    Chapter 3 - System Setup 3 - 77
    3-9-16-3 Create a New Dataflow with Worklist and DICOM Storage (for CA-1000 sites)  This dataflow is intended for Vivid 7 scanners in a DI COM network, where the data should not be stored 
    locally on the scanner. A site with the integrated  EchoPAC PC and CA-1000 is an example of this.
    To create a new dataflow with Worklist an d DICOM Storage, do the following steps:
    1.) Add a new dataflow in Config screen, Conn ectivity, Dataflow tab. Enter a name, e.g. Worklist - 
    DICOM Storage . 
    2.) Select  OK.
     
    3.) Add the Worklist inpu t. Configure (at least) IP address, AE title  and Port Number . 
    4.) Select  OK.
     
    5.) Add the Dicom Storage output. 
    6.) Configure (at least)  IP address, AE title  and Port Number .
     If the site has inte grated EchoPAC PC 
    and CA-1000, enable  Allow SR and Allow SR Private Data .
     
    7.) Select OK.
    8.) If this dataflow should be the default at start-up, enable  Default for the dataflow.
     
    Figure 3-76   Dicom SR Settings 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 78 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-17 Query/Retrieve (Q/R) Setup  for Software v7.x/v6.x/v5.x/v4.x
    3-9-17-1 Overview
    The Query/Retrieve function  makes it possible to search for and  retrieve DICOM data from a DICOM 
    server for further analysis on the Vivid 7. 
    NOTE: You may have to set up Vivid 7 as a destination on the server.
    3-9-17-2 Query/Retrieve Setup on the Vivid 7 with Software 
    v7.x/v6.x/v5.x/v4.x
    1.) Press  CONFIG (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
    2.) Select Connectivity
    3.) Select Dataflow
    4.) Select Query Retrieve from the Name pull down menu
    Figure 3-77   Select Query Retrieve
    NAME PULLDOWN 
    MENU
    QUERY RETRIEVE 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 79
    3-9-17-2      Query/Retrieve Setup on the  Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d)
    In  Figure 3-78 , the selected dataflow, Query/Retrieve, is shown.
    5.) Select QueryRetrieve so it is hi ghlighted and then select Propertie s to display the Properties dialog.
    Figure 3-78   Query/Retrieve workflow
    Figure 3-79   Select Qu ery/Retrieve’s Properties
    SELECT 
    QUERY RETRIEVE 
    AND THEN SELECT 
    PROPERTIES
    QUERY RETRIEVE 
    PROPERTIES DIALOG 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 80 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-17-2      Query/Retrieve Setup on the  Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d)
    6.) Select the IP-address down-arrow to choose t he DICOM Query/Retrieve server from the pull down 
    menu. In some cases, the server to use  is the same as used for DICOM Storage.
    If the server to use is missing  from the list, select  from the pull down menu and edit the 
    setup for one of the predefined servers.
    7.) Enter the correct AE Title and Port Number for th e DICOM Query/Retrieve server in the respective 
    fields in the Query/Retrieve screen. Figure 3-80   DICOM Query/Retrieve properties
    IP-ADDRESS 
    DOWN-ARROW
    AE TITLE
    PORT NO
    DO NOT SELECT THE 
    SEARCH CRITERIA  
    BUTTON IF NOT TOLD 
    SO BY THE ONLINE 
    CENTER. 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 81
    3-9-17-2      Query/Retrieve Setup on the  Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d)
    Change Search Criterias
    It is possible to set up special Se arch Criterias for DICOM Query/Retrieve. In most cases you may leave 
    the Search Criterias as is , and skip this adjustment.
    Follow the steps below to change the Search Criterias parameters:
    1.) Select Search Criterias
    2.) Select the correct tag from the Select Tag pull-down menu.
    3.) If needed, type in the value.
    NOTE: In this example, tag 00080060 Modality has been added to the list with value “US”. This may be  useful if the studies cont ain non-ultrasound images.
    4.) Select Add to List
    5.) Select OK to close the Search Criterias window. Figure 3-81   Select Search Criterias 
    SEARCH CRITERIAS
    ADD TO LIST
    REMOVE 
    SELECT TAG  
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 82 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-17-3 Query/Retrieve Verification
    Follow the Steps Below to do a First Test (TCP-IP Ping) of the Connection
