    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 59
    7-7-3-3 Monitor Test Double-click Monitor Test  from the PC Diagnostic interactive tests menu.
    The monitor test dialog box presents you with three options:
    • Combination Test  - Checks the monitor for alignment, co lor depth, and monitor resolution. It will 
    put four cross-hairs in each corner, RGB (Red Green Blue) spectrum arrays and horizontal and 
    vertical alignment bars on the screen.
    NOTE: In most cases, you will ch oose the Combination Test.Figure 7-62   Monitor Test
    Figure 7-63   Monitor Test 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 60 Section 7-7 - Common Diagnostics
    7-7-3-3      Monitor Test (cont’d) •Solid Color Test  - The solid color test is for laptop and  LCD (Liquid Crystal Display) panel displays. 
    This test displays solid color screens to allow yo u to make sure all color pixels are operating 
    correctly.
    • VESA (Video Electronics Standard s Association) Test Patterns - The VESA test patterns 
    checks luminance, geometry,  and focus. The VESA tests are designed for professional 
    measurement use.
    Luminance
    Figure 7-64   Moni tor, VESA Test Pattern, Luminance 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 61
    7-7-3-3      Monitor Test (cont’d)
    Geometry Figure 7-65   M onitor, VESA Test Pattern, Geometry 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 62 Section 7-7 - Common Diagnostics
    7-7-3-3      Monitor Test (cont’d)
    Focus
    Always shutdown the system and reboot after a diagnostics session. Figure 7-66   Monitor, VESA
     Test Pattern, Focus 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 63
    Section 7-8 
    Acquisition Diagnostics
    7-8-1 Acquisition Di agnostics Introduction
    • If the scanner is running, but an error that may be a hardware failure, has been registered, then 
    continue to read. 
    7-8-1-1 Start System Test To start the System Test,  power up the scanner, see  4-2-2 - Power ON/ Boot UP  on page 4-3 for detailed 
    instructions.
    After scanner initializat ion is complete, continue with the steps below;
    1.) Press  CONFIG
     (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    2.) Select  SYSTEM
     from the bottom of the menu that appears on the monitor.
    3.) Select  TEST
    .
    .
    4.) Select   GO TO SYSTEM TEST to select the Calib ration/Test screen.
    Figure 7-67   System > Test selection screen 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 64 Section 7-8 - Acquisition Diagnostics
    7-8-2 Calibration Screen
    After following the steps in  7-8-1-1 Start System Test on page 7-63 , the default screen will be the 
    Calibration screen. 
    The lower part of the Calibration  screen is different on units with Front-End Processor version 1 (FEP1) 
    and Front-End Processor version 2 (FEP2).
    7-8-2-1 Functions Available from the Calibration Screen These functions are available from the Calibration screen.
    • DC Offset Calibration (DC Offs. Cal) of  the A/D converters in the Front-End.
    • Reset the calibration (Cal. Reset) to the factory default setting.
    • Show a graph regarding the calibration (Show graph).
    • Deselect/Select channels are used in manufactur ing to isolate noisy channels (Channel Control).
    With a linear probe connected to the system, it is po ssible to disconnect groups of channels either 
    on the TX Board or on the RX Board to isolate the noise source.
    • The buttons in the lower part of the  screen are used for board testing (see 7-8-3 - System Test on 
    page  7-66). 
    The number of buttons in the lower  row are different for units with FEP1 and FEP2. The illustration in 
    Figure 7-68  is the Calibration Screen on units with FEP1.  
    Figure 7-69 Calibration Screen on units with FEP2 on page 7-65  illustrates the Calibration Screen on 
    units with FEP2. Figure 7-68   Calibration Dialog on units with Front-End rack version - 1 (FEP1)
    Channel Control
    (used for factory 
    tests)
    DC Offs. Cal Cal. ResetShow graph
    “LEDs”
    Gray - Test not done
    Green - Test Passed
    Red - Test failed 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 65
    7-8-2-1      Functions Available from the Calibration Screen (cont’d) To start a test, select  the respective button.
