    Chapter 3 - System Setup 3 - 117
    3-11-8      Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d)
    8.) Select the  DATAFLOW
     tab to view the Dataflow Services window.
    9.) Select  Remote Archive - Remote HD  on the drop-down menu to see  the contents of that dataflow. 
    Figure 3-126   Dataflow Services Window
    Figure 3-127   Obtaining Dataflow Contents Overview
    Dataflow 
    selection 
    area
    Dataflow 
    Services 
    area
    DATAFLOW tab
    Locale Archive -Internal 
    HD service
    DATAFLOW tab
    Dataflow Name drop-
    down menu
    Remote Archive -Remote HD 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 118 Section 3-11 - Connectivity Setup - Software v2.x
    3-11-8      Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d) Figure 3-128   Remote Archive - Remote HD Dataflow Contents Overview
    Remote Archive -
    Remote HD 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 119
    3-11-8      Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d)
    10.) Select the  SERVICES
     tab (located near the top of the window).
    11.) Select the EchoPAC PC Workstation you want to connect to, from the “Name” drop-down menu. In  this example, it is called “EchoPAC7-000000”.  Figure 3-129   Dataflow Services
    Figure 3-130   Dataflow Services, Choices and Setup
    Name drop-down 
    menu
     
    SERVICES
     tab
    Local Archive - Local 
    Storage
    Select Server
    Name drop-down 
    menu 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 120 Section 3-11 - Connectivity Setup - Software v2.x
    3-11-8      Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d) 12.) Verify that the IP-address is correct, in dicating that you have selected the right unit.
    If the IP number is wrong, it means that yo u are addressing the wrong device (computer). 
    Correct the address and do a new verification before you continue. 
     
    Figure 3-131   Verify/change IP address to selected device.
    IP address for selected 
    device 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 121
    3-11-8      Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d)
    13.) In the list of “Services”, select “Remote Archive - Remote Storage”. 
    14.) In the list of  Services select Remote Archive -  Remote Storage. 
    15.) Change the  Corresponding sql.ini name  to reflect the name of the ECHOPAC7-nnnnnn 
    Workstation. The  Corresponding sql.ini name  is a combination of the host name plus a space and 
    the two letters “DB”. 
    Example:  If the EchoPAC PC name is  ECHOPAC7-987654, the Corresponding sql.ini name  will be ECHOPAC7-
    987654 DB.
    16.) Change the  Placement of Image Files  and Placement of Report Files  names so they reflect the 
    name of the EchoPAC  PC Workstation. 
    Example:  If the EchoPAC PC name is  ECHOPAC7-987654, the Placement of Image Files  and Placement of 
    Report Files  are both ECHOPAC7-987654\Archive.
    This completes the EchoPAC PC connection setup. 
    NOTE: Remember that these settings have no effect  until you have logged on to the dataflow. 
    It may be necessary to log on and log off  this dataflow before the settings work.
    Figure 3-132   Dataflow services Choice and Setup
    Remote Archive - 
    Remote Storage
    Corresponding sql.ini 
    name Storage location for 
    Image Files
    Storage location for 
    Report Files 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 122 Section 3-11 - Connectivity Setup - Software v2.x
    3-11-9 HL7 Communication Se tup for Software v2.x
    3-11-9-1 Introduction
    The procedures below tell you how to set up the Vivid 7 workstation so it can connect to the Hospital 
    Information System (HIS), via th e Vivid HL7 Gateway. By connec ting to the HIS, demographic 
    information i.e Patient ID, Name, Gender etc. can  be pulled from the HIS, reducing “double-work” and 
    typing errors. Patient Reports can be sent back to the  HIS for storage together with the rest of the Patient 
    Records.
    3-11-9-2 Preparations 1.) Before you can set up the HL 7 Communicat ion, you need to collect some information. 
    - The IP address for  the Vivid HL7 Gateway
    - Port Number used by the Vivid HL7 Gateway (Default port: 3320)
    Typical source for this  information is the Mitra support personnel.