    1.) Select  Query retrieve  and then select  Properties.
    2.) Select the “Smiley” button to Ping the server.
    NOTE: The smiley checks if the remote server is  accessible (ping). It is not a DICOM Echo (DICOM 
    ping), so it does not check AE title or port number.
    3.) If the network connection to the server is  OK, it will be illustrated by an smiling “Smiley”  .
    A sad Smiley indicates that the  network connection is failing. 
    Typical causes:  
    - Network cable not connected.
    - Configuration error(s).
    4.) When ready, select OK to close the DICOM  Query/Retrieve properties and save changes.
    Figure 3-82   DICOM Query/Retrieve properties
    SMILEY 
    (NETWORK 
    PING)
    OK 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 83
    3-9-17-3      Query/Retrieve Verification (cont’d)
    Follow the Steps Below to Test th e DICOM Query/Retrieve Workflow 
    NOTE: This check uses both Ping and DICOM Ping. - Ping is used to verify the  TCP/IP connection to the server.
    - DICOM Ping is used to verify that  the DICOM application is answering.
    1.) Select QueryRetrieve from the Selected devices list.
    2.) Select Check.  - If the test pass, a pop-up dialog, reporting the results as OK, is displayed, see  Figure 3-84. 
    - Select OK to continue.  A green check mark to the left of the Selected  devices indicates that the test was successfully 
    passed. Figure 3-83   Check the DICOM Query/Retrieve workflow
    Figure 3-84   Test (Check) Passed
    Figure 3-85   Test (Check) Passed
    QUERYRETRIEVECHECK 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 84 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-17-3      Query/Retrieve Verification (cont’d)       Follow the Steps Below to Test  the DICOM Query/Retrieve Workflow (cont’d)
    - If the test fails, a pop-up dialog reporting the resu lts as “0 of 1 OK” is displayed. In the example 
    in  Figure 3-86,  the warning message pop up almost at onc e, and it states that for the Server 
    Check (Ping), 1 of 1 is OK but the Device Che ck failed (0 of 1 OK). This indicates that the 
    network connection functions OK, but the application on the remote server is not answering.
    If the test fails directly, as  described in this example, the AE Title is probably wrong.
    A red “X” to the left of the Selected devices indicates  that the test failed, but it doesn’t indicate the cause.
    - If the Check button fails after a long time (corresponding to the timeout), the Warning will look  like the example in  Figure 3-88, where both tests failed.
    In this case, the IP address or Port Number is probably wrong.
    If Retrieve is Failing ...
    If Retrieve is failing but Qu ery is functioning, verify if the scanner has been registered as an receiver on 
    the server. Figure 3-86   Test (Check) Fails directly
    Figure 3-87   Test (Check) Passed
    Figure 3-88   Both Tests Failed 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    Chapter 3 - System Setup 3 - 85
    3-9-18 Mapping of Parameters  from Vivid 7 to DICOM
    3-9-18-1 Search Screen
    Figure 3-89   Search / Create Patient Screen
    Table 3-17    Search / Create Patient Screen
    ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME
    1 LAST NAME, FIRST NAME (0010,0010) PATIENT’S NAME
    2 SEX (0010,0040) PATIENT’S SEX
    3 ACCESSION (0008,0050) ACCESSSION NUMBER
    4PAT I E N T  I D(0010,0020)PATIENT ID
    5BIRTHDATE(0010,0030)PAT I E N T ’ s  B I R T H D AT E
    6STUDY ID(0020,0010)STUDY ID
    THE COLUMN 
    EXAM DATE, WAS 
    INTRODUCED FOR 
    SOFTWARE 
    VERSION V.6.0.
    1
    2
    3
    4
    5
    6 
    						
    							
    GE HEALTHCARE
    DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
    3 - 86 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
    3-9-18-2 Patient Information Screen Figure 3-90   Patient Information Screen
    Table 3-18    Patient Information Screen
    ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME
    1 OPERATOR (0008,1070) OPERATOR’S NAME
    2 REF. DOC (0008,0090) REFFERING PHYSICIAN’S NAME
    3 DIAGN.PHYS. (0008,1050) PERFORMING PHYSICIAN’S NAME
    4WARD/DEPT(0008,1040)INSTITUTIONAL DEPARTMENT NAME
    5WEIGHT(0010,1030)PATIENT’S WEIGHT
    6DESCRIPTION(0008, 1030)STUDY DESCRIPTION
    7 HEIGHT (0010,1020) PATIENT’S SIZE
    8 CONTRAST (0018,0010) CONTRAST/BOLUS AGENT
    1
    2
    3
    4
    5
    6
    7
    8 
    						