    of 536
    							Stress Echo
    Vivid7/Vivid7 PRO Users Manual147
    FC092326-03
    Chapter 4 
    Stress Echo
    This chapter includes the following information:
    • Introduction  ................................................................................... ..  148
    • Selection of a stress test protocol template  .............................. ..  149
    • Image acquisition .......................................................................... ..  150
    • Starting acquisition  ................................................................... 151
    • Continuous capture mode  ........................................................ 155
    • Analysis  ......................................................................................... ..  163
    • Image selection for analysis  ..................................................... 163
    • Scoring acquired loops ............................................................. 165
    • Quantitative TVI Stress echo analysis  ........................................ ..  169
    • Accessing QTVI Stress analysis tools ...................................... 171
    • Vpeak measurement  ................................................................ 171
    • Tissue Tracking  ........................................................................ 175
    • Quantitative analysis  ................................................................ 176
    • Editing/creating a template  .......................................................... ..  177
    • Entering the Template editor screen  ........................................ 177
    • Template editor screen overview  ............................................. 178
    • Editing/Creating a template  ...................................................... 181 
    						
    							Stress Echo
    148Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Introduction
    The Vivid 7/Vivid 7 PRO ultrasound unit provides an integrated 
    stress echo package, with the ability to perform image 
    acquisition, review, image optimization, and wall segment 
    scoring and reporting for a complete, efficient stress echo 
    examination.
    The stress package provides protocol templates for exercise, 
    as well as, pharmacological stress examinations. In addition to 
    preset factory protocol templates, templates can be created or 
    modified to suit users needs. Users can define various quad 
    screen review groups, in any order and combination, that will 
    suit their normal review protocol. When reviewing stress 
    examination images, the images are viewed at their original 
    image quality, and different post-processing and zoom factors 
    may be applied to the images under review for effective image 
    optimization. The protocol template may be configured for 
    Continuous capture. In addition to standard wall motion scoring 
    analysis, the user can perform quantitative stress analysis 
    based on tissue velocity information (TVI).
    A stress echo examination consists of three steps:
    •  Selection of a stress test protocol template (page 149)
    •  Image acquisition (page 150)
    •  Stress analysis (page 163)
    •  Quantitative Stress analysis (page 169) 
    						
    							Stress Echo
    Vivid7/Vivid7 PRO Users Manual149
    FC092326-03
    Selection of a stress test protocol template
    1. Press PROTOCOL to enter the stress echo mode.
    The Protocol screen is displayed (see Figure 4-1) showing 
    the default stress protocol for the current probe.
    To create or edit a 
    template see 
    page 177.2. •To use the current template:
    Turn freeze off to initiate scanning.
    •To use another template:
    Press the assignable 
    TEMPLATE.
    The template list is displayed.
    3. Trackball to the desired template.
    4. Press 
    SELECT.
    5. Turn freeze off to initiate scanning using the new template.
     Figure 4-1: The Protocol screen
    1. Projection selection
    2. Level
    3. Current acquisition
    4. Projection
    5. Group of views 
    						
    							Stress Echo
    150Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Image acquisition
    Images are acquired in a pre-defined order, according to the 
    selected template. The highlighted cell (green) of the matrix, 
    displayed in the Clipboard window indicates which view is 
    currently being acquired (see Figure 4-2). The names of both 
    the view and the level for the current cell is displayed on the top 
    corner of the image area and under the template matrix.
     Figure 4-2: The stress mode acquisition screen
    1. Current view label
    2. Template matrix
    3. Current view (Green 
    cell)
    4. Timers 
    						
    							Stress Echo
    Vivid7/Vivid7 PRO Users Manual151
    FC092326-03
    Star ting acquisition
    To use the Timer, 
    see page 154.1. Turn freeze off to initiate scanning.
    2. Perform a scan that conforms with the view that is 
    highlighted in the template matrix on the Clipboard window.
    Smart Stress is 
    turned on by de-
    fault in factory tem-
    plates. If the selected template has the option Smart Stress 
    turned on (see page 180), a subset of the image 
    acquisition settings for each view in the baseline level will 
    be stored and automatically reused in the corresponding 
    views in the next levels.
    3. Press 
    STORE.
    For further infor-
    mation on stress 
    test configuration, 
    see page 177.•  If the actual stress level is configured to preview 
    cineloop before storing, use the cineloop controls to 
    select the most appropriate heart cycle and, if desired 
    adjust the loop markers (see ’Cineloop operation’ on 
    page 62 for further information). Press 
    STORE to save 
    the selected cineloop.
    •  If the actual stress level is not configured to preview 
    cineloop before storing, the system will automatically 
    store the last cardiac cycle. 
    When storage of the cineloop is completed, the actual 
    highlighted cell in the template matrix displays a 2D icon 
    indicating that the view has been acquired. After storing 
    the loop, the system automatically highlights the next view 
    in the matrix to be acquired.
    For further infor-
    mation on stress 
    test configuration, 
    see page 177.Stress levels can be configured for side by side display/
    comparison of the reference loop from baseline or previous 
    level and the loop to acquire (see Figure 4-3).
    4. Repeat previous steps until all required views are 
    completed.
    The template used can be configured so that analysis is 
    automatically started, displaying the first protocol group. The 
    wall segment scoring diagrams for each view is displayed in the 
    Parameters window on the right side of the screen (see 
    Figure 4-9, page 165).
    CAUTIONIf using DICOM Server dataflow for stress-echo acquisition, 
    images should not be saved to permanent archive before the 
    complete protocol exam is acquired. 
    						
