    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 67
    3-9-14-6      DICOM Storage Properties Adjustments (cont’d)
    DICOM SR
    If SR is enabled, all M&A that is supported by DICO M is sent to the DICOM server (in a separate job).
    Allow SR Private Data (introduced with software versions v6.0, v5.2 and v4.2.)
    By default, this function is disabled.
    Allow SR Private Data  is intended for receivers that can handle “Private Data”, like the CA-1000.
    In the examples in Figure 3-67 and Figure 3-68 (below),  Allow SR Private Data has been enabled.
    Figure 3-67   Dicom Storage properties Figure 3-68   Dicom Media properties 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 68 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-14-6      DICOM Storage Properties Adjustments (cont’d)
    Retry
    • Max # - The number of times the  software will try to connect if the connection doesn’t succeed the 
    first time(s).
    • Interval - The interval in seconds between each attempt to connect.
    Smiley
    Select the Smiley button to ping th e server. If the network connection  is OK, the Smiley will smile, if the 
    connection fail, the Smiley will turn sad.
    NOTE: The smiley checks if the remote server is  accessible (ping). It is not a DICOM Echo (DICOM 
    ping), so it does not check AE title or port number.
    Reopen per Image
    If Reopen per image is enabled, Vi vid 7 will create a new connection (asso ciation) for each image. This 
    may be useful for DICOM servers that do not acce pt different image types in the same association.
    Storage Commitment
    If you want to enable Storage Co mmitment, check the check-box an d select the button “Storage 
    Commitment”. 
    In the dialog, enter IP address, Name, AE Title and Port Number.
    Figure 3-69   Storage Commitment 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 69
    3-9-14-6      DICOM Storage Properties Adjustments (cont’d)
    MPPS
    If you want to enable MPPS, check  the check-box and select the button “MPPS”. In the dialog, enter IP 
    address, Name, AE Title and Port Number.
    Time-out
    Adjust Time-out setting
    Introduced with software versions v6.0, v5.2 and v4.2:  The retry settings can be used to make jobs 
    retry on bad networks. There is no need to set retries for mobile (off-line) use when these software 
    versions are used.
    Software v5.1.x/v4.1.x and lower:  The retry settings for DICOM Storage are by default for mobile (off-
    line) use. They ensure that  jobs in the spooler will be sent automatically if the user reconnects within the 
    total retry period. If the customer does not have the  mobile case, you can reduce the number of retries 
    and the interval. This will make the jo b fail faster if there is a network or server problem (and let the user 
    forward the job to another server if desired).
    OK/Cancel
    •Select  OK to save all changes. The changes will take  place after the scanner has been restarted.
    •Select  Cancel to return without  saving any changes.
    Figure 3-70   MPPS 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 70 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-15 Export Configuration
    The destination for Export of patient records to Exce l and MPEG must be configured prior to use. This 
    is done from the  Dataflow screen.
    3-9-15-1 Setup on the Remote Share Required setup on the remote share (remote server):
    1.) Add user and password.Predefined user:  E1c2h3o4C5l6i7e8n9t
    Predefined password:  u1l3t5r7a
    2.) Set Share permissions.
    3.) Set security permissions. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 71
    3-9-15-2 Display the Dataflow Screen Follow these steps to display the Dataflow Screen:
    1.) Press CONFIG (F2)  and log on as administrator.
    2.) Select  Connectivity  > Dataflow .
    The  Dataflow sheet  is displayed, see  Figure 3-71 The Dataflow sheet on page 3-71 .
    3.)
    Select the dataflow  Misc. Export in the Name pull-down menu.
    Figure 3-71   The Dataflow sheet
    Software version v4.x/v5.x Software version v6.x: 
    A new check-box,  Show Only Imp/Export  
    has been added. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 72 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-15-3 Export to Excel Configuration 1.) Select the  Excel storage  device in the Selected devices pane  and press Properties .
    The  Excel properties window  is displayed.
    2.) Select a removable media or a network volume re mote path as the destination in the Destination 
    pull-down menu.
    NOTE: Remote paths for network volumes must be entered once in the Remote path field before they  can be selected from the Destination Pull-down menu.
    3.) Select OK.
    4.)Press CONFIG (F2) if you want to quit Config.