    Figure 2-2 Pre-quote Wo rksheet for DICOM Network  Information on page 2-14 can be used for 
    gathering this information.
    2.) Provide the computer name of  each EchoPAC PC Workstation and  Vivid 7 being configured to the 
    Mitra support personnel.
    3-11-9-3 Configuration of Modality Worklist Services
    Identify Dataflow to be Used
    1.) Identify which dataflow is to be used at the si te. If they will be using DICOM Modality Worklist, then 
    the dataflow must include a Worklist service.
    2.) On the Connectivity screen, select the Dataflow tab. 
    3.) Select the name of the sites dataflow from the pull-down.
    4.) In the Services section, locate the Worklist service. 
    5.) Record the services name  and destination device.
    Configure the DICOM Worklist Service
    1.) Select the Services tab. 
    2.) Select the Worklist destination  device from the pull down menu.
    3.) Enter the IP address for the Vivid HL7 Gateway.
    4.) Enter a descriptive name in the Name field. Exam ple: Vivid HL7 Gateway. The use of this field is 
    optional).
    5.) Enter AE Title. Use the Computer  Name for the unit you are configuring.
    6.) Enter Port Number. Port 33 20 is default, but Mitra may have  provided another port number.
    7.) Set Time-out to 90.
    8.) Leave the rest of the fields unchanged. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 123
    3-11-9-3      Configuration of  Modality Worklist Services (cont’d)
    Configure the Search Criterias
    NOTE: This configuration is required. 1.) Select the Search Criterias button.
    2.) Select the criteria Modality and select the Remove button.
    3.) Press the OK button.
    Configure Search Criterias:  Constrain Query by AE Title
    NOTE: This configuration is optional. It should only be  performed if the site is directing orders to the 
    Vivid 7 that is scheduled to perform the study. 
    Please contact the Mitra support personnel  to verify that this step is needed.
    1.) In the Search Criterias dialog box, select 00400001 Scheduled Station AE Title from the Select Tag  pull down menu.
    2.) Enter the Vivid 7 computer name for the Value.
    3.) Press the Add to List button.
    4.) Press the OK button.
    Configure Search Criterias: Constrain Query by MRN
    NOTE: This configuration should be per formed if the site will need to constrain the Worklist queries by 
    MRN (patient id)
    1.) In the Search Criterias dialog box, select 00101 000 Other Patient IDs from the Select Tag pull down 
    menu.
    2.) Enter a 9 digit number for the Value.
    3.) Select the Add to List button.
    4.) Select the OK button.
    Assign Dicom Worklist name to Dataflow
    1.) Select the Dataflow tab. 
    2.) From the Dataflow name pull down menu,  select the Dataflow which will be used.
    3.) Remove the Dicom Worklist service  currently used in the Dataflow.
    4.) From the Select Service pull down menu,  select the Gateway Modality Worklist service.
    5.) Select the Add button.
    6.) Ensure that the service is displaying the value In  for the Dir parameter and that its role is Primary. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 124 Section 3-11 - Connectivity Setup - Software v2.x
    3-11-9-3      Configuration of  Modality Worklist Services (cont’d)
    Test Connectivity with the Vivid HL7 Gateway
    The purpose of this test is to verify that the  Vivid 7 has connectivity with the Vivid HL7 Gateway.
    1.) From either the Current or Dataflow view, expand the Dataflow incorporating the Gateway Modality  Worklist service.
    2.) Expand the Gateway destination device.
    3.) Select the Gateway Modality Worklist service. 
    4.) Select the Check button. - If connectivity was achieved, then there should  be a green check next to the name of the 
    Gateways Modality Worklist service.
    - If connectivity was not achieved, then there will  be a red X next to the name of the Gateways 
    Modality Worklist service.
    NOTE: If the Check button fails directly, the AE Title is probably wrong. If the Check button fails after a long time (corres ponding to the timeout), the IP address or Port 
    Number is probably wrong.
    Troubleshoot Connectivity with the Gateway
    Typical causes if connection between Vivid 7 and HIS failed:
    1.) No network connection.
