    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 107
    3-10-10-3      Query/Retrieve Verification (cont’d)
    Follow the Steps Below to Test th e DICOM Query/Retrieve Workflow 
    NOTE: This check uses both Ping and DICOM Ping. - Ping is used to verify the  TCP/IP connection to the server.
    - DICOM Ping is used to verify that  the DICOM application is answering
    1.) Select QueryRetrieve from the Selected devices list
    2.) Select Check.  - If the test pass, a pop-up dialog is  displayed, see left illustration in Figure 3-114. 
    NOTE: If the Check button fails directly, the AE Title is probably wrong. If the Check button fails after a long time (corres ponding to the timeout), the IP address or Port 
    Number is probably wrong.
    - Select OK to continue. A green check mark to th e left of the selected devices indicates that 
    the test was successfully passed. Figure 3-113   Check the DICOM Query/Retrieve workflow
    Figure 3-114   Test (Check) Passed
    QUERYRETRIEVE
    CHECK 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 108 Section 3-10 - Connectivity Setup - Software v3.x
    3-10-10-3      Query/Retrieve Verification (cont’d)       Follow the Steps Below to Test  the DICOM Query/Retrieve Workflow (cont’d)
    - If the test fail, a pop-up warning is  displayed, see left illustration in Figure 3-115. 
    Select OK to continue. A red “X” to the left of the Selected devices indicates that the test failed.
    If Retrieve is Failing ...
    If Retrieve is failing but Qu ery is functioning, verify if the scanner has been registered as an receiver on 
    the server. Figure 3-115   Test (Check) Failed 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 109
    Section 3-11
    Connectivity Setup - Software v2.x
    NOTE: FMI 76037 updates softwa re versions v2.3.0 and below to Software version v2.3.1.
    NOTE: If connected as a stand-alone network, only  scanner, EchoPAC PC Workstation and an optional 
    network printer in a separate network, you should use default delivery settings. 
    3-11-1 Introduction
    To be able to use the network functions when con nected to a hospital network, the scanner must have 
    a proper network address. 
    • Before you can set up the scanner, you need to collect some information. The  Pre-quote 
    Worksheet for DICOM Network Information on page 2-14  can be used for gathering this 
    information. Typical source for this in formation is the network administrator.
    • Follow the steps below to prepare the scanner for use on the network.
    3-11-2 Contents in this Section
    • 3-11-3 Compatibility on page 3-109
    • 3-11-4 Select TCP/IP Set-up Screen - Software v2.x on page 3-110
    • 3-11-5 Set the Scanner’s Network Information - Software Version v2.x on page 3-112
    • 3-11-6 Set the Remo te Archive’s Network Information - Software v2.x on page 3-113
    • 3-11-7 Save the New Settings - Software v2.x on page 3-113
    • 3-11-8 Verification of a Connection Set Up  to EchoPAC PC - Software v2.x on page 3-114
    • 3-11-9 HL7 Communication Setup for Software v2.x on page 3-122
    • 3-11-10 Set Up Connection to a DICOM Server in a Network - Software v2.x on page 3-126
    3-11-3 Compatibility
    3-11-3-1 Vivid 7 with So ftware Version v2.4.x
    Vivid 7 with Software Version v2.4.x (Syst em Software v1.5.1) can communicate with:
    • EchoPAC PC with software version v2.4.x
    • EchoPAC PC with software version v3.x
    • EchoPAC PC with software version v4.x
    • EchoPAC PC with software version v5.x
    • EchoPAC PC with software version v6.x
    • EchoPAC PC with software version v7.x
    • Image Vault 3
    • Image Vault 4
    • EchoServer 7 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 110 Section 3-11 - Connectivity Setup - Software v2.x
    3-11-4 Select TCP/IP Set-up  Screen - Software v2.x
    1.) Press CONFIG (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    Figure 3-116   Measuring and Analyzes - Global Level
    Connectivity 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 111
    3-11-4      Select  TCP/IP Set-up Screen -  Software v2.x (cont’d)
    2.) Select CONNECTIVITY (in the lower part of the window).
    3.) Select the TCP/IP TAB (it is named Tcpip). (See example in  Figure 3-118.)
    Figure 3-117   Connectivity Opening Screen
    Figure 3-118   TCP/IP Set-up Screen, Overview
    Connectivity
    TCP/IP Tab
    IP settings area for 
    the scanner
    Remote Archive 
    Setup area 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 112 Section 3-11 - Connectivity Setup - Software v2.x
    3-11-5 Set the Scanner’s  Network Information - Software Version v2.x
    In the IP settings area of the screen, enter the;
    • IP Address for scanner. (Default IP  Address from factory: 10.0.0.3).
    • Subnet Mask for scanner. (Default Subn et Mask from factory: 255.255.255).
    • IP address for Default Gateway. (Default  Subnet Mask from factory: 0.0.0.0).
    NOTE: For software v2.x, the Port Number has been set to 104. It can not be adjusted.
    .
    Figure 3-119   TCP/IP Set-up for Vivid 7
    IP Address
    Subnet Mask Default Gateway 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 113
    3-11-6 Set the Remote Archive’s Ne twork Information - Software v2.x
    In the Remote Archive Setup area of the screen (see example in  Figure 3-118.), enter the;
    1.) Remote Archive IP address. (Default  IP Address from factory: 10.0.0.4).
    2.) Remote Archive Name. (Default Remote Ar chive Name from factory: ECHOPAC7-000001).
    .
    3-11-7 Save the New Settings - Software v2.x
    1.) Press SAVE SETTINGS to save the new settings. 
    The new settings are saved to a common settings file.  After a restart, the settings are also included in 
    other screens.
    2.) Restart Vivid 7 to make the changes take place. Figure 3-120   TCP/IP Set-up fo
    r Remote Archive (EchoPAC PC)
    Remote Archive 
    IP address
    Remote Archive 
    Name Remote Archive 
    Setup area
    Save settings 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 114 Section 3-11 - Connectivity Setup - Software v2.x
    3-11-8 Verification of a Connection Set  Up to EchoPAC PC - Software v2.x
    1.) Press CONFIG (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 . 
    2.) Select  CONNECTIVITY
     (in the lower part of the window).
    3.) Select the  VIEWS
     tab, located in upper part of window,  to display the Current Dataflow Views 
    screen, see  Figure 3-121 .
     