    Figure 3-72   The Excel properties window
    EXCEL STORAGE
    PROPERTIES 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 73
    3-9-15-4 Export to MPEGVue Configuration 1.) Select the  MPEGVue device in the  Selected devices  pane and press  Properties .
    The  MPEGVue properties window  is displayed.
    2.) Select a removable media or a network volu me remote path as the destination in the Destination 
    Pull-down menu .
    NOTE: Remote paths of network volumes must be enter ed once in the Remote path field before they 
    can be selected from the Destination Pull-down menu.
    3.) Check the options as required.
    4.) Select OK.
    5.) Press CONFIG (F2) if you  want to exit Configuration.
    Figure 3-73   The MPEGVue properties window 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 74 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-15-5 eVue Configuration 1.)
    Press CONFIG (F2) and log on as administrator.
    2.) Select the  Connectivity category and  Dataflow subgroup. The Dataflow sheet is displayed.
    3.) Select the dataflow  Local Archive - Int. HD/eVue  in the Name pull-down menu.
    4.) Select the  eVue device in the  Selected devices  pane and press Properties. The eVue properties  
    window is displayed.
    5.) Select a removable media or a network volu me remote path as the destination in the Destination 
    pull-down menu.
    NOTE: Remote paths of network volumes must be enter ed once in the Remote path field before they 
    can be selected from the Destination Pull-down menu.
    6.) Check the options as required.
    7.) Select  OK and press  CONFIG (F2) .
    Figure 3-74   The eVue dataflow
    Figure 3-75   The eVue Properties 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 75
    3-9-16 Create a New Dataflow
    3-9-16-1 Overview It is possible to make new dataflows by combining  the predefined settings. The table below describes 
    the legal combination of inputs and outputs in a dataflow.
    3-9-16-2 Create a New Dataflow with Worklist, Database and DICOM Storage (Introduced for software v4.0)
    To create a new dataflow with Work list, Database and DICOM Storage,  do the following in exactly these 
    steps:
    1.) Add the Worklist Input, highlight it, select proper ties, and name it (i.e.: DR PACS Worklist or just 
    Worklist).
    Go to the IP Address field, scro ll down and select . This  will allow you to edit or add IP 
    Addresses (New Destinations/devices).  If the correct IP Address is already in the list, just select it.
    2.) Add the Database Input/Output, highlight it, select  properties, and name it (i.e.: Local Archive - Int 
    HD).
    Go to the IP Address field and (if the Database is to be the Local Archive) verify that My Computer 
    is selected.
    The IP Address should then be correct and the Database Name and Archive File Location entries 
    should auto-populate correctly.
    Table 3-16    Allowed combinations of inputs and outputs in a dataflow
    NOTICEFailure to add destinations to the 
    dataflow in the correct order may or will caus e intermittent errors.
    Output  
    Input  
    No 
    Output
    Database
    Remote
    Database
    DICOM
    MOD5.25DICOM
    MOD3.5
    DICOM
    StorageDICOM  Print
    Database  +
    DICOM MOD
    5.25/3.5Database  +
    DICOM
    Storage
    Remote 
    Database  +
    DICOM MOD
    5.25/3.5Remote 
    Database  +
    DICOM
    Storage
    Database
    XXX
    Remote DatabaseX XX
    DICOM WorklistX
    DICOM CDX
    DICOM MOD 5.25X
    DICOM MOD 3.5X
    Query/RetrieveX
    EchoPAC MacX
    Worklist/databaseXX
    Worklist/rem databaseXX
    No input deviceXX 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 76 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-16-2      Create a New Dataflow with Worklist, Database and DICOM Storage (cont’d) 3.) Add the DICOM Storage Output, highlight it, select properties, and name it (i.e.: DR PACS Store).Go to the IP Address field, scroll down and select  . This will allow you to edit or add IP 
    Addresses (New Destinations/devices).
    If the correct IP Address is alre ady in the list, just select it.
    If you create the Dataflow in any  other order or remove and add an  entry later, you may or will receive 
    “you are not connected to the Data base” errors when trying to create a patient locally. The Worklist will 
    work, however they can NOT save patients “created ” on the Vivid 7 to the Archive while using that 
    dataflow.
    You can change the information on each entry in a Dataflow as long as you do NOT remove or add to 
    the Dataflow out of sequence.
    You should also set Direct Search to “All patients” for the new dataflow. 
    						