    							Stress Echo
    152Vivid7/Vivid7 PRO Users Manual
    FC092326-03
     Figure 4-3: Display of the Reference loop during acquisition
    Selecting a view during acquisition
    A fixed protocol is provided for scanning, based on the selected 
    template. The system automatically highlights the next view to 
    be acquired in the template matrix, as images are stored. 
    However, the order of scanning may be changed manually as 
    follow:
    Manual selection of a view during acquisition
    1. Use the arrow keys on the alphanumeric keyboard to 
    highlight the cell that represents the view that is to be 
    acquired.
    The selected cell in the template matrix is highlighted in 
    red, indicating non-default position and is blinking if it 
    contains a previously stored acquisition.
    2. Turn freeze off to initiate scanning.
    3. Scan and save the selected loop as explained in the 
    previous section.
    After storage the system automatically highlights the next 
    available view to be acquired.
    1. Current acquisition 
    loop
    2. Corresponding 
    reference loop 
    						
    							Stress Echo
    Vivid7/Vivid7 PRO Users Manual153
    FC092326-03
    Replacing an acquired image
    1. Use the arrow keys on the alphanumeric keyboard to 
    highlight the cell that represents the view that is to be 
    replaced.
    The selected cell in the template matrix is highlighted in 
    red, indicating non-default position.
    2. Turn freeze off to initiate scanning.
    3. Scan and save the selected loop as explained in the 
    previous section.
    After storage the system automatically highlights the next 
    available view to be acquired.
    Moving an acquired image
    An Image can be moved from one cell to another during 
    acquisition. There are two ways to move images:
    Procedure 1
    1. When in the Protocol screen, press 
    MORE (assignable 
    menu).
    2. Press the assignable 
    MOVE IMAGES.
    3. Trackball to the image to move (source cell).
    4. Press 
    SELECT.
    5. Trackball to the destination cell.
    6. Press 
    SELECT.
    The image is moved from the souce cell to the destination 
    cell.
    Procedure 2
    1. In the Protocol screen, trackball to the cell containing the 
    image to move (source cell).
    2. Press and hold down 
    SELECT.
    3. With the 
    SELECT key still depressed, trackball to the 
    destination cell.
    4. Release the 
    SELECT key.
    The image is moved from the source cell to the destination 
    cell.
    Stored images can-
    not be moved.If the destination cell contains an image, the images from the 
    source and destination cells will be exchanged when moving 
    an acquired image. 
    						
    							Stress Echo
    154Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Blood Pressure and hear t rate data
    The blood pressure (systolic and diastolic) and heart rate can 
    be manually entered during acquisition (or during Wall motion 
    analysis).
    1. Press  
    MEASURE.
    The Measurement menu is displayed.
    2. Select Wall Motion.
    The systolic and diastolic blood pressure and the heart 
    rate for the active cell stress-level are displayed on the 
    button over the scoring diagrams (see Figure 4-9, 
    page 165). By pressing this button it is possible to enter 
    blood pressure and heart rate values for all stress levels.
    Timers
    Two timers can be displayed in the Stress mode acquisition 
    screen, beside the template matrix (see Figure 4-4).
     Figure 4-4: The timers in the acquisition screen
    • T1 displays the elapsed time from the start of the stress 
    examination.
    • T2 starts when entering live scanning on the second stress 
    level
    Both T1 and T2 timers can be manually stopped and restarted 
    during the acquisition from the System Menu (Press 
    MENU on 
    the control panel).
    The display of T1 and T2 is user-configurable (see page 177).
    1. Timers display 
    						
    							Stress Echo
    Vivid7/Vivid7 PRO Users Manual155
    FC092326-03
    Continuous capture mode
    Continuous capture mode enables the user to perform 
    acquisition continuously for all views at any level depending on 
    the selected template configuration. Continuous capture 
    consists of temporary saving images acquired in a storage 
    buffer. To enable best possible use of the limited storage buffer 
    capacity, a Pause/Capture mode is provided, as opposed to the 
    normal Freeze/Scan mode. The Pause mode enables scanning 
    and live display on the screen, without any capture, thereby 
    leaving the buffer available.
    To run Continuous capture, the user has to select a template 
    where this feature is activated (see page 177 about template 
    configuration).
    The buffer bar
    When entering a level with Continuous capture enabled, a 
    Buffer bar is displayed in the Info window (see Figure 4-5). The 
    Buffer bar displays the following information:
    •  The units scanning state: 
    • PAUSE (live scanning without storing)
    • CAPTURE (live scanning with storing to buffer)
    •  The percentage of the buffer that is filled
    •  The buffer filling progression showed by a green filling 
    gauge
    •  The capturing sessions, reflected by the red lines along the 
    Buffer bar 
    						
    							Stress Echo
    156Vivid7/Vivid7 PRO Users Manual
    FC092326-03
     Figure 4-5: The buffer bar in Continuous capture
    Controlling the capture process
    When entering a stress level with Continuous capture enabled, 
    the unit is automatically set in Pause mode.
    1. Press 
    IMAGE SRORE or 2D FREEZE to start image capture.
    “Capture” is displayed in the buffer bar, the gauge starts 
    filling and the percentage of filled memory buffer increases 
    (see Figure 4-5, page 156).
    2. Press 
    IMAGE SRORE or 2D FREEZE again to stop capture.
    “Pause” is displayed in the buffer bar.
    When 90% of the memory buffer is filled up, the text display in 
    the buffer bar turns red.
    The unit enters Freeze mode automatically once the buffer is 
    full and the captured loops are displayed in the Continuous 
    capture selection screen (see below).
    1. Scanners state:
    2. Capture session
    3. Pause session
    4. Buffer gauge
    5. Percentage of filled 
    buffer
    2
    345 
    						
    All GE manuals Comments (0)