    4.) Select the  DATAFLOW button to see all available data flows, see  Figure 3-122.
    5.) Select the  +
     beside the dataflow you want to check (here:  Remote Archive - Remote HD) to see its 
    contents.
     
    Figure 3-121   Current Dataflow Views screen
    Figure 3-122   Pre-configured Data flows in Vivid 7
    VIEWS tab
    CONNECTIVITY tab
    DATAFLOW 
    button
    DATAFLOW 
    button
    Select  + 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 115
    3-11-8      Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d)
    .
    The Remote Archive - Remote HD includes an EchoPAC PC with one Service also called  Remote 
    Archive - Remote HD . The name of this service indicates that the archive is located on a hard disk on 
    a remote server (EchoPAC PC is used as an server). 
    6.) Select the EchoPAC PC’s name to highlight it, see  Figure 3-123.
    7.) Select  CHECK
    , see Figure 3-124 , to verify if there is a connecti on between the ultrasound unit and 
    the EchoPAC PC Workstation. 
     
    Figure 3-123   Remote HD dataflow contents
    Figure 3-124   Verify dataflow connection
    Select (highlight) 
    the server’s 
    name
    Select  CHECK 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 116 Section 3-11 - Connectivity Setup - Software v2.x
    3-11-8      Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d) • In the example in  Figure 3-125, the red “X” next to the computer sign ( ), means that the check 
    failed. The cause may be;
    - the setup of the network information, either on Vivid 7 or on the EchoPAC PC, or
    - the network connection is broken somewhere between the Vivid 7 and the EchoPAC PC.
    NOTE: If the Check button fails directly, the AE Title is probably wrong. If the Check button fails after a long time (correspo nding to the timeout), the IP address or  Port 
    Number is probably wrong.
    • A green light next to the computer sign me ans that a network connection is in place.
    Figure 3-125   Verification of connection failed
    Red “X” 
    						