Home > GE > Ultrasound System > GE Logiq P5 User Manual

GE Logiq P5 User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq P5 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 878
    							Safety Precautions
    LOGIQ 5 Basic User Manual 2-23
    Direction 2300002-100 Rev. 2
    Patient Environmental Devices
     Figure 2-1. Patient Environmental Devices
    100-120V  150VA Max12
    3
    45
    86
    7
    9
    10
    11
    15
    16
    17
    12
    18
    19
    1320
    1421
    22
    23
    24
    25 
    						
    							Safety
    2-24 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Patient Environmental Devices (continued)
    1.  Peripheral Device (Signals I/O Port, Power In)
    2.  Front Panel (Signal I/O Port, Power Out)
    3. Non-Imaging Probes
    4. Imaging Probes
    5. Probe Port
    6. ECG Cable
    7. PCG Sensor
    8. Physio-Signal Input Panel
    9. Rear Panel
    10.  Signals I/O Port
    11 .  P o w e r  O u t
    12.  Signals I/O Port
    13. Footswitch Connector
    14. Power In
    15. Peripheral Devices
    16.  Signals I/O Port
    17. Power In
    18.  InSite Modem (Signal I/O Port)
    19.  Power Telephone Line
    20. Footswitch
    21. Power Line (AC~)
    22. Ground Line
    23.  Power Cable with Protective Earth
    24. MO Drive
    25. CD-RW 
    						
    							Safety Precautions
    LOGIQ 5 Basic User Manual 2-25
    Direction 2300002-100 Rev. 2
    Acceptable Devices
    The devices shown in Figure 2-1 are specified to be suitable for 
    use within the PATIENT ENVIRONMENT.
    Unapproved Devices
    Accessories, Options, Supplies 
    CAUTIONDO NOT connect any probes or accessories without approval 
    by GE within the PATIENT ENVIRONMENT.
    See ‘Peripheral Update for EC countries’ for more information.
    CAUTIONThe unapproved devices shall be used out of patient 
    environment.
    If devices are connected without the approval of GE, the 
    warranty will be INVALID.
    Any device connected to the LOGIQ 5 must conform to one or 
    more of the requirements listed below:
    1.  IEC standard or equivalent standards appropriate to 
    devices.
    2.  The devices shall be connected to PROTECTIVE EARTH 
    (GROUND).
    CAUTIONUnsafe operation or malfunction may result. Use only the 
    accessories, options and supplies approved or recommended 
    in these instructions for use. 
    						
    							Safety
    2-26 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Acoustic Output
    Located on the upper right section of the system display monitor, 
    the acoustic output display provides the operator with real-time 
    indication of acoustic levels being generated by the system. See 
    the Acoustic Output chapter in the Advanced Reference Manual 
    for more information. This display is based on NEMA/AIUM 
    Standards for Real-time Dispaly of Thermal and Mechanic 
    Acoustic Output Indices on Diagnostic Ultrasound Equipment.
    Acoustic Output Display Specifications
    The display consists of three parts: Thermal Index (TI), 
    Mechanical Index (MI), and a relative Acoustic Output (AO) 
    value. Although not part of the NEMA/AIUM standard, the AO 
    value informs the user of where the system is operating within 
    the range of available output. Depending on the examination 
    and type of tissue involved, the TI parameter will be one of three 
    types:
    • Soft Tissue Thermal Index (TIS). Used when imaging soft 
    tissue only, it provides an estimate of potential temperature 
    increase in soft tissue.
    • Bone Thermal Index (TIB). Used when bone is near the 
    focus of the image as in the third trimester OB examination, 
    it provides an estimate of potential temperature increase in 
    the bone or adjacent soft tissue.
    • Cranial Bone Thermal Index (TIC). Used when bone is 
    near the skin surface as in transcranial examination, it 
    provides an estimate of potential temperature increase in 
    the bone or adjacent soft tissue.
    The TI and MI is displayed at all times. The MI display starts at a 
    value of 0.0 and increments in steps of 0.1 while the TI display 
    starts at a value of 0.4 and increments in steps of 0.1 (values 
    less than 0.4 are displyaed as < 0.4). Display precision is ± 0.1, 
    and accuracy is ± 50%. 
    						
