Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Vivid 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 858
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 5 - Components and Functions (Theory) 5 - 15
    5-3-10-2 Location in the Unit
    5-3-10-3 Input DC Voltages
     
    NOTICEPlease be careful about the positions when replacing  the TX card and the RX card. The position of the 
    two cards are interchanged between FEP1 and FEP2.
    Figure 5-11   TX128-5 Board: Location in Unit
    Table 5-4    Input, DC Voltages
    INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM
    +6 Va DC voltage, distributed via Motherboard (Backplane). 
    The “a” indicates that this vo ltage is used for analog circuits. J7-A22,B22,C22,D22,E22
    DC Power Supply > 
    Backplane 
    +15 Va DC voltage, distributed via Motherboard (Backplane).  J6-C20,D20,E20
    J7-A18,A19,A20 DC Power Supply > 
    Backplane 
    -5 Va DC voltage, distributed via Motherboard (Backplane).  J6-C3,D3,E3 DC Power Supply > 
    Backplane
    -15 VDC voltage, distributed via Motherboard (Backplane). J6-C1,D1,E1DC Power Supply > 
    Backplane
    GNDGround is distributed via Motherboard (Backplane).DC Power Supply > Backplane 
    HV1
    TX Voltage 1 
    Voltage can be programmed to vary from +/- 2.5 V to +/- 95V Used to drive the transmitters in 2D Mode and M ModeHV1N: J6-A15HV1P: J6-A24TX Power Supply >  Backplane
    HV2 TX voltage 2
    Voltage can be programmed to vary from +/- 2,5 V to +/- 95V Used to drive the transmitters in Doppler (CW/PW) mode  and in Color Flow Mapping (CFM) mode HV2N: J6-A18
    HV2P: J6-A21 TX Power Supply > 
    Backplane
    FEP2 with two TX boards
    FRONT OF 
    SCANNER
    TX128 
    BOARDS (2X) 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    5 - 16 Section 5-3 - a  Front-End Processor (FEP)
    5-3-10-4 Control Signals
    5-3-10-5 Outputs
    5-3-10-6 Fuses, Jumpers and Dip-switches None Table 5-5    Input, Control Signals 
    INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM
    FE_D0-15 Front-End Data bus, Digital TTL J1 RFI > Backplane
    FE_PA0-9 Front-End Page Address bus, Digital TTL J1 RFI > Backplane
    FE_SA0-5 Front-End Device Select Address bus, Digital TTL J1 RFI > Backplane
    TX_TRIG_LTransmit trigger, Digital TTLJ1-D4RFI > Backplane
    Table 5-6    Output TX pulses
    OUTPUTDESCRIPTIONCONNECTED TO
    XD1 - 128  128 transmitter channels via the connectors to the XD Bus 
    Board TX128 > XD Bus Board > Relay Board 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 5 - Components and Functions (Theory) 5 - 17
    5-3-10-7 LEDs on the TX128 board
     
    Figure 5-12   TX128 Board: LEDs
    Table 5-7    LEDs on the TX128 Board
    LED #LED COLORDESCRIPTIONNORMAL STATE
    LD1  - Red Board Failure Normally OFF
    LD2  - Green Board OK Normally ON
    LD3  - Yellow Front-End Bus Protocol Error Normally OFF
    LD4 - YellowScanNormally OFF
    LD5 - YellowSpa reNormally OFF
    LD5LD4LD3
    LD2 LD1 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    5 - 18 Section 5-3 - a  Front-End Processor (FEP)
    5-3-10-8 Test points on the TX128 board
     
