ready for scanning or the Patient information 
    window is displayed (Figure 1-14) depending on system 
    configuration (see page 437).
    If the Patient name is not on the patient record list:
    The automatic 
    search tool display-
    ing matching pa-
    tient information in 
    the Patient list can 
    be turned off (see 
    page 437).1. press Create Patient.
    The unit is ready for scanning or the Patient information 
    window is displayed (Figure 1-14) depending on system 
    configuration (see page 437).
    If the unit is configured to display the Patient information 
    window, follow the steps below:
    1. Enter additional patient information if required.
    Select between cardiac, obstetric, gynecology... etc. to 
    enter application specific patient info (Displayed when the 
    button More is depressed, see Figure 1-14.).
    Press EXAM. LIST to 
    display the previous 
    examinations and 
    diagnosis informa-
    tion for the selected 
    patient.Enter addi-
    tional patient infor-
    mation if required.2. Press Begin exam or any active scanning key to start the 
    examination.
    In the scanning screen, the patient information is displayed 
    on the left side of the Title bar (see Figure 1-15).
    CAUTIONDo NOT use \ or ^ in patient information fields, as these 
    characters might cause problems with some DICOM devices. 
    						
    							Getting started
    Vivid7/Vivid7 PRO Users Manual51
    FC092326-03
     Figure 1-13: The Search/Create Patient window
    1. Press one of the headings to sort the list 
    accordingly.
    2. Select the column heading border and drag to 
    adjust column width.
    3. Select new archive and other pre-defined 
    services.4. Extended menu
    5. The system can be configured to display the 
    Advanced search tool as default (see page 437).
    6. Expended Patient record displaying belonging 
    examinations 
    						
    							Getting started
    52Vivid7/Vivid7 PRO Users Manual
    FC092326-03
     Figure 1-14: The Patient Information window
     Figure 1-15: The Patient information on the scanning screen
    1. The date format is configurable (see page 443).
    2. The window can be configured to display the 
    expanded patient info as default (see page 437).3. The Address field is configurable (see page 437).
    4. Select patient information category. 
    1. The patient 
    information on the 
    scanning screen is 
    configurable (see 
    page 397). 
    						
    							Getting started
    Vivid7/Vivid7 PRO Users Manual53
    FC092326-03
    Selecting a Probe and an Application
    The combination 
    Probe-Application 
    may be user-de-
    fined. See page 400 
    for information on 
    probe/application 
    configuration. Probes and their related applications are selected from the 
    Probes and applications pop-up menus as described below. 
    Only probes currently connected are displayed in the pop-up 
    menu. Only applications appropriate for the type of probe 
    selected are shown.
    To select a probe and an application
    1. Press PROBE on the control panel.
    To change applica-
    tion without chang-
    ing the current 
    probe, press 
    APPL. 
    on the control pan-
    el. 
    A list of the connected probes is displayed.
    2. Trackball to the desired probe.
    3. Press 
    SELECT.
    An Application menu for the selected probe is displayed.
    4. Trackball to the desired application.
    To select a probe 
    with the default ap-
    plication, press 
    SELECT twice on the 
    actual probe.
    5. Press SELECT to launch the application.
    CAUTIONMake sure that the probe and application names displayed on the 
    screen correspond to the actual probe and application selection.
    Check that the correct TI category is displayed (see ’Thermal 
    Index’ on page 471). TIB must be displayed when a fetal 
    application is selected. 
    						
    							Getting started
    54Vivid7/Vivid7 PRO Users Manual
    FC092326-03 
    						
    							Basic scanning operations
    Vivid7/Vivid7 PRO Users Manual55
    FC092326-03
    Chapter 2 
    Basic scanning operations
    This chapter describes basic operations related to scanning. 
    Some operations described in this chapter are fully described 
    in the respective chapters throughout the manual.
    This chapter includes the following information:
    • Assignable keys and Soft Menu Rocker  ..................................... ....  57
    • Using the Soft Menu Rocker  ...................................................... 58
    • Trackball operation  ....................................................................... ....  59
    • Trackball assignment  ................................................................. 61
    • The system menu ....................................................................... 61
    • Cineloop operation  ....................................................................... ....  62
    • Cineloop overview  ...................................................................... 62
    • Cineloop controls  ....................................................................... 64
    • Using cineloop ............................................................................ 65
    • Storing images and cineloops  ..................................................... ....  66
    • To store a single image  .............................................................. 66
    • To store a cineloop ..................................................................... 66
    • Using removable media ................................................................ ....  67
    • Formatting removable media  ..................................................... 67
    • Ejecting removable media  .......................................................... 67
    • Recording images on VCR  ........................................................... ....  69
    • Zoom  .............................................................................................. ....  70
    • To magnify an image (Display zoom) ......................................... 71
    • To activate the HR zoom ............................................................ 71
    • Performing measurements  .......................................................... ....  72
    • Physiological traces  ..................................................................... ....  73
    • Connecting the ECG/Respiration  ............................................... 74
    • Connecting the Phono ................................................................ 75
    • Connecting the Pulse pressure transducer  ................................ 75 
    						
    							Basic scanning operations
    56Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    • Physio overview  .......................................................................... 77
    • Physio controls  ........................................................................... 78
    • Displaying the physiological traces ............................................. 79
    • Adjusting the display of physiological traces  .............................. 80
    • Annotations ................................................................................... .....  82
    • To insert an annotation  ............................................................... 83
    • To edit annotation  ....................................................................... 83
    • To erase annotation  .................................................................... 84
    • Configuration of the pre-defined annotation list  .......................... 84
    • Bodymarks  .................................................................................. 86 
    						
    All GE manuals Comments (0)