    							Safety Precautions
    LOGIQ 5 Basic User Manual 2-27
    Direction 2300002-100 Rev. 2
    Controls Affecting Acoustic Output
    The potential for producing mechanical bioeffects (MI) or 
    thermal bioeffects (TI) can be influenced by certain controls. 
    Direct. The Acoustic Output control has the most significant 
    effect on Acoustic Output. 
    Indirect. Indirect effects may occur when adjusting controls. 
    Controls that can influence MI and TI are detailed under the 
    Bioeffects portion of each control in the Optimizing the Image 
    chapter. 
    Always observe the Acoustic Output display for possible effects. 
    Best practices while scanning
    NOTE:  Refer to the sections of the Optimizing the Image chapter for a 
    complete discussion of each control. 
    Acoustic Output Default Levels
    In order to assure that an exam does not start at a high output 
    level, the LOGIQ 5 initiates scanning at a reduced default output 
    level. This reduced level is preset programmable and depends 
    upon the exam category and probe selected. It takes effect 
    when the system is powered on or New Patient is selected.
    HINTSRaise the Acoustic Output only after attempting image 
    optimization with controls that have no effect on Acoustic 
    Output, such as Gain and TGC. 
    WARNINGBe sure to have read and understood control explanations for 
    each mode used before attempting to adjust the Acoustic 
    Output control or any control that can effect Acoustic Output. 
    Acoustic
    Output
    HazardUse the minimum necessary acoustic output to get the best diagnostic image 
    or measurement during an examination. Begin the exam with the probe that 
    provides an optimum focal depth and penetration.  
    						
    							Safety
    2-28 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Warning Label Locations
    Console Labels
     Figure 2-2. Label Location (a)
    1.  Caution for Possible Injury: “Never put any device onto the 
    Monitor.
    2.  Caution for Hazardous movement (attached inside of OP 
    rear cover).
    3.  Defibrillator Label: “Do not use with Defibrillator” (attached 
    on optional ECG).
    (1)
    (2)
    (3) 
    						
    							Safety Precautions
    LOGIQ 5 Basic User Manual 2-29
    Direction 2300002-100 Rev. 2
    Console Labels (continued)
     Figure 2-3. Label Location (b)
    220-240V  500VA MaxIncluding front printer panel
    100~120V  500VA MaxIncluding front printer panel
    Japan/USA/Asia Console (100~120V) Europe/USA Console (220V)
    (1)
    (3)
    (4)
    (5)
    (6)
    (2)
    (7)
    AC Line Input
    Circuit Breaker
    100-120V~ 50/60HzOn
    Off 1250VA Max
    I
    AC Line InputCircuit Breaker220-240V~ 50/60HzOn
    Off 1250VA Max
    Ior (8)(9) 
    						
    							Safety
    2-30 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Console Labels (continued)
    Table 2-8: Label Location Explanations
    1.  Possible shock hazard. Do not 
    remove covers or panels. No user 
    serviceable parts are inside. Refer 
    servicing to qualified service 
    personnel.
    2.  Do not use the following devices near 
    this equipment: cellular phone, radio 
    receiver, mobile radio transimitter, 
    radio controlled toy, etc. Use of these 
    devices near this equipment could 
    cause this equipment to perform 
    outside the published specifications. 
    Keep power to these devices turned 
    off when near this equipment.
    3.  The equipment weighs approximately 
    180 kg (397 lbs). To avoid possible 
    injury and equipment damage when 
    transporting from one area of use to 
    another: 
    •  Be sure the pathway is clear.
    •  Limit movement to a slow careful 
    walk.
    •  Use two or more persons to 
    move the equipment on inclines 
    or long distance.4.  Prescription Device (For U.S.A. Only)
    5.  The CE Mark of Conformity indicates 
    this equipment conforms with the 
    Council Directive 93/42/EEC.
    6. CISPR  CAUTION: The LOGIQ 5 
    conforms to the CISPR11, Group 1, 
    Class A of the international standard 
    for Electromagnetic disturbance 
    characteristics. 
    7.  Voltage Range (Indication label)
    8.  Circuit Breaker location and rating
    9.  Signal ground point label CAUTION: 
    This is only for “FUNCTIONAL 
    GROUNDING”, NOT “PROTECTIVE 
    EARTH”. 
    						
    							Safety Precautions
    LOGIQ 5 Basic User Manual 2-31
    Direction 2300002-100 Rev. 2
    Console Labels (continued)
     Figure 2-4. ETL and Identification/Rating Plate Label Location (c)
    Korea 220V Console
    (1)(2)
    (3)
    (4)
    (5) (6)
    GE MEDICAL SYSTEMS KOREA LTD.
    LOGIQ 5
    1250VA
    1250VA
    1250VA120V5
    GE Medical Systems Korea GE MEDICAL SYSTEMS KOREA LOGIQ 5 
    						
    							Safety
    2-32 LOGIQ 5 Basic User Manual
    Direction 2300002-100 Rev. 2
    Console Labels (continued)
    1.  ETL Label: NRTL Listing and Certification Mark is used to 
    designate conformance to nationally recognized product 
    safety standards.
    2.  Identification and Rating Plate–USA/Asia 120V Console.
    3.  Identification and Rating Plate–Europe/Asia/Latin America 
    220V Console.
    4.  Identification and Rating Plate–Japan 120V Console.
    5.  Identification and Rating Plate–Korea 220V Console. 
    						
    All GE manuals Comments (1)

    Related Manuals for GE Logiq P5 User Manual