    NOTE: The Voltages on these test points are approxi mately one tenth (1/10) of the TX Voltages used 
    by the transmitter circuits. (The exact [calcu lated] value is 10/101). The measured values 
    depends on selected scanning mode  and the Acoustic Power adjustment.
    Figure 5-13   TX128 Board: Test points
    Table 5-8    Test points on the TX128-5 Board
    TESTPOINT #DESCRIPTIONTYPICAL VALUE
    TP11 HV1+ /10 Between +0.25 VDC and +9.5 VDC
    TP12 GROUND 0 VDC
    TP13 HV1- /10 Between -0.25 VDC and -9.5 VDC
    TP14HV2- /10Between -0.25 VDC and -9.5 VDC
    TP15HV2+ /10Between +0.25 VDC and +9.5 VDC
    TP12
    TP14
    TP15
    TP11
    TP13 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 5 - Components and Functions (Theory) 5 - 19
    5-3-11 Transmitter Board, TX12 8 (version used before BT’05)
    5-3-11-1 General Description
    The Transmitter board (TX128) contains 128 individually controlled transmit channels and provides 
    transmit pulses via the XDBUS board to the Relay boa rd and then to the transducer array (the probes).
    Frequency-, delay- and pulswidth parameters for the TX128 board are received on the FE_BUS from 
    the Front End Controller (FEC) on units with FEP1,  and from the Radio Frequency Interface (RFI) on 
    units with FEP2. The parameters are decoded and stored in the cache.
    A pulse, TX_TRIG_L, trigs the Tran smit Pulse Generators. The transmit pulses are then amplified to the 
    correct level in the 128 TX transmitters. 
    High Voltage 1 and High Voltage 2 from the TX Power supplies the TX Transmitters with the needed 
    voltages to generate the correct power. 
    The VDRIVER voltage is used by the TX Dr ivers (drivers for the TX transmitters).
    Two TX128 boards are used on units with the 3D option. Figure 5-14   Block Diagram for the TX128 board
    8
    TRANSMIT
    (128 CH.)
    TX
    128
    TRANSMITTERS CACHE
    FRONT END
    INTERFACE
    PULSE GEN.
    FROM TX POWER
    HV1&2
    TX_TRIG_L FROM FEC/RFI
    FE_BUS
    XD1-128
    TO RLY VIA XDBUS HV1P
    HV2P
    HV2N
    HV1N
    VDRIVER
    (+10V)
    FROM DC POWER
    FROM FEC/RFI 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    5 - 20 Section 5-3 - a  Front-End Processor (FEP)
    5-3-11-2 Location in the Unit
    5-3-11-3 Input DC Voltages
     
    NOTICEPlease be careful about the positions when replacing  the TX card and the RX card. The position of the 
    two cards are interchanged between FEP1 and FEP2.
    Figure 5-15   TX128 Board: Location in Unit
    Table 5-9    Input, DC Voltages
    INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM
    +6 Va DC voltage, distributed via Motherboard (Backplane). 
    The “a” indicates that this vo ltage is used for analog circuits. J7-A22,B22,C22,D22,E22
    DC Power Supply > 
    Backplane 
    +10 Va DC voltage, distributed via Motherboard (Backplane). 
    The + 10Va is only used by this board. J6-C5,D5,E5
    DC Power Supply > 
    Backplane 
    -5 Va DC voltage, distributed via Motherboard (Backplane).  J6-C3,D3,E3 DC Power Supply > 
    Backplane
    -15 VDC voltage, distributed via Motherboard (Backplane). J6-C1,D1,E1DC Power Supply > Backplane
    GNDGround is distributed via Motherboard (Backplane).DC Power Supply > Backplane 
    HV1
    TX Voltage 1 
    Voltage can be programmed to vary from +/- 2.5 V to +/- 80V Used to drive the transmitters in 2D Mode and M ModeHV1N: J6-A15HV1P: J6-A24TX Power Supply >  Backplane
    HV2 TX voltage 2
    Voltage can be programmed to vary from +/- 2,5 V to +/- 40V Used to drive the transmitters in Doppler (CW/PW) mode  and in Color Flow Mapping (CFM) mode HV2N: J6-A18
    HV2P: J6-A21 TX Power Supply > 
    Backplane
    TX128 
    BOARD
    FRONT OF 
    SCANNER
    FEP1
    FEP2 with two TX boards
    FRONT OF 
    SCANNER
    TX128 
    BOARDS (2X) 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 5 - Components and Functions (Theory) 5 - 21
    5-3-11-4 Control Signals
    5-3-11-5 Outputs
    5-3-11-6 Fuses, Jumpers and Dip-switches None Table 5-10    Input, Control Signals 
    INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM
    FE_D0-15 Front-End Data bus, Digital TTL J1 Front-End Controller/RFI > 
    Backplane
    FE_PA0-9 Front-End Page Address bus, Digital TTL J1 Front-End Controller/RFI > 
    Backplane
    FE_SA0-5 Front-End Device Select Address bus, Digital TTL J1 Front-End Controller/RFI > 
    Backplane
    TX_TRIG_LTransmit trigger, Digital TTLJ1-D4Front-End Controller/RFI > 
    Backplane
    Table 5-11    Output TX pulses
    OUTPUTDESCRIPTIONCONNECTED TO
    XD1 - 128  128 transmitter channels via the connectors to the XD Bus 
    Board TX128 > XD Bus Board > Relay Board 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    5 - 22 Section 5-3 - a  Front-End Processor (FEP)
    5-3-11-7 LEDs on the TX128 board
     