    The “LEDs” in the Calibration Screen will  change color when a test has been done;
    • GREEN indicates that the test PASSED.
    • RED indicates that the test FAILED. Figure 7-69   Calibration Screen on units with FEP2
    Channel Control
    (used for factory 
    tests)
    DC Offs. Cal Cal. ResetShow graph
    “LEDs”
    Gray - Test not done
    Green - Test Passed
    Red - Test failed 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 66 Section 7-8 - Acquisition Diagnostics
    7-8-3 System Test
    7-8-3-1 Select the Test Screen From the Calibration  screen, select  TEST
     or select one of the buttons (not  CAL) in the lower part of 
    the screen. See  Figure 7-70 - Select Test Window (FEP1)  and Figure 7-71 - Select Test Window (FEP2).
    The  System Test  screen (Figure 7-72 /Figure 7-73 ) gives you access to the tests needed for 
    troubleshooting the Front-End Card Rack. Figure 7-70   Select Test Window (FEP1)
    Figure 7-71   Select Test Window (FEP2)
    Press 
    TEST 
    or press one of these 
    buttons
    Press  TEST 
    or press one of
     these buttons 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 7 - Diagnostics/Troubleshooting 7 - 67
    7-8-3-2 The System Test Screen - FEP1 Figure 7-72   System Test Screen (FEP1)
    Table 7-4    Choices on the System Test Screen on units with FEP1
    ITEMDESCRIPTIONCOMMENTS
    1 MODULE NAME ETC.
    2 SYSTEM:TEST ALL See  7-8-3-5 Test Loops, System on page 7-69
    3 TEST LOOPS See  7-8-3-5 Test Loops, System on page 7-69
    4BREAKSee 7-8-3-5 Test Loops, System on page 7-69
    5“BLOCK DIAGRAM” WITH SINGLE TESTSSELECT THE CARD YOU WANT TO TEST, see  7-8-3-7 “Block Schematics” with 
    Single Tests on page 7-71
    6STATUS FIELD
    7 PROGRESS FIELD
    8 STATUS LIGHT RED = FAILED, GREEN = PASSED
    9 DATA FIELD DISPLAYS INFORMATION ABOUT THE TESTS
    10MODULE SPECIFIC TESTSSEE 7-8-3-9 Relay Board Test on page 7-72
    11MODE SELECT: TEST OR CALIBRATION SELECT  CALIBRATION TO GO TO THE CALIBRATION SCREEN, SEE  7-8-2 
    Calibration Screen on page 7-64
    1
    2
    3
    45
    67
    8
    910
    11 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    7 - 68 Section 7-8 - Acquisition Diagnostics
    7-8-3-3 The System Test Screen - FEP2 Figure 7-73   System Test Screen (FEP2)
    Table 7-5    Choices on the System Test Screen on units with FEP2
    ITEMDESCRIPTIONCOMMENTS
    1 MODULE NAME ETC.
    2 SYSTEM:TEST ALL See  7-8-3-5 Test Loops, System on page 7-69
    3 TEST LOOPS See  7-8-3-5 Test Loops, System on page 7-69
    4BREAKSee 7-8-3-5 Test Loops, System on page 7-69
    5“BLOCK DIAGRAM” WITH SINGLE TESTSSELECT THE CARD YOU WANT TO TEST, see  7-8-3-7 “Block Schematics” 
    with Single Tests on page 7-71
    6STATUS FIELD
    7 PROGRESS FIELD
    8 STATUS LIGHT RED = FAILED, GREEN = PASSED
    9 DATA FIELD DISPLAYS INFORMATION ABOUT THE TESTS
    10MODULE SPECIFIC TESTSSEE 7-8-3-9 Relay Board Test on page 7-72
    11MODE SELECT: TEST OR CALIBRATION SELECT  CALIBRATION TO GO TO THE CALIBRATION SCREEN, SEE  7-8-2 
    Calibration Screen on page 7-64
    1
    2
    4
    5
    8
    9
    3
    67
    10
    11 
    						