    2.) IP address and/or port number was incorrectly entered when configuring the Gateway Modality  Worklist service.
    3.) Mitra support personnel have  not configured the Gateway yet to accept Modality Worklist queries 
    from the Vivid 7.
    4.) Verify that the Vivid 7 computer name matc hes the value specified for the Gateway Modality 
    Worklist service and that it also matches the  value supplied to the Mitra support personnel.
    Configure the Export to HL7 Path
    1.) Enter the configuration  screen for the EchoPAC PC. 
    2.) Press the Connectivity button. 
    3.) Select the Tools tab.
    4.) In the section labeled Export Paths, make the following changes:
    a.) For the Export To HL7 path,  enter the following information:
    \\ < Name or IP Address of  the Vivid HL7 Gateway>\POLL_DIR
    Example:  \\Gateway\POLL_DIR
    b.) Verify that Text is select  for the Export file format. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 125
    3-11-9-3      Configuration of  Modality Worklist Services (cont’d)
    Test Connectivity with the Vivid HL7 Gateway
    1.) Start a new study and acquire an  image, or use an existing study.
    2.) In the Measurements mode, make some of the measurements.
    3.) In the Report screen, cr eate and store a report.
    4.) In the Patient screen, select the study that you  created the report in from the Examination List.
    5.) From the More pull down menu, select HL7 Results.
    6.) If the export was successful, then you should get the message:  Export exam status: OK.
    7.) If the export was not successful, then you should get the message, Export exam status: Unable to copy file or something similar.
    Troubleshoot Connectivity with the Vivid HL7 Gateway
    Typical causes if Connectivity  with the Vivid HL7 Gateway failed:
    1.) Path provided in Export to HL7 export path was incorrect.
    2.) Mitra support personnel did not crea te the user account for the Vivid 7.
    NOTE: Required setup on the remote share:  User: E1c2h3o4C5l6i7e8n9t
    Password: u1l3t5r7a
    3.) The POLL_DIR directory on the Vivid HL7  Gateway was not shared or the permissions are 
    incorrect. Verify with Mitra support personnel.
    4.) Domain or other network issue.
    5.) If Report files (ReportXXX.chm)  are not exported to the Vivid HL7 Gatewa y, then the reports may 
    not have been stored when they were created on the EchoPAC PC. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 126 Section 3-11 - Connectivity Setup - Software v2.x
    3-11-10 Set Up Connection to a DICOM Se rver in a Network - Software v2.x
    No special service tools are required to do this set-up.
    3-11-10-1 Preparations Obtain the following information  from the network administrator:
    • DICOM server’s IP address, subnet mask (and gateway, if required).
    • Server’s Port Number.
    • Server’s AE Title.
    • IP address for the Vivid 7.
    3-11-10-2 Establish Network Connection to Hospital’s Network Connect an Ethernet cable from the Vivid 7 to the wall network outlet.
    3-11-10-3 Set up TCP/IP address on Vivid 7 1.) Press CONFIG
     (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    2.) Select  CONNECTIVITY
    .
    3.) Select  TCPIP
     tab.
    4.) Enter IP-Address, Subnet Mask (and gateway, if required). 
    5.) Select  SAVE SETTINGS
    . You will be asked to reboot,  but wait with that until later.
    3-11-10-4 DICOM Server’s Set Up  It may be necessary for the administrator to enter in formation about Vivid 7 into the DICOM server’s set-
    up. 
    This information is typical Vivid 7’s AE Title an d maybe also the IP address which administrator 
    provides. 
    (See  3-11-10-5 Vivid 7’s AE Title on page 3-126  for a description on how to obtain Vivid 7’s AE Title.)
    3-11-10-5 Vivid 7’s AE Title 1.) Select CONNECTIVITY
    .
    2.) Select the  SERVICES
     tab. 
    Vivid 7’s AE Title is listed below My  Computer (here it is VIVID7-000000). 
    Figure 3-133   Vivid 7’s AE Title
    Vivid 7’s AE Title 
    						