    Figure 5-16   TX128 Board: LEDs
    Table 5-12    LEDs on the TX128 Board
    LED #LED COLORDESCRIPTIONNORMAL STATE
    LD1  - Red Board Failure Normally OFF
    LD2  - Green Board OK Normally ON
    LD3  - Yellow Front-End Bus Protocol Error Normally OFF
    LD4 - YellowScanNormally OFF
    LD5 - YellowSpa reNormally OFF
    LD5LD4LD3
    LD2 LD1 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 5 - Components and Functions (Theory) 5 - 23
    5-3-11-8 Test points on the TX128 board NOTE: If the test points below have not been mounted, and you want to verify the TX voltages, you  must use the TXP test, available via the Di agnostics Test Screen. Please refer to 7-8-3 System 
    Test on page 7-66 .
     
    NOTE: The Voltages on these test points are approxi mately one tenth (1/10) of the TX Voltages used 
    by the transmitter circuits. (The exact [calcu lated] value is 10/101). The measured values 
    depends on selected scanning mode and the Acoustic Power adjustment. Figure 5-17   TX128 Board: Test points
    Table 5-13    Test points on the TX128 Board 
    TESTPOINT #DESCRIPTIONTYPICAL VALUE
    TP11 HV1+ /10 Between +0.25 VDC and +8.0 VDC
    TP12 GROUND 0 VDC
    TP13 HV1- /10 Between -0.25 VDC and -8.0 VDC
    TP14HV2- /10Between -0.25 VDC and -4.0 VDC
    TP15HV2+ /10Between +0.25 VDC and +4.0 VDC
    TP12
    TP14
    TP15
    TP11
    TP13 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    5 - 24 Section 5-3 - a  Front-End Processor (FEP)
    5-3-12 Transducer Bus Boards, XD BUS
    5-3-12-1 General Description The two XD BUS Boards plug into the rear edge  connectors on the Relay Board, the TX128 Board(s) 
    and on the RX-128 Board. 
    FEP2 ONLY:  The XD Bus boards used on units with two TX  boards, with 4D (RT3D), are not backward 
    compatible and cannot be used on units with one TX card (without 4D (RT3D)). And the XD Bus board 
    used on units with one TX board cannot be used on units with two TX boards.
    FEP2 ONLY:  The connectors for the Relay Board connection, are color coded to match the colors on 
    the connectors on the Relay Board, to reduce th e risk of mixing the two types of XD Bus boards.
    The XD signals to and from the probes (via the Relay Board) are routed via this board. 
    5-3-12-2 Location in the Unit
    NOTICETwo different versions of the XD Bus boards are us ed due to the different positions for TX board and 
    RX board on FEP1 and FEP2. The two types can NOT be interchanged.
    Figure 5-18   XD BUS Boards : Location in Unit with FEP1.
    XD BUS BoardFront of 
    Scanner
    XD BUS Board 
    						
    All GE manuals Comments (0)